Joshua S. Greer1,2, Xinzeng Wang2, Ivan Pedrosa2,3, and Ananth J. Madhuranthakam2,3
1Bioengineering, University of Texas at Dallas, Richardson, TX, United States, 2Radiology, UT Southwestern Medical Center, Dallas, TX, United States, 3Advanced Imaging Research Center, UT Southwestern Medical Center, Dallas, TX, United States
Synopsis
Cardiac triggering is regularly used in arterial spin
labeled (ASL) pulmonary perfusion imaging to minimize pulsatile flow effects. A
cardiac-triggered pseudo-continuous ASL (pCASL) sequence was previously implemented
to improve the SNR of lung perfusion, but was found to be sensitive to
variations in heart rate due to the prolonged TR, causing the data acquisition
to occasionally occur during systole. The purpose of this study was to investigate
methods to reduce this cardiac cycle sensitivity in pulmonary perfusion images
using pCASL, including cardiac triggering the acquisition and alternative
readouts such as a projection acquisition.Introduction
Cardiac
triggering is essential for accurately quantifying pulmonary perfusion using
arterial spin labeling, and is regularly used in pulsed-ASL methods in the
lungs at 1.5T [1,2]. Compared to pulsed ASL, pseudo-continuous ASL (pCASL) offers
higher signal to noise ratio (SNR) [3]. A pCASL approach was recently applied to
measure pulmonary perfusion [4]. To minimize the sensitivity of pulmonary
perfusion to cardiac motion, the beginning of the pCASL pulse sequence was
cardiac triggered. However, the prolonged repetition times of pCASL (~5-6 sec), due to long labeling duration and long post-label delay, renders the acquisition
sensitive to variations in the heart rate (fig. 1). Thus, the purpose of this
work was to investigate methods to reduce cardiac cycle induced signal
variations in pulmonary perfusion using pCASL.
Methods
Pulmonary
perfusion weighted images are typically acquired with a single shot turbo spin
echo (SShTSE) due to its reduced sensitivity to B
0 inhomogeneities. However,
it has been previously shown that the SShTSE signal in the lungs varies
throughout the cardiac cycle [5], and that these variations are significant
enough to mimic perfusion-weighted images [6]. Similar signal variations were
observed in pulmonary perfusion using pCASL, even when the sequence was cardiac
triggered, due to slight variations in the cardiac cycle between the control
and label acquisitions (fig. 1). Various approaches were explored to overcome
this challenge including: 1) pCASL labeling of the right pulmonary artery close
to the lungs, 2) cardiac triggering the acquisition after the pCASL labeling, rather
than from the beginning of the sequence (fig. 2), and 3) using flow insensitive acquisitions such as T1 fast
field echo (T1-FFE) with a projection acquisition.
Normal
volunteers were scanned on a 3T Ingenia scanner (Philips Healthcare, Best, The
Netherlands) with IRB approval and written informed consent. To demonstrate the
signal differences that may arise from improper cardiac triggering, systolic
and diastolic proton density images (ECG triggered using 100ms and 400ms
delays, respectively) were acquired using a SShTSE without labeling. Similar
images were also acquired using a T1-FFE with a projection acquisition without
labeling. We hypothesized that the projection acquisition will be insensitive
to cardiac signal variations due to the acquisition of the center of k-space
throughout the cardiac cycle.
Subsequently, pCASL perfusion-weighted images of
the right lung were acquired using a SShTSE and a T1-FFE with a projection
acquisition. Sagittal pCASL labeling was applied perpendicularly to the right
pulmonary artery for 2.5 seconds, followed by a 500ms post-label delay to allow
labeled blood to perfuse the lung tissue. To ensure that the data acquisition with
SShTSE occurred during diastole, the cardiac trigger was placed after the
post-label delay (fig. 2). This approach minimized signal variations due to
pulsatile flow, even in cases with significantly varied heart rate between
control and label acquisitions.
Results
Figure 3
shows the difference between images acquired during systolic and diastolic
phase using a proton density weighted SShTSE (fig. 3b) and T1-FFE with Projection
acquisition (fig. 3c). Even without labeling, SShTSE generates a
“perfusion-weighted” image due to its sensitivity to cardiac cycle induced
signal variations, while the T1-FFE with projection acquisition is insensitive
to such cardiac signal variations. Figure 4 shows the pulmonary
perfusion-weighted images acquired with sagittal pCASL labeling of the right
pulmonary artery using cardiac triggered SShTSE acquisition (fig. 4b) and
T1-FFE with projection acquisition (without triggering) (fig. 4c) showing true
perfusion-weighted signal.
Discussion
The
sensitivity of SShTSE acquisition to cardiac cycle induced signal variations
has been demonstrated previously, to generate “perfusion-weighted” images [6]. However,
the variations in signal intensities depending upon the cardiac phase rendered
these images inconsistent and challenging to quantify. In this work, we
demonstrated a true perfusion-weighted image using pCASL labeling by cardiac
triggering the SShTSE data acquisition. This ensured that the only signal
variations between control and label images are from spin labeling and not due
to cardiac cycle. However, this poses a new challenge of variable post-label
delay between the control and label acquisitions that may introduce errors in
quantification. As an alternate, our approach with T1-FFE with projection
acquisition is insensitive to cardiac induced signal variations (fig. 3c) and
thus may be a viable acquisition strategy without cardiac triggering (fig. 4c).
Future work will quantify the perfusion maps with both acquisitions and compare
the perfusion maps against pulsed-ASL approaches, such as FAIR [7].
Acknowledgements
No acknowledgement found.References
[1] Bolar, D. S., et al. "Quantification of regional pulmonary blood flow using
ASL-FAIRER." Magnetic resonance in medicine 55.6 (2006): 1308-1317.
[2] Henderson,
A. Cortney, et al. "Characterizing pulmonary blood flow distribution
measured using arterial spin labeling." NMR in biomedicine 22.10 (2009): 1025-1035.
[3] Alsop,
David C., et al. "Recommended implementation of arterial spin-labeled
perfusion MRI for clinical applications: A consensus of the ISMRM perfusion
study group and the European consortium for ASL in dementia." Magnetic Resonance in Medicine 73.1 (2015): 102-116.
[4] Joshua
S. Greer, Yue Zhang, Ivan Pedrosa, and Ananth J. Madhuranthakam. “Non-Contrast Pulmonary
Perfusion using pseudo-Continuous Arterial Spin Labeling of the Inferior Vena
Cava.” Proceedings of the
International Society for Magnetic Resonance in Medicine, 23rd Scientific
Meeting, Toronto, Canada. (2015).
[5] Knight-Scott,
Jack, et al. "Temporal dynamics of blood flow effects in half-Fourier
fast spin echo 1H magnetic resonance imaging of the human lungs."Journal
of Magnetic Resonance Imaging 14.4
(2001): 411-418.
[6] Tadamura, Eiji, and Hiroto Hatabu. "Assessment of
pulmonary perfusion using a subtracted HASTE image between diastole and
systole." European
journal of radiology 37.3
(2001): 179-183.
[7] Bolar,
D. S., et al. "Quantification of regional pulmonary blood flow using ASL-FAIRER." Magnetic resonance in medicine 55.6 (2006): 1308-1317.