Jurgen H Runge1, Jules L Nelissen2,3, Larry de Graaf2, Barbara Molenkamp4, Suzan van der Meij4, Klaas Nicolay2, Gustav J Strijkers3, Jaap Stoker1, Anneloes E Bohte1, Aart J Nederveen1, Ondrej Holub5, and Ralph Sinkus5
1Radiology, Academic Medical Center, Amsterdam, Netherlands, 2Biomedical NMR, Eindhoven University of Technology, Eindhoven, Netherlands, 3Preclinical and Translational MRI, Academic Medical Center, Amsterdam, Netherlands, 4Surgery, Academic Medical Center, Amsterdam, Netherlands, 5Biomedical Engineering, King's College London, London, United Kingdom
Synopsis
Distinction between benign and malignant breast lesions remains difficult with conventional (dynamic) contrast-enhanced MRI. MR Elastography (MRE) can distinguish benign and malignant tissues based on their
viscoelastic properties but breast MRE has not found widespread use in daily
clinical practice, because of the complex equipment required and cumbersome data acquisition. Here we present a compact, easy to handle breast MRE set-up that allows the acquisition of high quality, artefact-free MRE data. This set-up was designed, built and tested at two different institutions in volunteers and a patient.Target audience
Clinicians, scientists interested in cancer research, breast
imaging, and MR Elastography.
Purpose
Breast cancer is the second most common cancer and one in
eight women is expected to develop this type of cancer during her life.
Treatment options include (neo-adjuvant) chemotherapy, radiotherapy and
surgery, with many women opting for breast-saving therapy: excision of the
tumour while the breast remains in place. To reduce breast cancer-related
mortality, screening programs have been implemented in many countries to detect
early cancers. Population screening is done with X-ray mammography, often with
follow-up by (dynamic) contrast-enhanced MRI. Although MRI has
high sensitivity, its specificity is only moderate due to overlap between
benign and malignant lesions resulting in false-positive findings and hence,
relatively low positive predictive value
1. MR Elastography (MRE) has
shown to be a promising technique for distinguishing between benign and
malignant tissues based on their viscoelastic properties
2. However,
breast MRE has not found widespread use in daily clinical practice, probably
due to the relatively complex MRE actuator set-ups necessary to generate shear
waves for MRE. Many proposed breast MRE actuator set-ups use a coil as engine
(engine-coil) to generate the shear wave, which is closely placed to the
breast. The B-fields generated by these coils disturb the imaged FOV in close
vicinity to the engine-coil, which often results in artefacts that reduce the
valid image volume. In this study we present a compact clinically easy to
handle breast MRE set-up where the engine-coils are placed at a distance from
the breast, resulting in high quality artefact-free MRE images.
Materials & Methods
Breast MRE setup
The set-up, shown in Fig. 1,
consists of (A) two bilaterally
placed 8 Ohm engine-coils, wound on a custom 3D ABS plastic printed holder, fixed
on a (B) in-house designed PEEK plastic
mechanical spring to transmit mechanical waves to both breasts simultaneously.
The latter is done via (C) splined
nylon low friction rods (Sullivan Products, UK), movable in a shaft, coupled to
(D) another PEEK spring part mounted
inside a (E) 7-channel breast
RF-coil (Philips Healthcare, The Netherlands). A (F) 3D printed C-shaped breast-holder is fixed to this (D) PEEK spring part while an (G) Urias® inflatable splint (Arden
Medical Ltd, United Kingdom) allows custom cup-size fixation of the breast with
an (H) oppositely placed C-shaped
breast-holder. For patient safety the engine-coils were positioned in a (I) closed box. The (J) head support was designed to cover
the (C) rods and to provide a
comfortable place for the positioning of the head. Overview of the whole set-up
positioned on top of the MR scanner bed, is shown in Fig. 2.
In
vivo MRI
MRE was performed on two clinical 3T MR Scanners (Achieva and Ingenia, Philips
Healthcare, The Netherlands) in both the Academic Medical Centre (AMC),
Amsterdam, The Netherlands and King’s College London (KCL), London, United
Kingdom to test the set-up. In total 5 healthy volunteers (2×KCL, 3×AMC) and 1
post-lumpectomy patient (1×AMC) were scanned. MRE acquisitions were performed according
to Garteiser et al. with in-phase TE=9.6ms,
fMECH=30Hz and 60Hz, fMEG=120Hz and 2.5×2.5×2.5 mm3
voxels in 3 minutes per motion-encoding direction3. A waveform
generator and amplifier were used to drive the engine-coils. Raw phase images
were visually inspected for artefacts and processed in dedicated software to
obtain elasticity maps4.
Results & discussion
Artefact free breast MRE data were acquired in all subjects.
Fig. 3 shows a representative
example of the curl of the displacement (
fMECH=60Hz)
for the X, Y, and Z directions of both breasts of a healthy volunteer.
Excellent shear wave coverage is observed through the differently sized breasts.
The movable C-shaped breast-holder (
Fig.
1H) for adjusting the fixation
to the breast size assures good coupling. In
Fig. 4 the wavelength image and the elastogram of the same subject
as in
Fig. 3 is shown together with
a high-resolution anatomical scan. The regions with longer wavelength and higher
shear stiffness accurately co-localize with the parenchyma in the anatomical
image. Finally,
Fig. 5 shows a
same-breast comparison of the curls of the displacement for the X direction for
the two
fMECH,
demonstrating excellent coverage at both frequencies.
Conclusion
A compact and easy to handle breast MRE set-up was
successfully designed, built and tested in two different institutions on two different
types of Philips 3T MR scanners. Artefact free high quality MRE images of both
breasts were acquired simultaneously with the set-up.
Acknowledgements
No acknowledgement found.References
1. Mahoney
MC, Gatsonis C, Hanna L et al. Positive predictive value of BI-RADS MR
imaging. Radiology. 2012;264(1):51-8. 2. Siegmann
KC, Xydeas T, Sinkus R et al. Diagnostic value of MR elastography in addition
to contrast-enhanced MR imaging of the breast-initial clinical
results. Eur Radiol. 2010;20(2):318-25. 3. Garteiser P, Sahebjavaher RS, Ter Beek
LC et al. Rapid acquisition of multifrequency, multislice and multidirectional
MR elastography data with a fractionally encoded gradient echo
sequence. NMR Biomed. 2013;26(10):1326-35. 4. Sinkus R, Siegmann KC, Xydeas T et al.
MR elastography of breast lesions: understanding the solid/liquid duality can
improve the specificity of contrast-enhanced MR mammography. Magn Reson
Med. 2007;58(6):1135-44.