Atsushi FUKUYAMA1, Haruo ISODA2, and Shuji KOYAMA2
1Radiological Sciences, Nagoya University Graduate School of Medicine, Nagoya, Japan, 2Brain & Mind Research Center, Nagoya University, Nagoya, Japan
Synopsis
Our
objective is to develop a phantom that can evaluate the image distortion along
the body axis objectively, and elucidate the relationship between the occurrence
and geometrical position of image distortion in MRA. As a result, the phantom that we have developed has made it
possible to evaluate image distortion objectively over a wide range of the
trunk. We made clear that when separated from the center of the magnetic
field in vertical or horizontal directions, image distortion became larger with
the increase in distance. Purpose
To develop a phantom
that can evaluate the image distortion along the body axis objectively, and
elucidate the relationship between the occurrence and geometrical position of
image distortion in
Magnetic
Resonance Angiography (MRA). Furthermore, we validated the characteristics and
utility of distortion
correction filters.
Methods
We used a MAGNETOM Verio 3T (Siemens, Erlangen,
Germany) as the MR device and the Body matrix coil and Spine matrix coil
together for the receiver coil. The phantom for evaluating image distortion was
an acrylic container 40cm in width, 50cm in depth, and 6cm in height. Seven
acrylic pipes (with an inner diameter of 11mm and an outer diameter of 15mm) were
arranged in equal intervals of 50mm inside the container. In this experiment, we
enclosed a 1% aqueous solution of gadolinium inside the acrylic pipes and
filled distilled water around it (Fig.1).
First, we placed the phantom in such a way that it coincided with the acrylic
pipe placed right at the center of the
magnetic field of the MR device. While doing so, we adjusted the height of the
phantom using polystyrene foam. We performed coronal plane
imaging around the acrylic pipe in the center, and performed sagittal plane imaging individually at three acrylic pipes - one at the center, one
shifted to the left by 50mm, and one shifted by 100mm respectively. Next, we
used polystyrene foam of 50mm thickness and carried out imaging at three
locations similar to the above mentioned method: one at 50mm above the center
of magnetic field, one at 50mm below the center of magnetic field and one at
100mm below the center of the magnetic field (Fig.2).
The MR images acquired in the coronal planes and
sagittal planes were reconstructed in the transverse plane, and the central coordinates of the acrylic pipe were obtained
from the measurements at seven places. Then we evaluated the
image distortion from the front view and
lateral view.
Results
When the acrylic pipe was placed in the center of
the magnetic field, image distortion did not occur at all (Fig.3). When the acrylic
pipe was shifted up or down and left or right, the foot and the head of the acrylic pipes became distorted towards the center of the
magnetic field. The distortion ratio was at its maximum of 5% at a location
10cm away from the center of the magnetic field in the direction of the body axis,
and it was at a maximum of 10% at a location 15 cm away. As the distance from
the center of the magnetic field increased, the image distortion ratio also
increased. Regardless of whether the imaging plane was set to coronal or
sagittal, no significant difference was observed in the degree of distortion.
The two-dimensional distortion correction filter
was able to correct only the in-plane image
distortion that exists in the slice plane and it had no effect on the image
distortion that occurs across the slices (Fig.4). The three-dimensional distortion
correction filter was able to correct the image distortion across the slices as
well (Fig.5).
Discussion
The equipment used in this experiment has built-in technology to make the magnetic field cylindrically uniform based on the human body.
1This technology has enabled the imaging of areas that were not possible to date using the spherical magnetic fields. However, the image distortion cannot be reduced if the correction filters are not used. This would not be a problem in the imaging of cerebrovascular and neck blood vessels, where the imaging range along the direction of the body axis is not longer than 10 cm. However, since the imaging range is longer in MRA imaging of the body trunk, much image distortion occurs at locations that are away from the center of the magnetic field.
2-4 Since the image distortion that occurs in MR devices may hinder medical diagnosis, the image distortion should be proactively reduced by varying the imaging position and through the use of filters.
Conclusion
The phantom that we have developed has made it
possible to evaluate image distortion objectively over a wide range of the
trunk.
5,6 When separated from the center of the magnetic field in vertical or
horizontal directions, image distortion became larger with the increase in
distance. In the case of a two-dimensional distortion correction filter, no
significant effect can be obtained unless an optimal plane is not determined in
accordance with the distortion occurrence conditions. However, in the case of a
three-dimensional distortion filter, it was possible to correct the distortion
under any condition.
Acknowledgements
We would like to thank the Brain and Mind Research Center at Nagoya University for permitting us to use their MR device for this study.References
1. TrueForm Magnet and Gradient Design.
Siemens AG. 09, 2011. http://www.healthcare.siemens.com. Accessed November 1, 2015
2. Hong C, Lee DH, Han BS.
Characteristics of geometric distortion correction with increasing field-of-
view in open-configuration MRI. Magn Reson Imaging. 2014;32(6):786-790
3. Viard R, Betrouni N, Vermandel M, et al.
Characterization and 3D correction of geometric distortion in low-field open-
magnet MRI. Conf Proc IEEE Eng Med Biol Soc. 2008:3649-3652
4. Baldwin LN, Wachowicz K, Thomas SD, et al. Characterization, prediction, and correction of geometric distortion in 3 T MR
images. Med Phys. 2007;34(2):388-399
5. Wang D, Doddrell DM, Cowin G. A novel phantom and method for comprehensive 3-dimensional
measurement and correction of geometric distortion in magnetic resonance
imaging. Magn Reson Imaging. 2004;22(4):529-542
6. Mizowaki T, Nagata Y, Okajima K, et al. Reproducibility of geometric distortion in magnetic resonance imaging
based on phantom studies. Radiother Oncol. 2000;57(2):237-242