Kei Fukuzawa1, Rieko Ishimura2, Takashi Yoshida1, Junji Takahashi1, Chiharu Yoshihara1, and Masakatsu Tano1
1Radiology, Toranomon Hospital, Tokyo, Japan, 2Cardiovascular center, Toranomon Hospital, Tokyo, Japan
Synopsis
To obtain the highest R/T ratio, for good
ECG synchronization,
The R/T ratio increased as the angle of
lead II axis became further away from the body axis and as the apex
electrode and the apex of the heart became a little further apart than the
basic placement position.
The basic placement position was the
lowest in the order of R/T ratios among the five kinds of electrode positions
(p < 0.01).Introduction & purpose
Electrodes for vectorcardiography are used
in cardiac MRI (CMR) to synchronize with ECG. In a high magnetic field at or
above 3 Tesla, T-wave amplitude increases inside a bore compared with outside a
bore, and R-wave amplitude decreases relatively. This results in poor ECG
synchronization, such as double triggering and erroneous synchronization with T-waves.
Since poor synchronization leads to low image quality or discontinuation of
examinations, it is necessary to place electrodes so that an R/T-wave amplitude
ratio (R/T ratio) becomes higher in 3 Tesla CMR. In a survey for ten radiology
technicians from multiple facilities who were experienced in CMR imaging, 14
out of 14 answered that it was good to line up the basal and apex electrodes
(namely, lead II) with the heart axis. Also, 11 out of 14 answered that
synchronization was poorer in obese patients whose heart axis is away from the
head-tail body axis due to eventration of the diaphragm. However, the clinical
study in our facility
1) found that the closer the heart axis was to the body
axis, the poorer synchronization was and that the wider the angle between the
heart axis and lead II axis was, the better T waves were controlled to gain high
R waves. Taking the results above into consideration, we examined ECG electrode
placement positions to increase the R/T ratio in healthy volunteers.
Method
The
study subjects were eight healthy volunteers (three males and five females, 28.1±3.2 years old). ECG waveforms were recorded inside
and outside a bore with a 3.0 Tesla MRI system (Ingenia, Philips Healthcare,
Best, The Netherlands). As shown in Figure 1, ECG electrodes were placed in the
following five positions: first, the basic placement position of lead II in line
with the heart axis, the second and third positions were variations of the
angle of lead II axis against the heart axis (angle of intersection), and the
fourth and fifth positions were where the lead II was fixed at the angle in line
with the heart axis, electrodes were moved in parallel to the left to create
variations of the distance between the apex electrode and the apex of the
heart. Then, R/T ratios in all five positions were compared, placed in the descending
order of R/T ratio, and statistically analyzed with the Kruskal-Wallis test.
Results
R/T
ratios in the basic placement position were 3.04±1.25 outside a bore and 1.12±0.28 inside a bore. T-wave elevation inside a bore
was verified in healthy volunteers without cardiac disease as well. The angle
of intersection in the basic placement position was 0.63 ± 3.7 degrees. When the angle of
intersection was 12 ± 3.2 degrees, the R/T ratio was 2.15 ± 0.66, and when the angle of
intersection was 29.7 ± 7.6 degrees, the R/T ratio was 2.11 ± 0.48. The relationship between
the angle of intersection and the R/T ratio was that the R/T ratio increased as
the angle of lead II axis became further away from the head-tail body axis. When
the distance between the apex electrode and the apex of the heart was 26.0 ± 7.1 mm, the R/T ratio was 1.79 ± 0.32 and when the distance
between the apex electrode and the apex of the heart was 28.3 ± 5.7 mm, the R/T ratio was 1.88 ± 0.59. The relationship of the
distance between the apex electrode and the apex of the heart to the R/T ratio
was that the R/T ratio increased as the apex electrode and the apex of the
heart became a little further apart than the basic placement position of 18.8 ± 7.2 mm. In all subjects, the
basic placement position was the lowest in the order of R/T ratios among the
five kinds of electrode positions (average order 5.0, p < 0.01). The results above found that there was a discrepancy
between the ECG electrode placement position considered ideal by radiology
technicians engaged in CMR and the placement position that actually increases
R/T ratio.
Conclusion
To
obtain the highest R/T ratio, the angle of lead II axis against the heart axis must
be closer to the horizontal plane and the apex electrode and the apex of the
heart must be apart. Further studies are necessary to decide which one of the
two points is more effective.
Acknowledgements
No acknowledgement found.References
K. Fukuzawa, et al.
Patient Factors and Electrode Placement Factors
that Cause 3T-CMR to Poorly Synchronize with Waveforms of Vectorcardiography –
Examination of Factors Worsening R/T-wave Amplitude Ratio –. ISMRM 2016 abstract submitted. No. 2048