Jose A. Rosado-Toro1, Ryan Avery2, Maria I. Altbach3, Aiden Abidov4, and Jeffrey J. Rodriguez1
1Electrical and Computer Engineering, University of Arizona, Tucson, AZ, United States, 2Radiology, University of Arizona, Tucson, AZ, United States, 3Medical Imaging, University of Arizona, Tucson, AZ, United States, 4Medicine, University of Arizona, Tucson, AZ, United States
Synopsis
We present a semi-automated algorithm for right ventricle segmentation in 4-chamber cardiac MR images. The algorithm takes post-processor landmarks in the end-diastolic and end-systolic frames and generates a segmentation of the right ventricle throughout the cardiac cycle. For the 175 images we analyzed (i.e., 7 patients with 25 frames per patient), the Dice metric was within human variability.Purpose
The analysis of cine MR images plays an important role in the
assessment of cardiac function. Quantitative analysis of cardiac function
requires manual tracing of the ventricular contours by an experienced post-processor,
which is time-consuming. To reduce the
burden on post-processors, semi-automated and automatic algorithms have been
developed. Most algorithms have been designed for the segmentation of the
left ventricle (LV); however, functional parameters for the right ventricle
(RV) are essential for main disease processes such as pulmonary hypertension
and congenital heart disease1.
Most segmentation algorithms applied to the heart fall
into two categories: those using training data and those using low-level
features, such as intensity and gradients.
The main drawback of the algorithms that rely on training data is that the
performance degrades with pathological cases2. The algorithms that focus on low-level
features may converge to an incorrect contour; for example, many experts draw the LV
endocardial border around the blood pool, outside the strongest edge3.
In this abstract we propose a novel
segmentation algorithm for the RV in 4-chamber (4-CH) cardiac images. The new
algorithm uses low-level features and incorporates user input to identify some
landmarks in the RV in 4-CH cardiac images.
Methods
Given a representative cine 4-CH image of the heart, the RV
is outlined as follows. First, the
image is automatically cropped to include only the chambers of the heart (ventricles and
atria). This is accomplished by looking
at the variance of the pixels through the cardiac cycle (e.g., across 25 cardiac frames), thresholding the variance image and finding the bounding box of the largest connected
component in the variance image. Once
the chambers of the heart have been cropped, the post-processor is asked to
identify the end-diastolic frame. This
is the frame where the RV has the largest area. In the end-diastolic frame, the post-processor
has to select four landmarks associated with the RV: the apex, the lateral
and medial sides of the tricuspid valve (TV), and the middle of the blood pool.
These landmarks are shown in Fig. 1a.
Then the post-processor is asked to identify the end-systolic frame.
This is the frame where the RV area is the smallest and can be identified
because the TV is closed. In the
end-systolic frame, the post-processor identifies the TV landmarks (lateral and
medial). These landmarks are shown in Fig. 1b.
Given the landmarks, the algorithm extrapolates the location of the
landmarks throughout the cardiac cycle, as shown in Fig. 2, and performs a
2D segmentation on each frame of the cardiac cycle using a variation of a polar
dynamic programming4 technique developed by our group. The final step is to smooth the 2D
segmentation to ensure temporal smoothness, as shown in Fig. 3.
To evaluate the segmentation performance, we analyzed 175
images coming from 7 different patients (25 frames per patient). Images were acquired using a cine steady-state
free-precession pulse sequence (TR = 2.7, 2.8 ms, TE =
1.16, 1.17 ms, acq. matrix = 168 x 192, FOV = 29.8 x 34 cm, slice thickness = 6
mm) in a Siemens 1.5T Aera scanner.
The segmentation algorithm described above was used
to outline the RV in the 4-CH cardiac images. The semi-automated segmentation result was
compared to the manual tracing of two experienced cardiovascular post-processors or tracers, using
the mean and standard deviation of the Dice metric, which is a measure of mutual
overlap between ground truth and the segmented region.
Results
The segmentation results for three different patients are shown in Fig. 4. The corresponding Dice metric values (calculated using tracer #1 as the ground truth) are 0.913, 0.887 and 0.905. Fig. 5 shows the Dice metric matrix. The matrix shows the intertracer variability
(comparison between tracer #1 and #2) as well as the Dice metric between the
semi-automated algorithm and each of the manual tracers. Note that the semi-automated
technique is within human variability. Specifically, the mean Dice metric between tracer #1
and the semi-automated technique is larger (i.e. higher overlap) than between
tracer #1 and tracer #2. Our technique took on average 2 minutes per patient, as
compared to 20 minutes per patient for the tracers.
Conclusion
We have developed a semi-automated algorithm that accurately segments RV 4-CH MR images within human variability. The RV segmentation in the 4–CH image can be used to measure the ventricular wall motion in the RV
similar to the work done by Nagel et al. for the LV
5.
Acknowledgements
NIH Grant No. T32-HL007955.
References
[1] Driessen MM, Baggen VJ, Freling HG, et al., Pressure overloaded right ventricles: a multicenter study of the importance of trabeculae in RV function measured by CMR, Int. J Cardiovasc Imaging 2014; 30:599-608.
[2] Antunes S, Colantoni C, Palmisano A, et al., Automatic right ventricle segmentation in CT images using novel multi-scale edge detector approach, Comput Cardiol. 2013; 40:815-818.
[3] Mitchell SC, Lelieveldt BP, Van der Geest RJ, et al., Multistage hybrid active appearance model matching: segmentation of left and right ventricles in cardiac MR images, IEEE Trans Med Imag. 2001; 20(5):415-423.
[4] Timp S and Karssemeijer N, A new 2D segmentation
method based on dynamic programming applied to computer aided detection in
mammography, Med Phys. 2004; 31(5):958-971.
[5] Nagel E, Lehmkuhl HB, Bocksch W, et al., Noninvasive diagnosis of ischemia-induced wall motion abnormalities with the user of high-dose dobutamine stress MRI, Circulation 1999; 99(6):763-770.