Cristian Montalba1, Jesus Urbina1,2, Julio Sotelo1,3, Marcelo Andia1,4, Cristian Tejos1,3, Pablo Irrarazaval1,3, Israel Valverde5,6, and Sergio Uribe1,4
1Biomedical Imaging Center, Pontificia Universidad Católica de Chile, Santiago, Chile, 2School of Medicine, Pontificia Universidad Católica de Chile, Santiago, Chile, 3Electrical Engineering, Pontificia Universidad Católica de Chile, Santiago, Chile, 4Radiology Department, Pontificia Universidad Católica de Chile, Santiago, Chile, 5Institute of Biomedicine of Seville, Universidad De Sevilla, Seville, Spain, 6Cardiology Unit, Hospital Virgen del Rocio, Universidad de Sevilla, Seville, Spain
Synopsis
4D flow is a MRI technique characterized by long scanning times. Because
of that, it is difficult to study the variability of flow parameters when subjected
to changes of the MR parameters. The purpose of this work is to study
the variability of different flow parameters due to changes of spatial and
temporal resolutions in 4D flow acquisitions through controlled experiments
using a realistic normal adult thoracic aortic phantom. We conclude that changing
the spatial and temporal resolutions in the 4D flow imaging greatly affects
different flow parameters with induced errors of up to 23.9%.Purpose
4D flow is a MRI technique characterized by long scanning times, which
leads to that the acquisition of these data sets is usually confined to a small
volume with restricted spatial and temporal resolution. Recently, there has
been a consensus regarding the minimum requirements of the MR parameters for
the acquisition of 4D flow data (1). However, due to the long scanning
times, it is difficult to study the variability of the results obtained when subjected
to changes of the MR parameters without having other variables influencing the
results. The purpose of
this work is to study the variability of different flow parameters due to
changes of spatial and temporal resolutions in 4D flow acquisitions through
controlled experiments using a realistic normal adult thoracic aortic phantom.
Methods
The experiments were performed using a whole-body
1.5T MR scanner (Philips Achieva, Best, Netherlands) with a 4-element
phased-array body coil. The phantom model (T-S-N-005,
Elastrat Sarl, Geneva, Switzerland) is used with a pulsatile MR-compatible flow
pump (Simutec, London, Ontario, Canada), with the same features as explained by
Urbina et al (2). We are able to control the pump in
terms of flow waveform, heart rate and cardiac output, which allows us
to study six different hemodynamic conditions: heart rates of 68 and 88 bpm,
each one with different maximum flow rate of 200, 230 and 260 mL/s. For each
hemodynamic condition, two 2D phase contrast (PC) sequences were acquired in
the ascending and descending aorta (Figure 1). Also, nine 4D flow data were
acquired in the entire phantom with different combinations of spatial and
temporal resolutions (Table 1). Furthermore, two acquisitions with medium
(2.0x2.0x2.0mm3) and low spatial resolution (1.5x1.5x1.5mm3) and with 40ms of temporal resolution were
re-acquired for reproducibility analysis. Each phantom condition was scanned at
different days and the data acquisitions lasted about 5 hours each session.
We compared the peak flow, mean velocity, maximum
velocity between all 4D flow data with the 2D PC sequence. The analyses were
performed using a commercial software GT Flow 2.2.15 (Gyrotools LLC, Zurich,
Switzerland).
Results
Figure 2 summarizes the results of peak flow,
mean velocity and maximum velocity in each hemodynamic condition comparing all
4D flow data with 2D PC data at the ascending (Ao1) and descending aorta (Ao5).
We observed a high variability of the results obtained when subjected to
changes of the studied MR parameters, particularly sensitive to changes of the
temporal resolution. Considering the 2D PC sequence as gold standard, the mean,
minimum and maximum peak flow errors are showed in Table 2. We observed minimum
errors when the data is acquired with high temporal resolutions and greater
errors when the data was acquired with low temporal resolutions. Result of the
reproducibility experiment, including the mean difference (bias) and the
standard deviations of difference between the first and second measurement are
summarized in Table 3.
Discussion
We
observed a high variability of the results obtained and a greater accuracy of
the flow parameters when using high temporal resolution compared to 2D PC
values. From Figure 2 we observed that the temporal resolutions greatly
affected the results obtained. However, for a given temporal resolution the results
showed small variations for the different spatial resolutions. These findings
can be appreciated for any hemodynamic condition studied. Low spatial
resolution images obtained result more similar than high spatial resolutions. Result
of the reproducibility studied showed an excellent agreement between the two
measurements for the medium and low resolutions in Ao1 and in Ao5, showing a
maximum bias of 3.2% for both spatial resolution studied.
Conclusion
Changing the spatial and temporal resolutions
in 4D flow imaging greatly affects different flow parameters that induced
errors of up to 23.9%, being the 4D flow sequence particularly sensitive to
changes of the temporal resolution.
Acknowledgements
Grant
Sponsor: Anillo ACT1416, Grant
Sponsor: Fondo Nacional de Desarrollo Científico y Tecnológico (FONDECYT),
Ministerio de Educación, Chile. Grant Number: FONDECYT #1141036.References
1.- Dyverfeldt et al. (2015). 4D flow
cardiovascular magnetic resonance consensus statement. J Cardiovasc Magn Reson. 2015 Aug 10; 17(1):72.
2. - Jesus Urbina et al. A realistic MR
compatible thoracic aortic phantom to study coarctations using catheterization
and cine PC-MRI sequences. ISMRM 2014.