David L Thomas1,2, Fabio Nery3, Isky Gordon3, Chris A Clark3, Sebastien Ourselin2, Xavier Golay1, David Atkinson4, and Enrico De Vita1,5
1Department of Brain Repair and Rehabilitation, UCL Institute of Neurology, London, United Kingdom, 2Translational Imaging Group, UCL, London, United Kingdom, 3Developmental Imaging and Biophysics Section, UCL Institute of Child Health, London, United Kingdom, 4Centre for Medical Imaging, UCL, London, United Kingdom, 5Lysholm Department of Neuroradiology, National Hospital for Neurology and Neurosurgery, London, United Kingdom
Synopsis
Multi-shot
3D acquisition schemes offer an efficient method to obtain ASL data with good
SNR and spatial resolution. However, multi-shot techniques are susceptible to
motion-induced artefacts that can severely degrade image quality. In this work,
we investigate the use of the autofocus algorithm to correct k-space phase
inconsistencies caused by inter-shot motion, and demonstrate its effectiveness at
improving image sharpness and removing artefacts in motion-corrupted 3-shot 3D
GRASE data. As such, autofocus offers a retrospective approach to improve the quality
of multi-shot ASL data, with the associated improvements in CBF quantification
accuracy and reproducibility.Purpose
To use an automatic iterative
optimisation reconstruction approach (autofocus) to reduce motion artefacts in
multi-shot 3D GRASE arterial spin labelling data by improving between-segment k-space phase consistency.
Introduction
Arterial spin labelling
(ASL) is a non-invasive MRI technique for measuring tissue perfusion
1,2. For
applications in the brain, fast 3D imaging methods, such as 3D GRASE or 3D
stack-of-spirals, have been recommended in order to minimise acquisition time
and enable repeat measurements for averaging and SNR improvement
3. To
achieve acceptable brain coverage, 3D techniques can acquire the entire 3D
k-space in a single shot
4, but this typically leads to a poor point spread
function and signal blurring along the second phase-encoding (PE) dimension. More
typically, 3D data sets are acquired using a multi-shot approach, with the echo
train duration of each individual acquisition being limited to 300ms or less
3. While this improves the intrinsic resolution of the images, it introduces
motion-sensitivity because the different shots are acquired several seconds
apart. Subject movement between shots leads to ghosting and loss of resolution
in the images. Autofocus
5,6 is a retrospective motion-correction approach whereby trial phase corrections
are iteratively performed on raw k-space data and then evaluated according to a
particular cost function (e.g. an
image quality metric), typically after reconstruction. Here, we investigate using
autofocus to reduce motion artefacts and improve image quality in multi-shot 3D
GRASE ASL images.
Methods
MR acquisition: Data
from a healthy volunteer were acquired on a 3T Tim Trio (Siemens Healthcare,
Erlangen). Background suppressed FAIR Q2TIPS 3D GRASE ASL volumes were acquired
with the following parameters: TE=26.84ms; TI1/TI2=800/2000ms; spatial resolution=3.1x3.1x5mm3;
acquisition matrix 128x104x12, with 5/8 partial Fourier applied the second PE
and segmentation applied in the primary PE direction (3 shots), resulting in an
echo train duration of 215ms; 140° refocusing pulse flip angle; TR=4s.
Data were acquired in batches of 4 repeats: during the first batch, the subject
was asked to remain as still as possible; during the second batch, the subject
was asked to make translational head movements during the first and third
repeat. Raw k-space data were exported and processed in Matlab (The MathWorks, Inc., Natick, MA,
United States).
Autofocus algorithm: It is known from the Fourier shift
theorem that translations in the image domain correspond to phase ramps in the
frequency domain. Therefore, the autofocus algorithm applies trial phase ramps
to the acquired raw data to compensate for rigid body translations that have occurred
between the acquisitions of the different k-space segments. After each
shift and following reconstruction, the image quality of the resulting image is
quantified using a quality metric (image entropy). Minimizing the entropy of an
image reduces blurring and removes ghosts from otherwise reduced signal
intensity regions, and this forms the basic principle of the autofocus method5,6.
To avoid a lengthy brute-force optimization, a basic iterative algorithm was
implemented whereby the algorithm starts applying phase shifts in a coarse
grid, which is iteratively refined (Figure 1). This allows the algorithm to
start with a large range of possible phase ramps and then iterate down to very
small phase increments to correct motion at the sub-pixel level.
Results
The autofocus algorithm was able to converge to an optimal minimum-entropy solution for the 3-shot 3D GRASE data sets. Figure 2 shows the effect
of applying the autofocus algorithm to one of the ASL control acquisitions with
intentional subject movement. Two main improvements are clear: (i) blurring is
reduced, which is apparent as a visible increase in image ‘sharpness’ when
comparing Fig 2a and 2b, and (ii) image ghosting is also reduced, as demonstrated
by the difference (pre-post correction) images in Fig 2c.
Discussion
Autofocus is a
post-processing algorithm that can be applied retrospectively to motion-corrupted
MRI k-space data. For many applications, the computational overheads associated
with the approach make it impractical. However, for multi-shot 3D data, because
intra-shot motion can be neglected, the limited number of shots makes the
method much more feasible. In the data presented here, a three shot
acquisition has two degrees of freedom (as the first shot is used as the
reference position), and therefore the iterative algorithm converges quickly on
an optimal solution. In the work presented here as a proof of principle, motion was restricted to
in-plane translation, but extension of the method to 3D
translations and rotations is straightforward. Future work will investigate the
effect of autofocus correction on CBF quantification and reproducibility.
Conclusions
We have demonstrated the
use of autofocus to improve image quality in multi-shot 3D GRASE ASL
acquisitions.
Acknowledgements
DLT is supported by the UCL
Leonard Wolfson Experimental Neurology Centre (PR/ylr/18575). We would also like to acknowledge the EPSRC
(EP/H046410/1) and the UCLH/UCL CBRC Strategic Investment Award (Ref. 168).References
1. Detre et al, Magn Reson Med 23; 37-45 (1992)
2. Barker et al (eds), Clinical Perfusion MRI: Techniques and
Applications. Cambridge University Press (2013)
3. Alsop et al, Magn Reson Med 73; 102–16 (2015)
4. Guenther et al, Magn Reson Med 54; 491-498 (2005)
5. Atkinson et al, IEEE Trans Med
Imaging 16; 903-910 (1997)
6. Atkinson et al, Magn Reson Med 41; 163-170 (1999)