Jianxun Qu1, Hairui Xiong2, Bing Wu1, and Junhai Zhang2
1MR Research China, GE Healthcare, Beijing, China, People's Republic of, 2Department of Radiology, Huashan Hospital Affiliated to Fudan University, Shanghai, China, People's Republic of
Synopsis
In body
perfusion measurement with arterial spin labeling technique, single shot acquisition
was usually used, which suffers from susceptibility induced distortion. A
straightforward way was to reduce readout echo train, which however lead to
lower spatial resolution. In this work, we performed flow sensitive alternating
inversion recovery (FAIR) with reduced FOV excitation to shorten echo train
length while maintaining spatial resolution in prostate cancer perfusion
imaging.Purpose
Perfusion
measures the level of blood supply to body tissues, and its abnormality conveys
important physiological information for diagnosis, especially for cancer.
Comparing to conventional contrast based perfusion, arterial spin labeling
(ASL) based perfusion is contrast free and offers quantification without the
need to time the arrival of contrast bolus. Although growing popular in the
neuroimaging, the use ASL based perfusion in body remains largely in the
research field, mostly constrained by the susceptibility related image
distortion since single shot acquisition is generally used. A straightforward
way to reduce these effects is to shorten the readout train length, which however
leads to lowered spatial resolution. In this work, we investigate the use of
reduced FOV excitation [1] in ASL based perfusion to shorten the readout train
length while maintaining the spatial resolution, in the application of prostate
cancer.
Method
Flow sensitive
alternating inversion recovery (FAIR) [2] based pulsed ASL perfusion with rFOV
excitation was implemented as illustrated in
Fig.1. In FAIR perfusion module, a hyperbolic scant pulse was used
to achieve selective and global inversion. A Type-I 2D RF [1] was used for
excitation with spatial selection in both phase encoding and slice selective
direction. Single shot spin echo EPI (SE-EPI) was used for readout. All MR
experiments were performed in a 3.0T whole body system (GE Discovery 750). A patient
diagnosed with prostate cancer underwent MR examination including multi-phase FAIR
perfusion acquisition and multi-phase DCE liver acceleration volume acquisition
(LAVA). Consent form was acquired prior to the study. The imaging parameters
for FAIR were following: FAIR, FOV 80 * 48 mm; slice thickness 10 mm; matrix 48
* 48. Inversion time (TI) of 1.0, 1.2, 1.4 s were used with recovery time of
3.0 s. Selective inversion slab with a thickness of 20 mm was used to uniformly
invert imaging plane. Sixteen pairs of label and control acquisitions were made
for averaging. LAVA used the following parameters: TE/TR = 1.3/3.0 ms; FOV 360
* 360 mm; slice thickness 3.0 mm; slice number 64; matrix 256*160; and flip
angle 15
o. Forty phases were acquired with temporal resolution of
9.0 s. DCE pharmacokinetics was analyzed in GEN-IQ (GE Advanced workstation
4.6).
Result
T2, T2 fused K
trans
map and T2 fused rFOV ASL perfusion map (TI = 1.2 s) are shown in
Fig.2, in the latter case, the actual
excited FOV is indicated as a rectangular region. The hypo-intense regions in
T2 (
Fig.2a) indicate the likely
position of the lesions, which is verified by the hyper-intense DCE perfusion
signal in
Fig.2b. Corresponding
hyper-intense ASL perfusion regions as indicated by the pink arrows (1 and 2)
are also observed in
Fig.2c. The
other two hyper-intense perfusion regions (blue arrow 4, 5) on FAIR ASL image
are likely vascular regions. To verify the high perfusion nature of the lesion,
a normal tissue region was selected as indicated by arrow 3. The unnormalized
label and control signal level of repeated FAIR measurements for regions 1, 2,
and 3 are shown in
Fig.2d, clear
differentiation of the perfusion signal level can be seen. Although TI of 1.2 s
was shown, hyper-perfusion presented with all TI used.
Discussion
and Conclusion
In this work,
the feasibility of using reduced FOV ASL perfusion in prostate cancer is
investigated. The hyper-perfusion region observed in rFOV ASL was consistent
with that of conventional DCE, and shown as lesion in the T2 image. Not only
rFOV excitation reduces susceptibility related artifacts while maintaining
spatial resolution (In-plane resolution of 1.67mm was achieved that was
comparable to that of 1.41mm in DCE), it is also able to avoid region with
known artifacts such as motion or lipid. Sufficient SNR was obtained even with
rFOV excitation due to sufficient averages, as seen by the distinct perfusion
levels in lesion and healthy tissues. Although the use of TI was not
specifically optimized, all the TIs used led to sufficient level of perfusion
signal. It is hoped that the use of reduced FOV may help with the broad
application of ASL based perfusion in body imaging.
Acknowledgements
No acknowledgement found.References
[1] E. Saritas, et al. MRM 2008;
[2]
W Cai, et al. NMR Biomed 2014;