Kai Jiang1, Hui Tang1, Prassana K. Mishra2, Slobodan I. Macura2, and Lilach O. Lerman1
1Division of Nephrology and Hypertension, Mayo Clinic, Rochester, MN, United States, 2Department of Biochemistry and Molecular Biology, Mayo Clinic, Rochester, MN, United States
Synopsis
In this study, a saturation recovery
fast spin echo method was developed for fast T1 mapping of mouse
kidney at a temporal resolution of ~3 min. The validity of this method was
first demonstrated in vitro on a
phantom with different concentrations of MnCl2 by comparing to the
conventional spin echo T1 mapping method and a previously validated
saturation recovery Look-Locker (SRLL) method and then in vivo on mouse kidneys at both pre- and post-MnCl2
infusion by comparing to SRLL. Introduction
Assessment of renal
functionality using dynamic contrast-enhanced MRI (DCE-MRI) requires accurate T
1
mapping of kidneys at high temporal resolution. The inversion recovery
Look-Locker T
1 mapping method provides robust and relatively fast T
1
measurement, but requires a repetition time (TR) of 5 times T
1 for
complete relaxation of longitudinal relaxation.
1 Recently, a
saturation recovery Look-Locker (SRLL) method was developed for fast
in vivo T
1 mapping of mouse
myocardium within 3 min.
2 However, this method suffers from low SNR
due to a small flip angle (~10°). In the current study, a saturation recovery
fast spin echo (SRFSE) method was developed for fast and accurate T
1
mapping of mouse kidneys.
Materials and Methods
SRFSE Protocol The SRFSE sequence
diagram is shown in Fig. 1. During each repetition, three nonselective
saturation pulses with spoil gradients in all three directions are applied
before each fast spin echo (FSE) acquisition. A total of 20 images were
acquired with delay time (TD) from 0 to 3.8 s. An echo train length of 32 was
used to balance image resolution and acquisition time. To guarantee high SNR in
the acquired images, a centric encoding scheme was implemented.
Validation
Studies All MRI studies were performed on a vertical 16.4 T
animal scanner (Bruker Biospin, Billerica, MA) equipped with a 38 mm inner
diameter birdcage coil. The SRFSE method
was first validated in vitro using a
multi-compartment phantom with MnCl2 solutions concentrated from 30
to 500 μM. The SRFSE images from a single slice were acquired using the
following parameters: FOV 3.0×3.0 cm2; matrix size 128×128; echo
spacing 4.42 ms; slice thickness 1 mm; number of averages 1. The standard
spin-echo (SE) and SRLL methods were also implemented for validation of SRFSE. In
SE, fourteen TR values ranging from 50 to 10000 ms were used to sample the data
on the entire longitudinal recovery curve. In SRLL, a total of 20 Look-Locker
images were acquired with a sampling interval of 190 ms and an average of 1. The
central 64 lines were acquired in the phase encoding direction. The proton
density image was acquired with a TR of 3 s.
For in vivo validation, three 3-month old C57BL/6J mice were used. T1
maps of a short-axis slice were acquired using SRFSE and SRLL at baseline and
after the infusion of 0.1 mL 63mM MnCl2 solution through the tail-vein.
A FOV of 2.56×2.56 cm2 was used. In SRLL, the number of averages was
prescribed at 2 to increase SNR. All other imaging parameters were the same as
in the phantom study. T1 values of the cortex (CO), outer (OM) and
inner (IM) medulla were quantified.
Results and Discussion
Three representative
SRFSE images of the phantom acquired at 400, 1000, and 3800 ms are shown in
Fig. 2a-c. The continuous increase of the image intensity reflects the recovery
of longitudinal magnetization. The total acquisition time was 2.75, 7.25
(including the proton-density image), and 80.9 min for SRFSE, SRLL and SE,
respectively. The fitted T1 map is shown in Fig 2d, which showed a
good agreement with SE and SRLL (Fig. 2e). A linear regression analysis of the
T1 relaxation rate (R1) and Mn2+ concentration
showed that the relaxivity of Mn2+ in water was 5.83 s-1mM-1
at 16.4 T (Fig. 2e).
The acquisition time for
the in vivo study was 2.75 and 14.5 (including
the proton-density image) min for SRFSE and SRLL, respectively. Shown in Fig. 3
are the representative T1 maps by SRLL (a&c) and SRFSE (b&d)
at baseline (a&b) and post-Mn2+ infusion (c&d). Quantitative comparison showed no difference
in the measured T1 values of CO, OM and IM by SRLL and SRFSE either
at baseline or after Mn2+ infusion (Fig. 3e). Notably, the T1
value of OM was the same as CO at baseline but showed a slightly greater fall
(83.4% for CO and 87.5% for OM) after Mn2+ infusion (Fig. 3c-e), suggesting
more uptake of Mn2+ in the OM than CO.
Conclusion
A saturation recovery fast spin echo method was developed for
fast T
1 mapping of mouse kidneys. Validation studies performed in
MnCl
2 phantom and mouse kidneys at baseline and post-contrast showed
a good agreement with previously validated T
1 mapping methods.
Acknowledgements
None.References
1. Messroghli DR, Radjenovic A, Kozerke S, Higgins DM, Sivananthan MU, Ridgway JP. Modified Look-Locker inversion recovery
(MOLLI) for high-resolution T1 mapping of the heart. Magn. Reson. Med.
2004;52(1):141–6.
2. Jiang K, Li W, Li W, Jiao S, Castel L, Van Wagoner DR, Yu X. Rapid multislice T1
mapping of mouse myocardium: Application to quantification of manganese uptake
in α-Dystrobrevin knockout mice. Magn. Reson. Med. 2015;74(5):1370–9.