Niranjan Balu1, Jie Sun1, Daniel S Hippe1, Huijun Chen2, and Chun Yuan1
1Radiology, University of Washington, Seattle, WA, United States, 2Tsinghua University, Beijing, China, People's Republic of
Synopsis
We
have developed a single sequence (SNAP) for identification of high-risk
features such as stenosis, intraplaque hemorrhage and juxtaluminal
calcification. However measurement of plaque burden, an early marker for
disease, required a separate black-blood scan. In this work, we demonstrate
that plaque burden can be measured using an alternate reconstruction (SNAP2).
In patients with atherosclerosis, we compare vessel wall area measurements on
SNAP2 to established black-blood MRI measurements and show that our method
provides results comparable to established plaque MRI methods.Introduction
Atherosclerotic plaque
features such as intraplaque hemorrhage (IPH) and juxtaluminal calcification
(JCA) are considered to indicate high-risk status [1]. Carotid plaques with
these features are more likely to cause stroke due to thromboembolism. Recently
a phase-sensitive inversion-recovery (IR) based sequence (SNAP) [2] was extended
to include a PD weighting (SNAP2) that can identify calcification and show its
JCA status [3]. Even though IPH and JCA can be identified in this manner,
measurement of plaque burden which is an earlier marker of disease than
IPH/JCA, required a separate black-blood sequence. In this work, we show that
plaque burden can be reliably measured using the SNAP2 formulation alone.
Aims
To compare plaque burden measurement between SNAP2
against 3D black-blood (3D BB) carotid plaque MRI.
Materials and methods
Image acquisition: Bilateral
arteries (N=14) from 7 patients with 50-79% stenosis by ultrasound were
examined with MRI. Using an 8 channel phased array carotid coil, SNAP MRI was acquired
with parameters: TR/TE 10/4ms, flip angle 11/5 (corresponding to ⍺ and θ in fig 1), TI 500ms, Resolution 0.8x0.8x0.8mm, FOV (coronal)
16x16x3.2cm, turbo factor 98, scan time 5 min. Thus the first image (I1)
was T1-weighted and the second image (I2) was PD weighted.
Image
reconstruction: A polarity function P(x,y)
which takes values (-1 or +1) depending upon the longitudinal magnetization was
calculated using I1(x,y)
and I2(x,y) as
$$P(x,y)= (I_1 (x,y)I_2^* (x,y))/‖I_1 ‖‖I_2 ‖$$
where
* represents complex conjugation. T1-weighted corrected real image was then
obtained as $$$S_1 (x,y)=P(x,y)‖I_1 (x,y)‖$$$ and PD-weighted
corrected real image was obtained as $$$S_2 (x,y)=P(x,y)‖I_2 (x,y)‖$$$. A sigmoid function was used in
reconstructing S2(x,y)
such that gray-blood PD-weighted vessel wall image (SNAP2) was available for plaque
burden measurements. A previously validated 3D black-blood sequence [4] was
used for reference standard plaque burden measurements. Imaging parameters
were: TR/TE 10/4ms, flip angle 6, Resolution 0.8x0.8x0.8mm, FOV (coronal)
25x16x3.2cm, turbofactor 30, scan time 3.5 min.
Patient image review: The coronally acquired SNAP2 and 3D BB were reformatted
into axially oriented 2mm slices. Lumen and outer wall boundaries were drawn
separately on SNAP2 and 3D black-blood MRI by two different reviewers. Slices
were matched across the two sequences based on their location relative to the
carotid bifurcation. Percent wall volume (%WV) was computed as wall area/total
vessel area x 100% and summed over slices per artery. Wall, lumen and total
vessel areas were also averaged for each artery. Artery level measurements were compared
between the two sequences.
Statistics: The non-parametric
bootstrap and percentile method was used to calculate 95% confidence intervals
(CIs) for the ICCs. Overall bias was tested using a permutation test.
Results
Common coverage reviewed ranged from 4 to 30
slices (median: 19, mean: 17.4 ± 10.2 slices). On an artery level, there was no
significant difference in wall area between SNAP2 and 3D BB measurements (32.6
± 8.3 mm2 vs 35.8 ± 8.1 mm2, p=0.17). Figure 2 shows
representative slices of SNAP2 plaque boundaries vs 3D BB boundaries. Figure 3
shows corresponding Bland-Altman plot. There was good ICC for lumen and total
vessel areas but ICC was low for wall area. However %WV measure of plaque
burden showed improved ICC (table 1).
Discussion
Measures of plaque burden such as wall area and
%WV were comparable between SNAP2 and 3D BB. Lumen area and total vessel area
tended to be larger on SNAP2 which could be due to reviewer variability or
difference in sequence contrast and blood suppression capabilities.
Conclusions
Plaque
burden can be measured using SNAP2 and was comparable to
plaque measurements on 3D-BB. Thus use of SNAP2 along with other SNAP derived
image weightings (SNAP2 and I2) may provide a time-efficient means
for comprehensive high-risk plaque detection using a single sequence.
Acknowledgements
No acknowledgement found.References
[1] Saam Radiology 2007; 244(1):64-77,
[2] Wang MRM 2013; 69(2):337-45. [3] Balu ISMRM 2015. [4] Balu MRM 2011;
65(3):627-37