Patrick McDaniel1, Berkin Bilgic2, Audrey Fan3, Jeffrey Stout4, and Elfar Adalsteinsson1,4
1Electrical Engineering and Computer Science, MIT, Cambridge, MA, United States, 2Athinoula A Martinos Center for Biomedical Imaging, Massachusetts General Hospital, Charlestown, MA, United States, 3Richard M Lucas Service Center for Imaging, Stanford University, Stanford, CA, United States, 4Health Sciences and Technology, Harvard/MIT, Cambridge, MA, United States
Synopsis
Accurate,
quantitative determination of blood vessel oxygenation and caliber from GRE
data would be useful for assessing local
perfusion and oxygen consumption. However,
partial-volume effects confound both measurements. Vessel edges are blurred and
discretized, hampering assessment of vessel size. Oxygenation can be measured from GRE phase,
but partial-volume effects with other tissues often contaminate these
measurements. In this work, we present a novel approach that simultaneously
obtains accurate measurements of vessel caliber and oxygenation state for vessels
nearly parallel to B0. This approach is demonstrated in numerical
and in vivo experiments.Introduction
Veins affect voxel magnitude and phase signals in GRE
acquisitions by virtue of the high magnetic susceptibility of deoxygenated
blood. Estimation of vessel caliber and oxygenation from susceptibility would - along with blood velocity measurements - enable quantification of local CMRO
2 [1], but partial-volume effects confound
the estimation of both. Previously, we presented a method for accurately
measuring blood vessel oxygenation in the presence of partial-volume effects
called Joint Utilization of Magnitude and Phase (JUMP). In this work, we
present a novel approach that accurately measures both blood vessel caliber and
oxygenation state and is robust to partial-volume effects, which we call Multi-Voxel
Joint Utilization of Magnitude and Phase (MV-JUMP).
Theory
We
assume that blood vessel segments are straight and oriented nearly parallel to
the main field $$$B_0$$$. The blood
vessel voxel magnitude and phase signals in a GRE acquisition both depend on
the vessel oxygenation saturation ($$$Y_v$$$) and the fraction of the
voxel comprised of blood ($$$\alpha$$$). This relationship can
be described by a two-compartment model [2]. If we impose that all
voxels in a vessel have the same $$$Y_v$$$, then we can estimate $$$Y_v$$$ and each voxel's partial-volume fraction by
solving the following minimization problem: $$\min_{\{\hat{\alpha}_k\},\hat{Y}_v}\left[\sum_{k}\sum_{i}\max(\varphi_{v,ik},0.1)\cdot{\left\|\hat{M}_{v,i}(\hat{\alpha}_{k},\hat{Y}_{v})\cdot e^{j\hat{\varphi}_{v,i}(\hat{\alpha}_{k},\hat{Y}_{v})}-M_{v,ik}\cdot e^{j\varphi_{v,ik}}\right\|}^{2}\right]$$Here, $$$\hat{Y}_{v}$$$ is the estimated vessel oxygen saturation, $$$\{\hat{\alpha}_{k}\}$$$ are the estimated voxel partial-volume fractions,and $$$M_{v,ik}$$$ and $$$\varphi_{v,ik}$$$ are the measured voxel signal magnitudes and
phases. $$$\hat{M}_{v,i}(\hat{\alpha}_{k},\hat{Y}_{v})$$$ and $$$\hat{\varphi}_{v,i}(\hat{\alpha}_{k},\hat{Y}_{v})$$$ describe the forward signal model in
[2], and
further account for the $$$Y_v$$$-dependence of blood $$${R}_{2}^{*}$$$. The sub-indices $$$k$$$ and $$$i$$$ indicate voxel and echo time ($$$TE$$$),
respectively. The thresholding term limits the contribution of low-SNR voxels
to the $$$\hat{Y}_{v}$$$ solution. As a consequence of the sinc-shaped
voxel function implicit in k-space windowing, $$$\hat{\alpha}_k$$$ is constrained to be between -0.1628 and
+1.39. Venous cross-sectional area ($$$\hat{A}_{cs}$$$) is calculated by summing $$$\hat{\alpha}_{k}$$$ from all voxels comprising an axial section of
the vessel. The cross-sectional area is then: $$\hat{A}_{cs}=\hat{\alpha}_\Sigma \cdot \delta_x \cdot \delta_y \cdot \cos(\theta)$$ $$$\hat{\alpha}_{\Sigma}$$$ is the axial section $$$\hat{\alpha}_{k}$$$ sum, $$$\delta_x$$$ and $$$\delta_y$$$ are the nominal voxel dimensions along the x-
and y-axes, and $$$\theta$$$ is the angle made by the vessel with $$$B_0$$$. Averaging $$$\hat{A}_{cs}$$$ from all axial sections then gives an average vessel cross-sectional area.
Methods
Numerical vessels (Figure 1) and in vivo vessels (Figure 2) were analyzed.
Dual-echo ($$$TE=8.1ms, 20.3ms$$$) GRE data from [2] were used to analyze four
in vivo vessels from 2 subjects.
All phase images were unwrapped with the
robustunwrap tool [3] and processed with SHARP to remove
background phase [4]. Vessel ROIs were manually identified, and included all voxels that visibly coincided with the
vessel (Figures 1-2). Vessels were then analyzed with MV-JUMP. The MV-JUMP minimization problem was solved using the MATLAB fmincon tool (initial
guesses for all runs: $$$\hat{Y}_v=0.6$$$; $$$\hat{\alpha}_k=0.6$$$ if $$$\varphi_{v,k}>0$$$, $$$\hat{\alpha}_k=-0.1$$$ if $$$\varphi_{v,k}<0$$$ at $$$TE=20.3ms$$$).
Results
Figure 3 shows the measured $$$\hat{Y}_v$$$ and average $$$\hat{A}_{cs}$$$ for numerical vessels. $$$\hat{Y}_v$$$ was always within 10% of the ground truth
vessel $$${Y}_v$$$ for voxel sizes under 2mm isotropic. $$$\hat{A}_{cs}$$$ deviated from ground truth by under 4% at $$$Y_{v}=0.8$$$; by under 10% at $$$Y_{v}=0.7$$$; and by under 19%
at $$$Y_{v}=0.6$$$ for voxel sizes
under 2mm isotropic. Figure 4 shows the measured vessel $$$\hat{Y}_v$$$ and $$$\hat{A}_{cs}$$$ for two vessels in each subject. Vessels 3 and
4 could not be properly identified in Subject 2 at 0.6mm isotropic due to
reconstruction artifacts. Measured mean $$$\hat{Y}_v$$$ across acquisition resolutions varied between
0.575 and 0.701 for the 4 vessels, and all acquisitions for a vessel gave $$$\hat{Y}_v$$$ within 9% of the vessel mean. Measured mean $$$\hat{A}_{cs}$$$ across resolution varied between 0.545mm
2
and 2.2mm
2. All acquisitions for a given vessel gave $$$\hat{A}_{cs}$$$ within 16% of the vessel mean.
Discussion
MV-JUMP accurately measured $$$\hat{Y}_v$$$ and $$$\hat{A}_{cs}$$$ in vessels with $$$Y_v$$$ between 0.6 and 0.8 and with a tilt angle of 20°.
In in vivo experiments, MV-JUMP
provided average measurements of $$$\hat{Y}_v$$$ that fell within or near this range. Also,
variations in $$$\hat{Y}_v$$$ and $$$\hat{A}_{cs}$$$ for in vivo vessels at different acquisition resolutions were similar to the variations seen in simulation. This suggests that MV-JUMP accurately
measured both oxygenation and caliber from the vessels analyzed in this study.
This approach is currently limited to
vessels oriented close to $$$B_0$$$. Increasing extravascular
fields and the “magic angle” problem are ignored in the physical model used here, but
become important at large tilt angles. The MV-JUMP inverse problem is capable of using a model
that accounts for these concerns through modification of the form of $$$\hat{M}_{v,ik}$$$ and $$$\hat{\varphi}_{v,ik}$$$, and future work will look to
develop such a model.
Acknowledgements
This work was supported by NIH NIBIB:
R01EB017337, U01HD087211, P41-EB015896, 1U01MH093765, R00EB012107,
R24MH106096, T32 EB001680 and the Stanford Neurosciences Institute
Interdisciplinary Scholar Award.References
[1] Rodgers ZB, Jain V, Englund EK, Langham MC, Wehrli FW. High temporal resolution MRI quantification of global cerebral metabolic rate of oxygen consumption in response to apneic challenge. J. Cereb. Blood Flow Metab. 2013;33:1514–22. doi: 10.1038/jcbfm.2013.110.
[2] McDaniel
P, Fan AP, Bilgic B, Stout J, Adalsteinsson E. Improved Accuracy in Susceptibility-based OEF Measurements by
Mitigation of Partial-Volume Effects via Combined Magnitude and Phase
Reconstruction. In: Proc. ISMRM, Toronto, Canada, 2015. pg 3330.
[3] Cusack R, Papadakis N. New Robust 3-D Phase Unwrapping
Algorithms: Application to Magnetic Field Mapping and Undistorting Echoplanar
Images. Neuroimage 2002;16:754–764. doi: 10.1006/nimg.2002.1092.
[4] Schweser F, Deistung A, Lehr BW, Reichenbach JR. Quantitative
imaging of intrinsic magnetic tissue properties using MRI signal phase: An
approach to in vivo brain iron metabolism? Neuroimage 2011;54:2789–2807. doi:
10.1016/j.neuroimage.2010.10.070.