Patrick Magrath1,2, Eric Aliotta1,3, Shams Rashid1, Yutaka Natsuaki4, Xiaoming Bi4, Zhe Wang1,2, Kyung Sung1,2,3, Peng Hu1,2,3, Holden Wu1,2,3, and Daniel B Ennis1,2,3
1Department of Radiological Sciences, University of California, Los Angeles, CA, United States, 2Department of Bioengineering, University of California, Los Angeles, CA, United States, 3Physics and Biology in Medicine IDP, University of California, Los Angeles, CA, United States, 4Siemens Healthcare, Los Angeles, CA, United States
Synopsis
4D-flow MRI is used to quantify
blood flow in a variety of neurovascular pathologies including intracranial
aneurysms [1], but is limited by long scan times as well as moderate spatial
and temporal resolution. Conventional RF pulses have poor slab profiles that
contribute to low sequence efficiency by increasing the field-of-view needed to
avoid aliasing in the slab direction. Our objectives were to design a minimum
time, high Time Bandwith Product (TBW) VERSE pulse for 4D flow and to validate
the improvement in sequence efficiency, confirm flow accuracy, and evaluate
total SAR deposition for VERSE+4D-flow compared to our clinical 4D flow imaging
protocol.Background
4D-flow MRI is used to quantify blood flow in a variety of neurovascular
pathologies including intracranial aneurysms [1], but is limited by long scan
times as well as moderate spatial and temporal resolution. Conventional slice
selective RF pulses have long pulse durations and/or poor slab profiles, which reduce
scan efficiency by increasing the field of view (FOV) needed to avoid aliasing in
the slab direction. Hargreaves et al. [2] described VariablE
Rate Selective Excitation (VERSE) RF pulses which are optimized for short, high time bandwidth product
(TBW) excitation by maximizing either the slice select gradient amplitude or
the RF amplitude at all times. However, VERSE pulses have not previoulsy been reported
for use in 4D flow. Our objectives were: 1) To design and simulate steady-state
slab profiles for minimum time, high TBW VERSE pulses; 2) To compare the slice
profiles generated by TBW=10 (TBW
10) and TBW
28 VERSE
pulses to a conventional TBW
10 windowed sinc (WSinc) pulse used in
our clinical 4D flow sequence; and 3) To validate the improvement in sequence
efficiency, confirm flow accuracy, and evaluate total SAR deposition for VERSE+4D-flow
compared to our clinical 4D flow aneurysm imaging protocol.
Methods
Minimum time VERSE
pulses were created by optimizing high TBW WSinc pulses subject to the following
hardware limits: GMax=43mT/m, SRMax=172T/m/s, B1,Max=13.5µT,
gradient raster time=10µs and RF raster time=5µs. Pulses were designed and
simulated in Matlab (www-mrsrl.stanford.edu/~brian/mintverse/). TBW
10 and TBW
28 VERSE pulses
were implemented on a 3T scanner (Siemens Skyra: 20-channel head coil) in
the 4D flow sequence and evaluated in a stationary aqueous copper-sulfate phantom
and in healthy subjects. VERSE slice profiles were compared to the TBW
10
WSinc pulse in the phantom using a 40mm excited slab thickness (TR=6.41-6.51,
TE=3.80-3.87, FOV=200x182x60mm, matrix-size=176x160x60 FA=15°, BW=822 Hz/pixel). In vivo validation between WSinc
and the TBW
10 VERSE
pulse was performed in the
carotid arteries of healthy volunteers (N=5) using the following parameters: TE/TR=3.80-3.87/6.41-6.51,
FOV=200x182x44-52mm, matrix-size=140x140x44-48, GRAPPA=2, cardiac phases=7-12, scan-time=11:40-19:36,
FA=15°, BW=822 Hz/pixel. All volunteers signed statements of
informed consent.
Results
Bloch simulations of steady-state slab profiles for VERSE
pulses spanning TBW2 to TBW100 were used to select two
VERSE pulses for comparison to the conventional 900µs, TBW
10 WSinc
pulse: a fast 510µs TBW
10 VERSE pulse and an 870µs TBW
28
VERSE pulse with an improved slice profile. Steady-state slab profile
simulations (
Figure 2a) and phantom validation (
Figure 2b-c) demonstrate
that a TBW
28 VERSE pulse provides a sharper slab profile than either
the TBW
10 WSinc pulse or TBW
10 VERSE pulse. For a 40mm
slab a TBW
28 VERSE pulse decreases FWHM by 11% compared to the
standard WSinc pulse. Using a cutoff of 0.15•Mxy, phantom validation
shows that the WSinc pulse excited 9mm out of slab compared to 3mm for the TBW
28
VERSE pulse. Of the 40mm desired profile, 38mm and 36mm were excited above 0.75•Mxy
for the WSinc and TBW
28 VERSE pulse respectively. The results from
simulations and phantom validation suggest that the VERSE TBW
28 pulse
can be used with a 44mm FOV without significant aliasing artifacts. TBW
10
VERSE and TBW
28 WSinc pulses require a 49-52 mm FOV to avoid slab
aliasing artifacts, decreasing acquisition efficiency and extending acquisition times by 2.5 minutes (16%).
Table 1 demonstrates that forward
flow [mL/min] and peak velocity [cm/s] were similar within the right and left common
carotid artery between WSinc and TBW10 VERSE pulses, with a 6.4%
difference in through plane flow and an 8.6% difference in peak velocity.
Figure
3 shows axial and coronal slices using each pulse from a single volunteer. Initial
evaluations show that a decreased slab FOV and reduced number of kz
encodes made possible with the VERSE TBW
28 pulse resulted in an
average total scan time reduction of 2.5 minutes (16%) compared to either the
WSinc or TBW
10 VERSE pulses. In volunteers averaging the SAR for the
TBW
28 VERSE pulse was significantly higher than for the TBW
10
WSinc pulse (1.45±1.04 W/Kg vs. 0.572±.12 W/Kg, P<0.05), but did not
approach SAR limits.
Discussion
High TBW VERSE limits the need for slab oversampling compared to conventional slab selective WSinc pulses. Scan time reductions
were determined based exclusively on decreased phase oversampling. Combining the
slice select rephasing gradient following the VERSE pulse with the
flow compensation waveform will further optimize the sequence and potentially
reduce the TR by an additional 400µs. This would improve temporal
resolution by 6% and allow the flexibility to either acquire additional cardiac
phases or reduce scan time. Further validation and flow evaluation with higher
TBW VERSE pulses is warranted.
Conclusion
High TBW VERSE excitation in 4D flow provides
sharper slab
profiles than Wsinc pulses and improves sequence efficiency.
Acknowledgements
This work was supported by the Department of
Radiological Sciences at UCLA and Siemens Medical Solutions.References
1. Hope, Thomas A.,
Michael D. Hope, Derk D. Purcell, Cornelius von Morze, Daniel B. Vigneron,
Marcus T. Alley, and William P. Dillon. "Evaluation of intracranial
stenoses and aneurysms with accelerated 4D flow." Magnetic
resonance imaging 28, no. 1 (2010): 41-46.
2.
2. Hargreaves, Brian
A., et al. "Variable-rate selective excitation for rapid MRI
sequences." Magnetic resonance in medicine 52.3 (2004):
590-597.