Does Respiratory Motion Influence Tissue Phase Mapping Velocities?
Jan Paul1, Stefan Wundrak1, Peter Bernhardt1, Wolfgang Rottbauer1, and Volker Rasche1

1Internal Medicine II, University Hospital of Ulm, Ulm, Germany

Synopsis

Cartesian Tissue Phase Mapping (TPM) usually necessitates respiratory navigators or other means of motion selection/correction to avoid ghosting artifacts. In radial MRI, however, motion artifacts result in image blurring rather than ghosting, which might allow using all respiratory states for reconstruction. The aim of this study is to investigate the influence of respiratory motion on velocities obtained from radial Tissue Phase Mapping MRI. Only small biases towards reduced velocity peaks were found in ungated compared to motion-compensated reconstructions. Overall velocity agreement of ungated data was very high compared to gated reconstructions.

Motivation

In Cartesian Cardiac MRI, respiratory motion leads to distinct ghosting artifacts. This necessitates respiratory navigators or other means of motion selection/correction also for Cartesian Tissue Phase Mapping (TPM) [1]. With radial MRI, however, motion artifacts result in image blurring rather than ghosting, which might allow using all respiratory states for reconstruction. The aim of this study is to investigate the influence of respiratory motion on velocities obtained from radial Tissue Phase Mapping MRI.

Methods

Acquisition: In 10 healthy volunteers, a retrospectively triggered radial golden angle black-blood TPM sequence with 4-point balanced (Hadamard) velocity encoding and VENC = 30 cm/s was applied in 3 short axis slices. Acquisition parameters were: TR/TE = 5.6 ms / 3.4 ms, flip angle α = 15°, resolution = 1.6 x 1.6 mm2, slice thickness = 8 mm, and FOV = 340 x 340 mm2.

Reconstruction: Raw data was exported and processed with Matlab. An image-based self-gating signal [2] was generated for respiratory gating. The respiration signal was divided into 6 respiration bins with accepting window widths adapted to ensure similar undersampling in each bin. The bin with the most occurring respiratory position was defined as the reference bin. Bins with more than 6 mm acceptance window were rejected to avoid risk of intra-bin motion. The following cine reconstructions were performed:

· FREE: no respiratory gating (accept all data);

· REF: only use reference bin;

· AVG: reconstruct respiration bins separately, then average;

· MC: as AVG, but perform motion correction (affine image registration [3] in ROI around heart) before averaging.

Reconstructions were done by gridding (FREE; low undersampling) or iteratively (all others) by GRASP [4].

Analysis: Image quality and velocity measures for the different reconstructions were compared by Wilcoxon signed-rank tests with Bonferroni correction for multiple testing and p-values below 5% were considered significant.

Results

Figure 1 shows magnitude and velocity images for the different reconstruction methods. Visually, REF has less SNR than the other reconstructions. FREE and AVG appear slightly blurred, whereas MC is as sharp as REF. Velocity analysis can be appreciated in figure 2. Segmental velocities over the cardiac cycle are similar for all reconstruction methods, with only slightly spatial deviations in REF. Global velocities are also highly similar over the cardiac cycle. Slight reduction of peaks velocities, however, are apparent compared to REF, especially in FREE. Velocity peak times between the different reconstruction methods are neither different in the segmental nor in the global analysis.

Statistical analysis (Figure 3) confirms the visual impressions. SNR is improved in all methods compared to REF due to the usage of all acquired rather than only selected data. Sharpness and image contrast are decreased in FREE and AVG, but similar in MC to REF. Only some of the velocity peaks are reduced in FREE compared to REF. Overall agreement of velocities with REF as measured via velocity correlation and RMSE is higher in AVG and MC than in FREE.

Conclusion

While consideration of respiratory motion seems to be necessary in Cartesian TPM, free-breathing reconstruction is feasible for radial TPM without significant influence on the resulting velocities. The small bias towards reduced velocity peaks might be decreased via respiratory binning and (motion-compensated) averaging, thus keeping the 100% navigator efficiency of FREE compared to typically lower acceptance rates of 40%-60% in gated acquisitions (REF).

Acknowledgements

This study was partially funded by Philips Healthcare and the NVIDIA Hardware Donation Program.

References

[1] Jung et al.: MRM 2006, doi: 10.1002/mrm.20808

[2] Paul et al.: MRM 2014, doi: 10.1002/mrm.25102

[3] http://sourceforge.net/projects/niftyreg/

[4] Feng et al.: MRM 2013, doi: 10.1002/mrm.24980

Figures

Figure 1: Comparison of magnitude and velocity images in one volunteer for the different reconstruction methods. Noise appears increased in REF compared to the other methods. Slight image blurring is visible in FREE and AVG, but not in REF and MC.

Figure 2: Comparison of velocities over the cardiac cycle in one volunteer for the different reconstruction methods. Segmental velocity patterns appear similar for all methods (a), with slightly increased noise-like deviations in REF. Analysis of global velocities (b) reveals slight underestimations of peak velocities (arrows), mainly for FREE compared to REF. Abbreviations of the segments: A – anterior, AL – anterolateral, IL – inferolateral, I – inferior, IS – inferoseptal, AS – anteroseptal.

Figure 3: Statistical comparison of image and velocity information between the different reconstruction methods (“↑”: significant increase, “↓”: significant decrease, “=”: not significant). Compared to REF, SNR/VNR is increased. Sharpness and contrast are decreased in FREE and AVG, but similar in MC to REF. Some velocity peaks are reduced in FREE compared to REF, but similar otherwise. Overall agreement of velocities with REF (correlation and RMSE) is higher in AVG and MC than in FREE.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2585