Eva S. Peper1, Qinwei Zhang1, Bram F. Coolen2, Wouter V. Potters1, Pim van Ooij1, Dennis W.J. Klomp3, and Aart J. Nederveen1
1Radiology, AMC, Amsterdam, Netherlands, 2Biomedical Engineering and Physics, AMC, Amsterdam, Netherlands, 3Radiology, UMCU, Utrecht, Netherlands
Synopsis
Using
a novel 15 channel coil for carotid artery imaging we could prove that the
g-factor loss at higher acceleration factors is limited. This permits the use
of higher parallel imaging factors than commonly exploited in carotid MRI. A 4D
flow scan at high spatiotemporal resolution was accelerated 10 fold (SENSE 4 x
2.5), resulting in high quality images and consistent flow values acquired in a
scantime as short as 7 min.Purpose
MRI
of the carotid arteries has emerged as an important imaging field during the
last decade and has been shown valuable for risk assessment of plaque rupture.
1
The use of surface coils for the neck is mandatory in order to optimize the SNR
performance. Parallel imaging does even further increase the performance of
carotid MRI since several applications are nowadays hampered by excessive scan
duration and limited temporal resolution. Especially 4D flow measurements in
the carotids at sufficiently high spatiotemporal resolution
2 (<1mm,~50
ms) have not been proven feasible to date. We hypothesize that a dedicated 15
channel carotid surface coil allows for 10 fold accelerated 4D flow by
application of parallel imaging in the two phase encoding directions.
Methods
Two
different experiments were conducted in two healthy volunteers (1 male). We
compared SNR and g-factor performance between a conventional 8 channel
bilateral coil (Shanghai Chenguang Medical Technologies, Shanghai, China) and a
novel 15 channel unilateral coil (MR Coils, Drunen, The Netherlands). We
subsequently present a show case to prove the performance of the novel 15
channel coil for 4D flow in the carotids. The 8 channel coil has two coils with
4 coil elements in a 2x2 configuration. The 15 channel coil has a 3x5 configuration,
as described by Koning et al.3, with 5 coils in the
anterior-posterior direction. All data was acquired on a 3T MRI scanner
(Philips Ingenia, Best, the Netherlands). To compare SNR, a multi slice FFE
image stack and the corresponding noise images of the right carotid artery were
acquired using both coils; scan parameters were: FOV 200x200 mm, TR/TE/FA
9.1/4.2/6, 10 slices, 8 mm gap, resolution 1x1 mm, 1 mm slice thickness. The
SNR maps of the 8 channel and 15 channel coils were compared by plotting the
SNR along a manually drawn line at equal positions in the neck.
To compare
the parallel imaging capability of the 15 channel coil, we acquired multiple
single-slice TSE images with increasing parallel imaging factors (SENSE 2,3 and
4) in the AP-direction and g-factor maps were reconstructed on the scanner.
Scan parameters were FOV 150x150mm, TR/TE/FA 800/11/90, ETL 10, resolution
0.28x0.28 mm, 3 mm slice thickness. In addition, the noise correlation between
the 15 elements was assessed in vivo. The loaded and unloaded Q factor was
determined to assess the noise factor of the 15 channel array.
4D flow was
measured using the following sequence parameters: 4 point velocity encoding,
TR/TE 7.6/3.7 ms, Venc (RL,AP,FH) 60/60/100 cm/s, FOV (RL,AP,FH) 215x129x90 cm,
FA 15 degrees, non interpolated resolution (RL,AP,FH) 0.9x0.9x1.4 mm,
temporal resolution 61 ms, SENSE (AP,
FH) 4x2.5 (acceleration factor 10), scantime 7.14 min. For comparison a 2D flow scan was acquired in
the common carotid artery at high temporal resolution of 18 ms.
G-factor maps were
reconstructed for SENSE 4x2.5.
Results
The SNR of
the 15 channel coil was 1.5 times higher at 1 cm depths, whereas at the
location of carotid arteries the SNR was comparable (figure 1A-C). The g-factor
maps measured in a healthy volunteer revealed that with increasing SENSE the
g-factor loss in the 15 channel coil is substantially less compared to the 8
channel coil (figure 2). In the magnitude images (figure 2, gray images) a
better SNR at 3 cm depth for SENSE 4 could indeed be observed for the 15
channel coil, in agreement with the lower g-factor maps. The cross correlation
matrix (figure 1D) shows that the average cross correlation was limited to ~12%.
The ratio between unloaded Q factor (348) and loaded Q factor (97) is 3.6 fold,
demonstrating tissue loss dominance (and a low noise figure).
In the
second experiment 4D flow was successfully performed with large coverage at
high spatiotemporal resolution, as can be appreciated in the streamline
visualization infFigure 3. The flow curve over the heartcycle is still similar
to the flow curve obtained at a 3-fold higher temporal resolution and the peak
flow is captured correctly (figure 4). G-factors remained below 1.5 at the
location of the carotid artery (figure 5).
Discussion
Using a
novel 15 channel coil we could prove that the g-factor loss at higher
acceleration factors is limited. This permits the use of higher parallel
imaging factors than commonly exploited in carotid MRI.
Conclusion
4D flow at
sufficiently high spatiotemporal resolution in the carotid arteries becomes
feasible for the first time using a dedicated 15 channel coil. Once a bilateral 30
channel design becomes available in the near future this coil will allow for
routine 4D flow assessment of the carotid arteries within feasible scantime.
Acknowledgements
No acknowledgement found.References
1. Takaya N, Yuan C, Chu B et al. Association
Between Carotid Plaque Characteristics and Subsequent Ischemic Cerebrovascular Events
: A Prospective Assessment With MRI −−Initial Results. Stroke 2006; 37(3):818–823.
2. Nederveen AJ, Avril S and Speelman
L. MRI strain imaging of the carotid artery: Present limitations and future
challenges. J Biomech. 2014; 47(4):824-833.
3. Koning W, Bluemink JJ, Langenhuizen
EAJ et al. High-resolution MRI of the carotid arteries using a leaky waveguide
transmitter and a high-density receive array at 7T. MRM. 2013; 69(4):1186-1193.