Simultaneous Quantification of Blood Vessel Caliber and Oxygenation via Multi-Voxel Joint Utilization of Magnitude and Phase (MV-JUMP)
Patrick McDaniel1, Berkin Bilgic2, Audrey Fan3, Jeffrey Stout4, and Elfar Adalsteinsson1,4

1Electrical Engineering and Computer Science, MIT, Cambridge, MA, United States, 2Athinoula A Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3Richard M Lucas Service Center for Imaging, Stanford University, Stanford, CA, United States, 4Health Sciences and Technology, Harvard/MIT, Cambridge, MA, United States

Synopsis

Accurate, quantitative determination of blood vessel oxygenation and caliber from GRE data would be useful for assessing local perfusion and oxygen consumption. However, partial-volume effects confound both measurements. Vessel edges are blurred and discretized, hampering assessment of vessel size. Oxygenation can be measured from GRE phase, but partial-volume effects with other tissues often contaminate these measurements. In this work, we present a novel approach that simultaneously obtains accurate measurements of vessel caliber and oxygenation state for vessels nearly parallel to B0. This approach is demonstrated in numerical and in vivo experiments.

Introduction

Veins affect voxel magnitude and phase signals in GRE acquisitions by virtue of the high magnetic susceptibility of deoxygenated blood. Estimation of vessel caliber and oxygenation from susceptibility would - along with blood velocity measurements - enable quantification of local CMRO2 [1], but partial-volume effects confound the estimation of both. Previously, we presented a method for accurately measuring blood vessel oxygenation in the presence of partial-volume effects called Joint Utilization of Magnitude and Phase (JUMP). In this work, we present a novel approach that accurately measures both blood vessel caliber and oxygenation state and is robust to partial-volume effects, which we call Multi-Voxel Joint Utilization of Magnitude and Phase (MV-JUMP).

Theory

We assume that blood vessel segments are straight and oriented nearly parallel to the main field $$$B_0$$$. The blood vessel voxel magnitude and phase signals in a GRE acquisition both depend on the vessel oxygenation saturation ($$$Y_v$$$) and the fraction of the voxel comprised of blood ($$$\alpha$$$). This relationship can be described by a two-compartment model [2]. If we impose that all voxels in a vessel have the same $$$Y_v$$$, then we can estimate $$$Y_v$$$ and each voxel's partial-volume fraction by solving the following minimization problem: $$\min_{\{\hat{\alpha}_k\},\hat{Y}_v}\left[\sum_{k}\sum_{i}\max(\varphi_{v,ik},0.1)\cdot{\left\|\hat{M}_{v,i}(\hat{\alpha}_{k},\hat{Y}_{v})\cdot e^{j\hat{\varphi}_{v,i}(\hat{\alpha}_{k},\hat{Y}_{v})}-M_{v,ik}\cdot e^{j\varphi_{v,ik}}\right\|}^{2}\right]$$Here, $$$\hat{Y}_{v}$$$ is the estimated vessel oxygen saturation, $$$\{\hat{\alpha}_{k}\}$$$ are the estimated voxel partial-volume fractions,and $$$M_{v,ik}$$$ and $$$\varphi_{v,ik}$$$ are the measured voxel signal magnitudes and phases. $$$\hat{M}_{v,i}(\hat{\alpha}_{k},\hat{Y}_{v})$$$ and $$$\hat{\varphi}_{v,i}(\hat{\alpha}_{k},\hat{Y}_{v})$$$ describe the forward signal model in [2], and further account for the $$$Y_v$$$-dependence of blood $$${R}_{2}^{*}$$$. The sub-indices $$$k$$$ and $$$i$$$ indicate voxel and echo time ($$$TE$$$), respectively. The thresholding term limits the contribution of low-SNR voxels to the $$$\hat{Y}_{v}$$$ solution. As a consequence of the sinc-shaped voxel function implicit in k-space windowing, $$$\hat{\alpha}_k$$$ is constrained to be between -0.1628 and +1.39. Venous cross-sectional area ($$$\hat{A}_{cs}$$$) is calculated by summing $$$\hat{\alpha}_{k}$$$ from all voxels comprising an axial section of the vessel. The cross-sectional area is then: $$\hat{A}_{cs}=\hat{\alpha}_\Sigma \cdot \delta_x \cdot \delta_y \cdot \cos(\theta)$$ $$$\hat{\alpha}_{\Sigma}$$$ is the axial section $$$\hat{\alpha}_{k}$$$ sum, $$$\delta_x$$$ and $$$\delta_y$$$ are the nominal voxel dimensions along the x- and y-axes, and $$$\theta$$$ is the angle made by the vessel with $$$B_0$$$. Averaging $$$\hat{A}_{cs}$$$ from all axial sections then gives an average vessel cross-sectional area.

Methods

Numerical vessels (Figure 1) and in vivo vessels (Figure 2) were analyzed. Dual-echo ($$$TE=8.1ms, 20.3ms$$$) GRE data from [2] were used to analyze four in vivo vessels from 2 subjects. All phase images were unwrapped with the robustunwrap tool [3] and processed with SHARP to remove background phase [4]. Vessel ROIs were manually identified, and included all voxels that visibly coincided with the vessel (Figures 1-2). Vessels were then analyzed with MV-JUMP. The MV-JUMP minimization problem was solved using the MATLAB fmincon tool (initial guesses for all runs: $$$\hat{Y}_v=0.6$$$; $$$\hat{\alpha}_k=0.6$$$ if $$$\varphi_{v,k}>0$$$, $$$\hat{\alpha}_k=-0.1$$$ if $$$\varphi_{v,k}<0$$$ at $$$TE=20.3ms$$$).

Results

Figure 3 shows the measured $$$\hat{Y}_v$$$ and average $$$\hat{A}_{cs}$$$ for numerical vessels. $$$\hat{Y}_v$$$ was always within 10% of the ground truth vessel $$${Y}_v$$$ for voxel sizes under 2mm isotropic. $$$\hat{A}_{cs}$$$ deviated from ground truth by under 4% at $$$Y_{v}=0.8$$$; by under 10% at $$$Y_{v}=0.7$$$; and by under 19% at $$$Y_{v}=0.6$$$ for voxel sizes under 2mm isotropic. Figure 4 shows the measured vessel $$$\hat{Y}_v$$$ and $$$\hat{A}_{cs}$$$ for two vessels in each subject. Vessels 3 and 4 could not be properly identified in Subject 2 at 0.6mm isotropic due to reconstruction artifacts. Measured mean $$$\hat{Y}_v$$$ across acquisition resolutions varied between 0.575 and 0.701 for the 4 vessels, and all acquisitions for a vessel gave $$$\hat{Y}_v$$$ within 9% of the vessel mean. Measured mean $$$\hat{A}_{cs}$$$ across resolution varied between 0.545mm2 and 2.2mm2. All acquisitions for a given vessel gave $$$\hat{A}_{cs}$$$ within 16% of the vessel mean.

Discussion

MV-JUMP accurately measured $$$\hat{Y}_v$$$ and $$$\hat{A}_{cs}$$$ in vessels with $$$Y_v$$$ between 0.6 and 0.8 and with a tilt angle of 20°. In in vivo experiments, MV-JUMP provided average measurements of $$$\hat{Y}_v$$$ that fell within or near this range. Also, variations in $$$\hat{Y}_v$$$ and $$$\hat{A}_{cs}$$$ for in vivo vessels at different acquisition resolutions were similar to the variations seen in simulation. This suggests that MV-JUMP accurately measured both oxygenation and caliber from the vessels analyzed in this study. This approach is currently limited to vessels oriented close to $$$B_0$$$. Increasing extravascular fields and the “magic angle” problem are ignored in the physical model used here, but become important at large tilt angles. The MV-JUMP inverse problem is capable of using a model that accounts for these concerns through modification of the form of $$$\hat{M}_{v,ik}$$$ and $$$\hat{\varphi}_{v,ik}$$$, and future work will look to develop such a model.

Acknowledgements

This work was supported by NIH NIBIB: R01EB017337, U01HD087211, P41-EB015896, 1U01MH093765, R00EB012107, R24MH106096, T32 EB001680 and the Stanford Neurosciences Institute Interdisciplinary Scholar Award.

References

[1] Rodgers ZB, Jain V, Englund EK, Langham MC, Wehrli FW. High temporal resolution MRI quantification of global cerebral metabolic rate of oxygen consumption in response to apneic challenge. J. Cereb. Blood Flow Metab. 2013;33:1514–22. doi: 10.1038/jcbfm.2013.110.

[2] McDaniel P, Fan AP, Bilgic B, Stout J, Adalsteinsson E. Improved Accuracy in Susceptibility-based OEF Measurements by Mitigation of Partial-Volume Effects via Combined Magnitude and Phase Reconstruction. In: Proc. ISMRM, Toronto, Canada, 2015. pg 3330.

[3] Cusack R, Papadakis N. New Robust 3-D Phase Unwrapping Algorithms: Application to Magnetic Field Mapping and Undistorting Echoplanar Images. Neuroimage 2002;16:754–764. doi: 10.1006/nimg.2002.1092.

[4] Schweser F, Deistung A, Lehr BW, Reichenbach JR. Quantitative imaging of intrinsic magnetic tissue properties using MRI signal phase: An approach to in vivo brain iron metabolism? Neuroimage 2011;54:2789–2807. doi: 10.1016/j.neuroimage.2010.10.070.

Figures

Vessels were created for analysis with: $$$A_{cs}=1.152mm^2$$$; $$$Y_{v}=0.6, 0.7, 0.8$$$; and $$$\theta=20\deg$$$. GRE acquisitions were simulated with: $$$TE=8.1ms, 20.3ms$$$; Voxel Size=0.5 through 5mm isotropic (18 sizes); $$$B_0=2.89T$$$. Gaussian noise was added so that SNR=20 at 0.6mm isotropic voxel size to match the observed SNR in the In Vivo data.

Four vessels identified from two subjects are shown at 0.7mm and 1.2mm isotropic resolution in both GRE phase and GRE magnitude images. Vessel masks were created at all resolutions, and encompassed the entire cross-section of the vessels as was visible in the GRE magnitude image

(a) Oxygen saturation ($$$\hat{Y}_v$$$) in 3 vessels measured using MV-JUMP. $$$\hat{Y}_v$$$ measurements were always within 8.5% of ground truth for voxel dimensions under 2mm. (b) Mean cross-sectional area ($$$\hat{A}_{cs}$$$) of 3 vessels measured with MV-JUMP. $$$\hat{A}_{cs}$$$ measurements were always within 19% of ground truth for voxel dimensions under 2mm.

(a) $$$\hat{Y}_v$$$ and (b) $$$\hat{A}_{cs}$$$ in all vessels as measured using MV-JUMP. Vessels 3 and 4 could not be identified in the 0.6mm isotropic acquisition. $$$\hat{Y}_v$$$ deviated less than 9% from the mean across all acquisition resolutions, and $$$\hat{A}_{cs}$$$ deviated less than 16% from the mean across all acquisition resolutions.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2844