Courtney K Morrison1, Leah C Henze Bancroft1, Kang Wang2, James H Holmes2, Frank R Korosec1,3, and Roberta M Strigel1,3
1Medical Physics, University of Wisconsin-Madison, Madison, WI, United States, 2Global MR Applications and Workflow, GE Healthcare, Madison, WI, United States, 3Radiology, University of Wisconsin-Madison, Madison, WI, United States
Synopsis
Fat suppression can be
achieved in a variety of ways, including using water-only excitation or using a
Dixon method. Each of these methods exhibits strengths and limitations. In this
work, we evaluated the characteristics of water excitation compared to a 2-point
Dixon fat suppression method used in high spatiotemporal resolution DCE breast
MRI.Introduction
Several
techniques have been proposed to improve the temporal resolution of dynamic
contrast-enhance (DCE) breast MRI while maintaining the clinically-necessary
high spatial resolution.
1-4 One such technique is DIfferential
Subsampling with Cartesian Ordering (DISCO) which most commonly uses a 2-point
Dixon method for removal of fat signal.
5 However, the echo spacing at
3.0 T required for the Dixon method limits the number of readout points that
can be acquired in each echo and thus limits the achievable spatial resolution.
Fat suppression can also be achieved using a spatial-spectral radiofrequency pulse
that excites only water. This technique allows for higher spatial resolution
but is more affected by B0 and B1 inhomogeneities. Given the potential for
trade-offs between these methods, in this work we investigated the characteristics
of these two types of fat suppression when used with DISCO at 3.0 T for DCE breast
MRI.
Methods
DISCO
uses a pseudorandom k-space sampling scheme that fully samples the center of
k-space but undersamples the periphery of k-space in each time frame. Images
are then reconstructed by view sharing peripheral k-space data from nearby
frames. This sampling scheme was incorporated into a spoiled gradient echo
sequence. Fat suppression was achieved in one of two ways: with a
spatial-spectral radiofrequency pulse (water excitation) or with 2-point Dixon
fat-water separation (Dixon).
6 Twenty-five patients undergoing
clinical breast MRI were recruited for this study. Patients were scanned on a
3.0 T scanner (Discovery MR 750w, GE Healthcare, Waukesha, WI) using a 16-channel
breast coil (Sentinelle, Invivo, Gainsville, FL). Clinical DCE MRI was acquired
after injection of 0.1 mmol/kg of gadobenate dimeglumine (MultiHance, Bracco
Inc., Milan, Italy). A single, fully-sampled, mask phase of both DISCO water
excitation and DISCO-Dixon were acquired immediately following the clinical DCE
MRI. The order of the water excitation and Dixon sequences was randomized. DISCO-Dixon
was acquired with the highest in-plane resolution achievable for the selected
imaging parameters (1.0 x 1.0 mm
2, limited by echo spacing) while
water excitation was acquired at a clinically-desired in-plane resolution (0.8 x
0.8 mm
2). Other imaging parameters were: FOV = 32 x 32 cm
2,
124 slices with thickness of 1.6 mm, flip angle = 10 degrees, ARC parallel
imaging factor of 3 x 1, bandwidth = ±83.33 kHz for water excitation and ±166.67 kHz
for DISCO-Dixon. The acquisition time for the fully-sampled mask was 1:17 for
water excitation and 1:19 for DISCO-Dixon. Both sets of DISCO images were
scored by three radiologists with 1, 6, and 22 years of experience for
diagnostic image quality (1-non-diagnostic, 3-adequate, 5-excellent diagnostic quality)
in the following categories: depiction of anatomy, degree of fat suppression,
uniformity of fat suppression, water signal shading, and overall diagnostic
confidence. Radiologists also ranked their preference for the images based on
overall image quality and the quality of fat suppression.
Results
A
representative patient case is shown in Figure 1. In 24 of 25 cases, both DISCO
image sets were scored as providing good diagnostic confidence with mean values
of 4.0 and 4.5 for water excitation and Dixon, respectively (Figure 2). The
mean scores for the depiction of anatomy, degree of fat suppression, uniformity
of fat suppression, and water signal shading were 4.2, 3.9, 3.4, and 4.0 for
water excitation and 4.4, 4.8, 4.6, and 4.4 for Dixon (Figure 3). All radiologists
scored one technically challenging case poorly for both sequences; when
compared, this case received a similar poor assessment of the standard clinical
images. Reader 1 (6 years experience) preferred water excitation to Dixon in
15/25 cases for overall image quality and in 11/25 cases for fat suppression.
Reader 2 (22 years experience) preferred water excitation to Dixon in 8/24
cases and scored one case as equivalent for overall image quality and preferred
Dixon for fat suppression in all cases. Reader 3 (1 year experience) preferred
Dixon for overall image quality and fat suppression for all 25 cases.
Conclusion and Discussion
DISCO
water excitation allowed for higher spatial resolution whereas DISCO-Dixon
provided more uniform and a higher degree of fat suppression; however sequence
preference varied between readers and both sequences provided diagnostic image
quality. This study focused on the assessment of different fat suppression
techniques at 3.0 T with DISCO to ensure that the image quality and diagnostic
confidence of the current standard-of-care DCE breast MRI method was maintained;
however these techniques have other possible advantages that were not included
in this evaluation, including higher temporal resolution and fat images
generated using the Dixon technique.
Acknowledgements
Acknowledgements: We would like to thank Drs.
Frederick Kelcz and JulieAnn Stover for participating in the reader study. We
are grateful for support from our institutional departmental R&D Fund, the
Radiological Society of North America, NIH (T32CA009206), and GE Healthcare.References
1. Dougherty L, Isaac G, Rosen MA, et al. High frame-rate simultaneous
bilateral breast DCE-MRI. Magn Reson Med 2007;57(1):220-225.
2. Ramsay E, Causer P, Hill K. Plewes D. Adaptive bilateral breast MRI using
projection reconstruction time-resolved imaging of contrast kinetics. J Magn
Reson Imaging 2006;24(3):617-624.
3. Saranathan M, Rettman DW, Hargreaves BA, Lipson JA, Daniel BL. Variable
spatiotemporal resolution three-dimensional Dixon sequence for rapid dynamic
contrast-enhanced breast MRI. J Magn Reson Imaging 2014;40(6):1392-9.
4. Hermann KH, Baltzer PA, Dietzel M, et al. Resolving arterial phase and
temporal enhancement characteristics in DCE MRM at high spatial resolution with
TWIST acquisition. J Magn Reson Imaging 2011;34:973-982.
5. Saranathan M, Rettman DW, Hargreaves BA, Clarke SE, Vasanawala SS.
DIifferential Subsampling with Cartesian Ordering (DISCO): a high
spatiotemporal resolution Dixon imaging sequence for multiphasic contrast
enhanced abdominal imaging. J Magn Reson Imaging 2012;35(60:1484-1492.
6. Ma J. Breath-hold water and
fat imaing using a dual-echo two-point Dixon technique with an efficient and
robust phase-correction algorithm. Magn Reson Med 2004;52(2):415-419.