Byeong-Yeul Lee1, Xiao-Hong Zhu1, and Wei Chen1
1Center for Magnetic Resonance Research, Radiology, University of Minnesota, Minneapolis, MN, United States
Synopsis
Spatial
averaging of multiple high-resolution CSI (hrCSI) voxels is commonly employed to
gain SNR and improve quantification of metabolites. Using in vivo 17-oxygen 3D CSI, we compared SNR between spatial averaging
of multiple hrCSI voxels and a single voxel acquired with low-resolution CSI
(lrCSI) with matched sample volume and position. SNR from voxel averaging was much
lower than that of lrCSI caused mainly by the increased noise level by spectral
summation. This
study clearly demonstrates that the acquisition of high-resolution data with
spatial averaging faces a large trade-off of SNR. Therefore, it should be taken
consideration carefully for the choice of an appropriate voxel size of
high-resolution CSI for in vivo study
of neurological or metabolic diseases. Purpose
To
quantitatively examine the SNR implication of multiple-voxel averaging from high-resolution
CSI data, resulting in a significant SNR loss compared to a single voxel from low-resolution
CSI data with matched spatial resolution.
Introduction
High-resolution
chemical shift imaging (hrCSI) is highly desired for basic and clinical
research due to less partial volume effect, thus better specificity for
identifying and studying diseased tissue. However, lower signal-to-noise ratio
(SNR) caused by a small voxel size impedes reliable quantification of
metabolites, particularly challenging for low-γ
X-nuclear MRS, such
13C and
17O spins. As one of solutions
to overcome the SNR issue, spatial averaging of multiple hrCSI voxels taken
from the region of interest has been frequently employed. In this context, it
is interesting to examine if the averaged signal from multiple hrCSI voxels
could substantially sacrifice
a SNR compared to a large single voxel acquired from low-resolution CSI
(lrCSI) with matched spatial resolution and position, and if yes, how much? To
address this critical question, we conducted a thorough SNR comparison study of
X-nuclear
17O CSI of natural abundance water in phantom and human occipital lobe
in vivo with
two spatial resolutions.
Methods
All experiments were performed at 7.0T/90-cm bore scanner (Siemens) using
RF surface coil and 8 healthy volunteers participated this study. The Fourier
Series Window (FSW) technique
[1-2] was used to acquire all 3D
17O
CSI data under full relaxation condition (
17O
T1 = 4.5ms, TR = 100ms, 1610 k-space scans; spectral width = 30kHz;
9x9x7 phase encodes using two field of views (FOV): (8×8×6 cm
3) hrCSI and (24×24×18 cm
3) for lrCSI, resulting in 27 times of volume difference.
A total of 27 hrCSI voxels’ FIDs (see Figs.1 and 2) were summed in time
domain. No line broadening was applied
to measure true noise levels and SNR. The RF
transmission field (B
1+) of
17O RF coil was
mapped (Fig. 2) for calculating flip angle (B
1+) and its effect on the
17O water intensity.
The noise correlation among inter-voxels in the hrCSI was measured. Finally, the
effect of point-spread function (PSF) was estimated based on the cylindrical
voxel shape from the FSW-CSI method
[1-2].
Results and Discussion
Figure 1 shows a typical in vivo 17O CSI distribution of H217O signal and the selection of target voxels of interest used for SNR comparison. Noise and linewidth (FWHM) of single voxel spectra were not significantly different between hrCSI and lrCSI data, and independent on the voxel position. However, the summed spectrum from 27 hrCSI voxels showed a much high noise
level as well as higher signal intensity (Fig. 1D) compared to the single
voxel spectrum from the lrCSI with the similar sample volume and position (Fig.
1E). The signal level difference indicates that the true hrCSI voxel size was larger than the apparent voxel size due to the spatial
overlapping associated with PSF (Fig. 4). The estimated PSF from the FSW-CSI
method accounted ~41% signal increase. The B1+ maps shown
in Fig. 2 were highly inhomogeneous among hrCSI voxels and had
effects on the 17O signal, for example, counting for ~35% and ~33% signal
loss for phantom and in vivo,
respectively. In addition, PSF induced the noise coherence among
adjacent hrCSI voxels as illustrated in Fig. 3, resulting a high noise
level ratio (phantom ~7.8, in vivo ~
9.1) compared to low resolution CSI. With consideration and correction of the
PSF and B1+ effects on each hrCSI voxel, phantom and in vivo
brain data consistently suggest that multiple-voxel averaging of hrCSI data resulted
in a huge SNR reduction (~5.4 times) compared to that
of the lrCSI.
This experimental result was close to the theoretical
prediction of 5.2 based on the relationship
between the multiple-voxel averaged SNR (SNRm) and single voxel SNR (SNRs): $$$SNR_m = (\sqrt{NumberofMeasurement} \times SNR_s $$$
if inter-voxel
noise is completely random and independent (i.e., PSF=0), and B1
distribution is homogenous in space. As we
demonstrated, this theory might be invalid for practical CSI
application commonly with significant PSF and B1 inhomogeneity,
especially for a surface coil used in this study, thus, the SNRm calculation should be modified
accordingly: $$Signal_m =(\sum_{i=1}^{27}Signal_i\times sin(\alpha_i))\times PSF$$ $$Noise_m = NoiseCorrelation\times \sqrt{{Numberof Voxels}}\times \sigma_n$$ where σn is the mean voxel noise.
Conclusion
This study
demonstrates that the spatial voxel averaging of hrCSI data results in a huge SNR
trade-off, which should be useful for selecting an appropriate voxel
size of high-resolution CSI for
in vivo
application. A better strategy is to
optimize the CSI voxel size for achieving the desired SNR, rather to
averaging multiple hrCSI voxels for matching the same SNR that requires a much
long scanning time.
Acknowledgements
NIH grants of R24 MH106049, RO1
NS070839,
S10 RR029672, P41
EB015894 and P30 NS076408References
[1] Garwood et al., JMR, 75:244-261 (1987); [2] Zhu et al., MRM, 45:543-549 (2001);