Michael Loecher1, Peng Hu1, and Daniel B Ennis1,2
1Department of Radiological Sciences, University of California, Los Angeles, CA, United States, 2Biomedical Physics, University of California, Los Angeles, CA, United States
Synopsis
4D Flow phase contrast MRI acquisitions inherently require a measure of
background phase to remove phase contributions from non-velocity based
components. The temporal dynamics of
this background phase are not well understood. Consequently, the background phase may be
measured too infrequently or too often for accurate and/or time efficient measurements.
The purpose of this work was: 1) to measure the temporal dynamics of the
background phase with high temporal resolution; and 2) to demonstrate methods
of selecting time optimal background phase sampling strategies that improve the
measurement efficiency of 4D Flow acquisitions.Introduction
4D Flow phase
contrast MRI acquisitions inherently require an estimate of background phase to
remove phase contributions from non-velocity based components. Traditionally,
this measurement is acquired once during each cycle of velocity encoding steps, or every 4 TRs (~20-30ms), however many modern methods have been introduced that sample background phase much less frequently1-3.
This assumes that the background phase does not change by a significant amount
during each velocity encoding cycle, but that it does change substantially over
the cardiac cycle. The temporal dynamics of background phase are
not well understood for 4D Flow acquisitions.
Consequently, the background phase may be measured too infrequently or
too often for accurate and/or time efficient measurements.
From
results seen in 2D through plane PC-MRI4, it is expected that background
phase does not change as rapidly as velocity phase, and 4D Flow sequences could
be potentially accelerated by sampling background phase less frequently. The purpose of this work was to measure the
temporal dynamics of the background phase with high temporal resolution, and demonstrate
methods of selecting time optimal background phase sampling strategies to improve
the measurement efficiency of 4D Flow acquisitions.
Methods
Blood-mimicking
fluid was pumped (CardioFlow 5000, Shelley Medical) through a straight tube
phantom with three different programmed waveforms: sinusoidal, femoral, and
carotid. Two healthy human volunteers were also scanned covering the carotid
arteries and the middle cerebral artery (MCA).
4D Flow MR
imaging was performed at 3.0T (Skyra, Siemens) with: FOV=24x18x4cm, 1.25mm
isotropic, TE/TR=2.7/6.25ms, flip angle=10°, venc=70-100cm/s, and BW=800Hz/pixel.
The temporal resolution of the phantom experiments was 25ms (27 cardiac
phases) and the temporal resolution of the volunteer scans was 50ms (18-20
cardiac phases). High temporal resolution sampling of the background
phase was acquired with identical sequence parameters, but without
flow-encoding gradients. Consequently, the number of acquired phases was quadrupled
to 108 (phantoms) and 72-80 (humans).
Flow values
were extracted from the regular 4D Flow measurements using standard
post-processing methods. For the background phase measurements, all
phases were offset by the negative of the first cardiac phase, to mitigate
phase wrapping. Spectral decomposition of the background phase was analyzed
after mean detrending. The high temporal resolution background phase data
was used to estimate peak and mean velocity errors for various background
sampling rates, effectively varying the number of TRs (i.e. velocity encodes)
between background phase measurements.
Results
Figure 1 shows representative velocity waveforms and the
underlying background phase for a phantom waveform and a volunteer. For phantom experiments, the total change in
background phase (max-min of a vessel ROI mean) was 0.14±0.05 radians (1.8±0.6cm/s) for the 3 waveforms. For in
vivo experiments the total change in background phase was 0.29±0.001 radians (4.7±1.2cm/s) for the 2 volunteers. Figure 2
shows the frequency components of the background phase for all experiments. Figure 3 shows the results of simulating
different sampling rates on velocity error.
Analysis of the potential velocity error (Figure 3) shows increasing
error with slower sampling rates, however even the peak velocity errors seen
with the slowest sampling rate (12 TR, 68.75ms) would only amount to a velocity
error of roughly 10%, as compared to the roughly 5% error seen with traditional
4 TR sampling rates.
Discussion
This work shows how background phase can be measured with high temporal resolution and used to evaluate the requirements of background
phase sampling in 4D flow. The results confirm that the background phase is a
small, but significant source of potential error in 4D Flow MRI.
This study demonstrated methods of selecting sampling rates by
looking at the frequency components of the background phase variation as well as
simulating the effect of different sampling rates on the mean and peak error in
velocity measurements. It was seen that the
frequency components of the background phase contained significant low
frequency components on top of noise.
From analyzing the potential error due to changes in background phase,
it was seen that mean and max phase errors tended to increase with less
sampling, though the scale of the error was small (<0.1 radian or 2-3
cm/s). Depending on the nature of the
measurement, this additional error could be made up for with more efficient
measurements of velocity components, leading to net VNR improvements.
This study is limited in the number of volunteers/territories examined. Ideally all
vascular regions could be classified by the scale of their respective
background phase dynamics. Additionally,
this work only looks at periodic changes in background phase over the cardiac
cycle, however other sources of motion (e.g. respiratory) or temperature
changes were not investigated.
Acknowledgements
This research was supported, in part, by Siemens Healthcare and the Department of Radiological Sciences at UCLAReferences
1. K. M. Johnson and M. Markl, “Improved SNR in phase contrast velocimetry with five-point balanced flow encoding.,” Magn. Reson. Med., vol. 63, no. 2, pp. 349–55, Feb. 2010.
2. C. Binter, V. Knobloch, R. Manka, A. Sigfridsson, and S. Kozerke, “Bayesian multipoint velocity encoding for concurrent flow and turbulence mapping.,” Magn. Reson. Med., vol. 000, pp. 1–9, Jun. 2012.
3. N. R. Zwart and J. G. Pipe, “Multidirectional high-moment encoding in phase contrast MRI.,” Magn. Reson. Med., vol. 69, no. 6, pp. 1553–64, Jun. 2013.
4. D. Wang, J. Shao, S. Rapacchi, M. J. Middione, D. B. Ennis, and P. Hu, “Phase contrast MRI with flow compensation view sharing,” Magn. Reson. Med., vol. 23, p. 2737, 2014.