Synopsis
In CEST imaging, when saturation time is not sufficient long (empirically smaller than 0.8s), rotation effect would appear and contaminate with saturation effect, making the signal unavailable for further analysis. Considering the inhomogeneity of $$$B_{0}$$$ and $$$B_{1}$$$ in reality, we propose a novel Fixed Angle Single Rotation CEST(FASR-CEST) sequence to overcome the restriction, successfully reducing the saturation time to about 0.5s while keeping identical effect as CEST sequence with long saturation time, with the help of analytical calibration method in another abstract. Effect of the sequence is verified with vitro and in vivo data. PURPOSE
Rotation
effect is another effect during saturation apart from saturation effect, referring
to the process of magnetization perpendicular to the saturation direction ($$$\overrightarrow{\omega_{eff}}$$$, for detail see below). The rotation effect would contaminate CEST signal when saturation
time is not sufficiently long(empirically $$$t_{sat}<0.8s$$$), which
places constrains on repeatable quantification in practices
1. Traditionally
this requires restrictively long saturation time, while the concomitant SAR and
long scan time would impose a concern. Although sophisticated quantification
model can describe rotation effect
2, it is unavailable to be applied in reality.
In this work, a novel Fixed Angle Single Rotation CEST (FASR-CEST) method is
proposed which could effectively diminish the rotation effect even if a short
saturation period (<0.5s).
METHOD
Based
on Zaiss’es analytical solution of Z value
2,3, process of chemical saturation transfer with saturation
pulse amplitude
$$$B_{1}$$$ and offset $$$\Delta \omega$$$ could
be treated as $$$T_{1\rho}$$$ relaxation
along the direction of $$$\overrightarrow{\omega_{eff}}$$$ ($$$\overrightarrow{\omega_{eff}}=(\gamma B_{1}, 0, \Delta \omega)$$$),
combined with $$$T_{2\rho}$$$ relaxation
perpendicular to the direction of $$$\overrightarrow{\omega_{eff}}$$$
(Fig. 1b). If there is still nonzero magnetization on $$$T_{2\rho}$$$ direction ($$$M_{T_{2\rho}}$$$) at
the end of saturation pulse, it has net component in Z axis and the rotation
effect takes place. Short saturation time will very likely induce rotation
effect. To avoid this,
rotation pulses before or after rectangular saturation may be proposed to
decrease $$$M_{T_{2\rho}}$$$ or its projection to Z axis
3,4. Ideally, the flip
angle of the rotation pulse satisfies $$$\beta=\theta$$$, . Considering the inhomogeneity of both $$$B_{0}$$$ and $$$B_{1}$$$, the direction of $$$\overrightarrow{\omega_{eff}}$$$ is varied. Then, the flip angle β should not
always be $$$\beta=\theta$$$ as supposed, which would still induce rotation effect. To
overcome the defect, only a rectangular rotation pulse before saturation pulse
is used (Single Rotation). The rectangular pulse is used as it can decrease
difference between β and θ due to unideal $$$B_{1}$$$. Besides, because of inhomogeneous $$$B_{0}$$$, the rotation angles changes with $$$\Delta \omega$$$. Thus, the interpolation method for $$$B_{0}$$$ correction would no longer be suitable. To overcome
this issue, the rotation angle is fixed to a single value according to the
offset frequency the researcher is most interested in (Fixed Angle)
(Fig. 1a). To compare the signal from
FASR-CEST sequence (small $$$t_{sat}$$$) with that from conventional CEST, an analytical quantificationmethod (for details see our another abstract) is used, which
converts $$$MTR_{asym} $$$ value to $$${\Delta R}_{1\rho}$$$ or $$${\Delta R}_{1\rho}^{calib}$$$.
EXPERIMENTS
Firstly, phantoms with
different solutes (glucose and glutamate) densities and different relaxation
times (adjusted with MnCl
2) were prepared and used to test effects
of FASR-CEST sequence. Continuous wave rectangular saturation pulse followed by
a SE-EPI sequence was used for CEST Imaging. Conventional CEST sequence with $$$B_{1}=1\mu T$$$, $$$TR=2s$$$, $$$t_{sat}=1.5s$$$ was used first acquired. Then, a FASR-CEST sequence
with $$$B_{1}=1\mu T$$$, $$$TR=2s$$$, $$$t_{sat}=0.5s$$$ and fixed angle $$$\beta=14.93^{o}$$$ (corresponding to offset frequency around 1.25ppm) was
scanned. As comparison, the same CEST sequence with a reduced saturation period ($$$t_{sat}=0.5s$$$) was also scanned. Frequency offsets were from -6ppm to +6ppm. After phantom
experiment,
in vivo data was
acquired on a healthy subject. Three pairs of TR&
for CEST imaging were used: $$$TR=2s, t_{sat}=1.5s$$$; $$$TR=1.5s, t_{sat}=1s$$$; $$$TR=7s, t_{sat}=5s$$$, and FASR-CEST sequence was also
scanned with $$$TR=2s, t_{sat}=0.5s$$$ and fixed angle $$$\beta=14.93^{o}$$$ (corresponding to around 1.25ppm). For each experiment, corresponding $$$B_{1}$$$ map, $$$T_{1}$$$ map and $$$T_{2}$$$ map were also acquired for quantification.
RESULT
$$${\Delta R}_{1\rho}^{calib}$$$ urves from two different phantoms and two sequences
(CEST with $$$t_{sat}=1.5s$$$ and FASR-CEST) are shown in
Fig.
2b. It can be seen that the
$$${\Delta R}_{1\rho}^{calib}$$$
value from FASR-CEST around 1.25ppm fits well to that
from CEST with
$$$t_{sat}=1.5s$$$.
Besides
$$${\Delta R}_{1\rho}^{calib}$$$ map on 1.25ppm from FASR-CEST is also nearly the same with
the reference CEST
(Fig. 2c, 2d). However, when a short saturation period $$$t_{sat}=0.5s$$$ was used, rotation effect could be obviously observed in
an image with almost no effective contrast
(Fig. 2e). The effect of FASR-CEST would be similar
in vivo data.
(Fig. 3)DISCUSSION and CONSLUSION
Reproducible quantification in CEST is hurdled by the
rotation effect. In practice, it is often non-practical to have sufficiently
long saturation period for the rotation effects to be dissipated. In this work
a FASR-CEST sequence was proposed to overcome this constrain, phantom and in
vivo experiment demonstrated that even with a short saturation period
relatively constant CEST quantification can be obtained under varying imaging
parameters. FARS-CEST is easy to implement and also help to maintain a low SAR
level. Further optimization of this prototype may help its application to other
more general saturation strategies.
Acknowledgements
No acknowledgement found.References
1.Moritz Zaiss, Zhongliang Zu, Junzhong Xu et al., A combined analytical solution for chemicalexchange saturation transfer and semi-solidmagnetization transfer, NMR Biomed. 2015; 28(2): 217-230
2. Moritz Zaiss, Peter Bachert, Chemical exchange saturation transfer (CEST) andMR Z-spectroscopy in vivo: a review of theoreticalapproaches and methods, Phys. Med. Biol. 2013; 58(22): 221-269
3. Moritz Zaiss, Peter Bachert, Exchange-dependent relaxation in the rotatingframe for slow and intermediate exchange –modeling off-resonant spin-lock and chemicalexchange saturation transfer, NMR Biomed. 2013; 26(5): 507-518
4. Tao Jin, Seong-Gi Kim, Advantages of chemical exchange-sensitivespin-lock (CESL) over chemical exchangesaturation transfer (CEST) for hydroxyl– andamine–water proton exchange studies, NMR Biomed., 2014; 27(11): 1313-1324