Nithin N Vajuvalli1, Amaresh Konar1, Shivaprasad Ashok Chikop1, Darshan S Keelara1, Ashwini Kumnoor1, and Sairam Geethanath1
1Medical Imaging Research Centre, Dayananda Sagar Institutions, Bangalore, India
Synopsis
DCE MRI technique is widely used
in tissue characterization and tissue perfusion. Current work focuses on the design and development of a low
cost in vivo DCE phantom for
providing a reference standard for quantitative validation and the ability to
generate Signal Intensity curves similar to arterial input functions. Poly
vinyl Alcohol material was used in the phantom to mimic tissue perfusion
characteristics and obtain signal intensity curves for different flow rates. We
obtained similar curves compared to Arterial Input Function and were able to
control flow rates through different tube and pore sizes.Purpose
DCE MRI provides information about Pharmacokinetics (PK) in
tissues which aids in detection, staging and looking at the behaviour of
vascularized tumours. Current work focuses on the design and development of in vivo DCE phantom for providing a reference
standard for quantitative validation and the ability to generate Signal Intensity
(SI) curves that mimic Arterial Input Functions (AIF) of different anatomies.
Methods & Materials
The DCE MRI phantom setup is shown in the figure 1. The phantom
has three acrylic tubes (diameter: 2mm, 2mm and 3mm) for water flow which
mimics the arteries. Small pores of 1mm diameter was made at the centre of all three
tubes to mimic leakiness of Contrast Agent (CA) to the tissue mimicking
material.
The three tubes were connected to each other making a single
inlet and outlet as shown in figure 1(a-b). Further the inlet was divided into
two tubes by using the hose, where one of tubes is connected to the main water
source providing constant water flow and the other tube was used to inject the CA
using a syringe. Copper sulphate (CuSO4 0.25g/100ml) was used as the
CA. The water outlet was connected to a bucket outside the scanner.
Tissue mimicking was performed using Poly Vinyl Alcohol (PVA)1.10% of PVA solution was prepared by boiling PVA salt in water
at 1000 C. The mixture was heated until the solution was clear to obtain
the PVA cryogel. The PVA cryogel was cooled gradually to room temperate with
the lid closed so that the water content in cryogel is intact. The PVA cryogel
was poured into a container and placed at -180 C for 3 freeze-thaw cycles.
Multiple pores were made to make the material porous in order to obtain the wash-in and wash-out
curves. The set-up box was made from acrylic plastic and was machined in house.
The central part of the three tubes were then covered with the PVA material
thus prepared which can be seen in figure 1(b). The total bill of material for
all the components used to build the phantom was less than USD $75 and was
assembled with readily available material and tooling. Phantom experiments were
conducted on 1.5T Siemens MR scanner to obtain the SI curves. Images were
acquired using the 3D gradient recalled acquisition with parameters of
TR/TE=8.11/4.76 ms, number of average=1, 20 slices with 30 time points with
temporal resolution of 12 s.
The total scan time was about 6 mins with the CA being
injected after 1 min. SI curves were plotted to depict the wash-in and wash-out
characteristics of the phantom for the three different perfusion components
(seen in figure 1) with the one in the middle being able to let the flow
through at a constant rate. This served as the control for the other two
components. Tissue mimicking PVA was needle pricked to enable porosity around
the ROI1 and ROI3 region to ensure that CA leaks out rapidly
from that particular region.
Results
Figure
2(b) depicts the SI plots for the three different regions of tissue marked with
regions of interest in figure 2(a) with corresponding colors. It can be
observed that SI curves for the two ROIs were able to produce changes in SI with
respect to the control over the time for which data was being acquired. The curves
are similar to arterial input functions with different rates similar to that of
real in vivo data where there is a rapid
wash in and wash out depicted in Figure 2(c) for population averaged AIF
2.
It can be observed that the tube with 3mm diameter (ROI
3) resulted
in increased flow rate as is expected with the signal change rate of of R
1=18.71
(change in signal intensity /s) for ROI
3, R
2=5.11 for ROI
1
and R
3=1.26 for ROI
2.
Discussion
A perfusion
phantom that mimics the AIF curve has been developed at low cost. The increased
flow rate in tube 3 can be attributed due to the increase in the diameter of
the tube. The SI profile obtained from
ROI
2 is different to the other two ROIs due to the absence of
hand-made/needle pricked pores. In conclusion, a SI curve similar to that of
AIF was produced. Future studies of the phantom would be to mimic the different
tumour tissue curves, where there is a sudden wash in and slow wash out. This
would be then subject to application of relevant PK modelling based on the
understanding of the nature of perfusion of the tumour type.
Acknowledgements
Department of Science and Technology(DST). DST/TSG/NTS/2013/100References
[1] Dita Puspita Sari et.al J. FIS. DAN APL.,
VOL. 9, NO. 2, JUNE 2013
[2] Parker GJ et. al MRM 2006