MR visible localization device for ex vivo radiographic-pathologic correlation
Sonja Kinner1,2, Tilman B Schubert1,3, Christopher Brace1, Elisabetta A Nocerino1, Timothy J Colgan1, Shannon Hynes1, Scott B Reeder1,4,5,6,7, and Emily R Winslow8

1Department of Radiology, University of Wisconsin School of Medicine and Public Health, Madison, WI, United States, 2Department of Diagnostic and Interventional Radiology and Neuroradiology, University Hospital Essen, Essen, Germany, 3Clinic for Radiology and Nuclear Medicine, Basel University Hospital, Basel, Switzerland, 4Department of Medical Physics, University of Wisconsin School of Medicine and Public Health, Madison, WI, United States, 5Department of Emergency Medicine, University of Wisconsin School of Medicine and Public Health, Madison, WI, United States, 6Department of Medicine, University of Wisconsin School of Medicine and Public Health, Madison, WI, United States, 7Department of Biomedical Engineering, University of Wisconsin School of Medicine and Public Health, Madison, WI, United States, 8Department of Surgery, University of Wisconsin School of Medicine and Public Health, Madison, WI, United States

Synopsis

In this work we describe the design and preliminary validation of a novel MR-compatible localization device aimed at facilitating lesion detection in ex vivo tissue, such as resected liver. Experiments in two swine liver specimens with small microwave ablations to simulate lesions demonstrated the successful feasibility and ease of use of this localization device. Further evaluation in human liver tissue specimens containing metastatic lesions will be needed to demonstrate the performance of the device for ex vivo radiographic-pathologic correlation.

Introduction

Rapid analysis of pathologic specimen to assess tumor resection margins and tumor burden during surgery remains a challenging task [1]. Ex vivo identification of lesions found at in vivo imaging can be particularly demanding due to deformation and altered orientation of the resected specimen. In order to facilitate lesion detection and improve direct comparison of pathologic specimens with prior in vivo imaging, we developed an MRI compatible localization device, with MR-visible fiducial grid markings to facilitate ex vivo imaging and radiographic-pathologic co-localization and correlation. The purpose of the present study was to test the feasibility of this device and to evaluate the sensitivity and specificity in liver specimens.

Methods

Imaging device description

The device is entirely constructed from Plexiglas and designed to consist of two stationary MR visible grids and one removable grid that together create a 3D matrix (Figure 1). Total size of the device is 27x14x14cm3. Along the width of the box, laser-etched grid lines and labels were filled with silicone gel (Dow-Corning, Midland, Michigan, US) to facilitate MR-visible fiducial grid marks. Etched numbers (1-11) were oriented along the length, upper case letters (A-W) along the width, lower case letters (a-l) along the length. All grid lines were spaced 1cm apart (Figure 2).

When examining an ex vivo tissue specimen, alginate can be used as fixation for the histopathologic specimen to stabilize tissue (eg. liver) during imaging and cutting. Once the alginate hardens to a soft solid, MR examination of the device, containing the tissue specimen is performed. Accurate localization of lesions in three dimensions is determined by referencing the lesions in space relative to the three dimensions of the MR-visible grid.

After lesion identification in MRI, the removable vertically oriented grid along the length of the box is removed and the specimen can be sliced in a guided fashion along the vertical grid lines (Figure 3) using prefabricated grooves spaced 1cm apart to facilitate slicing at the identical locations corresponding to transaxial slices aligned with one dimension of the device grid.

Imaged specimen

To test the device, two specimens of swine liver were used for radiographic-pathologic correlation. In each specimen, lesions were created using a random number (0-15) of ultrasound-guided microwave ablation zones. The number and location of the ablations were blinded to the MR-readers. Ablations were carried out in three different sizes (5, 10 and 15 mm), based on previous calibration of the time-energy deposition relationship.

MR-protocol

MR imaging was performed on a clinical 3T scanner (GE Healthcare Discovery MR750, Waukesha, WI) using an 8-channel head coil. Acquisition parameters of a 3D spoiled gradient echo T1 weighted acquisition included: TR/TE=9.8/4.1ms, flip angle = 20deg, BW=+/-162 kHz, FOV=280x252x180mm3, 450x420x225 matrix, for true spatial resolution of 0.6x0.6x0.8mm3 interpolated to 0.5x0.5x0.4mm3 through zero-filling. Total scan time was 21:16min.

Imaging analysis

Two radiologists blinded to number and localization of lesions independently analyzed the MR images. The total numbers and coordinates of the lesions as well as lesion diameter (measured on MR images) were reported on a spreadsheet.

Histopathological analysis and correlation

After analysis of MR data, the specimens were sliced according to the MR-analysis spreadsheet and lesions were measured within each slice using a ruler with 1cm spaced etching. If lesions were identified within the coordinate-slice and not on the cutplanes between the coordinate slices, the cut section was cut again longitudinally. Histological findings were reported on the same spreadsheet as the radiographic data.

Discussion

In this work we described the design, and preliminary feasibility testing and evaluation of an MR-visible device intended for radiographic-pathologic co-localization and correlation for ex vivo tissue specimens with internal lesions. Use of the device proved feasible with 100% specificity and 100% sensitivity for both readers on a per specimen and per lesion basis. 5 and 6 lesions respectively were under- or overestimated in size (+/-3mm) according to pathological evaluation.

Conclusion

The MR-visible device was feasible for the detection and three-dimensional localization of liver lesions, and has potential to play an important role in lesion identification and in radiologic-pathologic evaluation. Future evaluation of the device will include human liver tissue specimens containing metastatic lesions.

Acknowledgements

The authors wish to acknowledge support from the NIH (UL1TR00427, K24 DK102595), the Societt of Surgeons (The Alimentary Tract Young Investigator Grant) as well GE Healthcare who provides research support to the University of Wisconsin.

References

1. Juettner et al., Eur J Cardiothorac Surg. 1990

Figures

Figure 1: Box components and their appearance on MR Imaging.

Figure 2: Maximum Intensity Projection MPR of the MR Images of the box with loaded liver (A). Note that the alginate (blue component in B) is not showing up in MR Imaging. Arrows in B point to the slicing channels every 1cm.

Figure 3: A view of the tissue specimen in the box (A), outside the box for lesion measurements and localization (B) and as appearing on MR Imaging (C). Note the exact correlation of the liver lesion (arrow) ex vivo and on MR Imaging. On MR Imaging the grid can be seen as marked lines and dots (arrowheads in C), depending on the orientation.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2521