Joseph G Woods1, Michael A Chappell2, and Thomas W Okell1
1FMRIB Centre, NDCN, University of Oxford, Oxford, United Kingdom, 2Institute of Biomedical Engineering, University of Oxford, Oxford, United Kingdom
Synopsis
This
study aims to investigate the ideal labeling and delay parameters for single-delay, multi-delay and
time-encoded pCASL and compare their theoretical performance in cerebral
blood flow (CBF) and bolus arrival time (BAT) estimation under realistic noise
levels over two BAT ranges: healthy (500ms-1500ms) and disease (500ms-3000ms).Introduction
Pseudo-continuous
arterial spin labeling (pCASL) is a non-invasive technique that magnetically
inverts blood for use as an endogenous tracer for perfusion quantification. In conventional
pCASL, only a single post label delay (PLD) is used. Multiple PLDs can also be
used and a suitable kinetic model fit to the data to estimate both the CBF and
the bolus arrival time (BAT), with the caveat that there will be less images at
each PLD to average over in a given scan time.
A
recent development, time-encoded (TE) pCASL1, splits the labeling
period into blocks using a Hadamard-type encoding scheme, resulting in greater
noise averaging compared to single and multiple delay methods, but the
effective labeling duration of each block is shorter, resulting in a smaller
signal. To counteract the effect of greater T1 decay for earlier TE blocks, Teeuwissee
et al.2 proposed adjusting the block durations such that the
signal from each block was equal at image excitation, assuming the signal only
decays with the T1 of blood (TE-T1adj).
The
aim of this study is to investigate ideal labeling and delay parameters for
these methods and compare their theoretical performance in cerebral blood flow
(CBF) and bolus arrival time (BAT) estimation under realistic noise levels over
two BAT ranges: healthy (500ms-1500ms) and disease (500ms-3000ms).
Methods
Simulations were performed using the general kinetic
model3; T1 of tissue (T1t) was assumed equal to T1 of blood (T1b) for
single-delay as a best-case scenario. Total scan
time was kept to 5 minutes. Readout time was assumed to be 1000ms. The number
of repeats available was calculated for each implementation. Noise standard deviation
(SD) was calculated from multi-delay pCASL data4 and added to all
control and tag images before subtraction or decoding. True CBF was kept
constant at 60ml/100g/min. A range of labeling durations (LDs) and PLDs were
tested for all methods. The TR for multi-delay varied with the PLD. A PLD of
49ms was used for both TE methods.
Perfusion
was calculated directly5 for single delay and by least-square
minimization for multi-timepoint data with a Metropolis-Hastings algorithm (500
iterations) for stable estimation of the maximum likelihood distribution; the
mode of a 50-bin histogram was taken for the parameter estimates. The bounds on
the estimations were CBF: 0-240 (ml/100g/min), BAT: 0-last timepoint (s).
Results and Discussion
We found that a LD of 1800ms was more accurate for
multi-delay pCASL. Figure 1 shows the
larger PLD range (250-2750ms) for multi-delay results in more accurate
perfusion estimates for the majority of the disease BAT range while not significantly
affecting the healthy range. The BAT estimates were comparable between the two
methods. The drop in late BAT error for 250-1500ms is because the last time-point
is earlier than for the longer range, resulting in the upper bound of the
fitting algorithm reducing the overestimation of BAT.
For both TE methods, we found that a LD of 4500ms was most accurate over
the disease range. Figure 2 demonstrates that 7-block TE pCASL and 11-block
TE-T1adj pCASL are best for CBF and BAT estimates over the entire BAT range.
The last time-points in 7-block TE-T1adj are relatively sparse so do not sample
the inflow well when
arrival is delayed, which results in the spike in CBF and BAT
estimations.
For single-delay we found that a PLD of
2600ms and a LD of 4000ms gave the most robust results for the disease range tested,
in agreement with Zun et al.6. This single-delay protocol was compared with the optimal
multi-delay, TE and TE-T1adj protocols as described above. Figure 3 and table 1 present the results of
our methods comparison. For the healthy BAT range, the CBF estimates for
multi-delay had a smaller SD, but TE had more accurate BAT estimates. Single-delay performed well in
the disease range because of the long PLD used, while TE-T1adj produced
the greatest CBF accuracy for the multi-timepoint methods. Multi-delay had the
highest BAT estimate accuracy averaged across the disease range. The SD of TE-T1adj
BAT estimates increased much more than the other methods at long BATs due to
the sparseness of samples at these times.
Conclusions
We
have demonstrated that, theoretically, TE methods can produce CBF and BAT
estimates comparable, or in certain cases better, than multi-delay pCASL. This
is in contrast to a
previous study7, though it should be
noted that different simulated parameters have been used. We hope to validate
these findings in-vivo.
Acknowledgements
Funding source: EPSRCReferences
1. GuentherM. Highly efficient accelerated acquisition of perfusion inflow series by cycled arterial spin labeling. ISMRM Abstract, 2007.
2. TeeuwisseWM, Schmid S, Ghariq E, Veer IM, and van Osch MJP. Time-encoded pseudocon- tinuous arterial spin labeling: Basic properties and timing strategies for human applications. Magn Reson Med, 00:00–00, 2014.
3. Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, and Edelman RR. A general kinetic model for quantitative perhsion imaging with arterial spin labeling. Magn Reson Med, 40:383– 396, 1998.
4. Okell TW, Chappell MA, Kelly ME, and Jezzard P. Cerebral blood flow quantification using vessel-encoded arterial spin labeling. J Cereb Blood Flow Metab, 33:1716–1724, 2013.
5. Alsop DC, Detre JA, Golay X, G¨ unther M, Hendrikse J, Hernandez-Garcia L, Lu H, Mac-Intosh BJ, Parkes LM, Smits M, van Osch MJP, Wang DJJ, Wong EC, and Zaharchuk G. Recommended implementation of arterial spin-labeled perfusion mri for clinical applications: A consensus of the ismrm perfusion study group and the european consortium for asl in dementia. Magn Reson Med, 000:000–000, 2014.
6. Zun, Z., Lebel, R. M., Shankaranarayanan, A. & Zaharchuk, G. What Is the Ideal Labeling Duration for Pseudocontinuous Arterial Spin Labeling? 22, 66506 (2014).
7. Johnston, M., Lu, K., Maldjian, J. & Jung, Y. Multi-TI Arterial Spin Labeling MRI with Variable TR and Bolus Duration for Cerebral Blood Flow and Arterial Transit Time Mapping. IEEE Trans. Med. Imaging 0062, 1–1 (2015).