Synopsis
Hyperpolarized
13C MR imaging can provide simultaneous assessments of metabolism
and perfusion to study disease processes. High resolution dynamic
imaging is needed to fully understand these processes, but is challenging, especially on clinically relevant systems. This project investigated new methods for
spectral selectivity with SNR-efficient bSSFP sequences to provide improved high
resolution 3D dynamic in vivo HP 13C
MR imaging at 3T. Purpose
Recent
advances in hyperpolarization (HP)
13C methods have allowed exceptional
characterization of metabolism and perfusion changes, specifically with HP
pyruvate and urea MRI, respectively, that underlie various diseases processes
1,2. The
ability to obtain both sets of information and/or resolve multiple compounds in
one scan via copolarization has been demonstrated
in vivo in prior hyperpolarized
13C studies
3. Several
different k-space acquisitions have been utilized to achieve this spectral
selectivity, including echo-planar spectroscopic imaging and echo-planar
imaging
4,5,
although these acquisitions typically have lower spatial resolution. The bSSFP
sequence offers the highest SNR per unit time, and consequently highest spatial
resolution, but exhibits off-resonance effects that make spectral
selectivity relatively more difficult
6.
While different methods for spectral selectivity have been successfully applied to the
bSSFP sequence, particularly at high field
7,8,
adapting the sequence for high resolution 3D dynamic acquisitions at the clinically
relevant field strength of 3T is more challenging. In this project we aimed to
investigate two different methods for spectrally selective high resolution 3D
dynamic bSSFP of copolarized [2-
13C]pyruvate and [
13C,
15N
2]urea
at 3T.
Methods
Two different methods for spectrally selective bSSFP tested on a 3T GE MR scanner were (referred to by number from here on):
(1) frequency encoding of both species simultaneously using a low gradient with
wide readout bandwidth (“multiband” encoding)
9;
(2) individual acquisition of each species applying spectrally selective RF
pulses with TR chosen to place the compound of interest in the passband and the
other in the stopband. Figure 1 depicts each method as it would be used to
acquire images of copolarized [2-13C]pyruvate and [
13C,
15N
2]urea.
13C thermal phantom tests were performed with a custom 3D bSSFP
sequence (single timepoint) to display efficacy of each of the methods. Two
spherical phantoms, one filled with bicarbonate and one filled with acetate,
were imaged. Method 1 aimed to separate the two phantoms by low bandwidth
frequency encoding, while method 2 aimed to alternately excite and suppress an
individual spectral line (i.e. bicarbonate signal) when on-resonance and
off-resonance, respectively.
In vivo experiments
were initially performed on one Sprague-Dawley rat with copolarized [2-
13C]pyruvate and [
13C,
15N
2]urea, which have a frequency
difference of 1412 Hz at 3T. Both methods were acquired
with a flip angle of 50°, after a 25° preparatory
pulse (no slice select gradients), 2.5 mm isotopic
resolution, 6 s temporal resolution for each compound, 10 dynamic timepoints for
a total scan time of 60 s, and the scans were started at 15 s after beginning
of injection. The different experimental parameters were as follows: (1) 48 x
24 x 8 matrix size, 1.6 ms pulse with a TBW of 4, TR of 21.9 ms, and rBW of
2.824 kHZ; (2) 24 x 24 x 8 matrix size, 2 ms pulse with a TBW of 2, TR of 7 ms,
and rBW of 15.625 kHZ. Both phantom and
in
vivo experiments were acquired using partial Fourier acquisition, with only ¾ of the
phase encodes acquired. A projection onto convex sets (POCS) reconstruction was
employed to reconstruct the raw data, and further reconstruction for the
multiband data was performed as described previously
9.
In
vivo DNP experiments used a HyperSense polarizer, and 3mL of copolarized 80 mM
[2-
13C]pyruvate/80 mM [
13C,
15N
2]urea
was injected via a tail vein catheter.
Results and Discussion
The
phantom experiments successfully showed the capability of using both methods
at 3T. Figure 2 shows the representative slice
13C images for both methods. Method
1 successfully features bicarbonate and acetate in two separate bands, and both spherical phantoms are in their correct location in
the recombined image. Method 2 successfully demonstrated that choosing the appropriate TR results in the bicarbonate signal being located in the stopband and consequently
the spherical phantom disappearing. Figure 3 shows the 3D urea images next to
a representative
1H image of the rat for both methods for the first dynamic
timepoint. Figures 4 and 5 show all 10 dynamic timepoints for both methods for
urea and pyruvate, respectively. Method 2 appeared to have better SNR and
longer lasting dynamics.
Conclusion
The
ability to acquire spectrally selective high resolution 3D dynamic images at 3T
via the bSSFP sequence is demonstrated here using two separate methods,
multiband encoding and spectrally selective RF pulses with an appropriate TR. Individual
images of [2-
13C]pyruvate and [
13C,
15N
2]urea
were successfully acquired within one scan. These methods can be potentially applied to other systems that require imaging of multiple
resonances, such as pH measurement with HP [
13C]bicarbonate. With
further optimization of each method, higher resolution can be achieved and
applied to various disease processes, and ultimately translated to clinical
imaging.
Acknowledgements
The
authors would like to thank Dr. Jeremy Gordon, Mark Van Criekinge, Lucas
Carvajal, Hsin-yu Chen, and Zihan Zhu for all their help and funding from the
NIH (P41EB013598).References
1. Kurhanewicz, J. et al. Analysis
of cancer metabolism by imaging hyperpolarized nuclei: prospects for
translation to clinical research. Neoplasia 13, 81–97 (2011).
2. Von Morze, C. et al. Investigating
tumor perfusion and metabolism using multiple hyperpolarized 13C compounds:
HP001, pyruvate and urea. Magn. Reson. Imaging 30, 305–311
(2012).
3. Von Morze, C. et al. Simultaneous
Multiagent Hyperpolarized 13C Perfusion Imaging. Magn. Reson. Med. 72,
1599–1606 (2014).
4. Larson, P. E. Z. et al. Fast
Dynamic 3D MRSI with Compressed Sensing and Multiband Excitation Pulses for
Hyperpolarized 13C Studies. Magn. Reson. Med. 65, 610–619 (2011).
5. Gordon, J. W., Machingal, S.,
Kurhanewicz, J., Vigneron, D. & Larson, P. Ramp-Sampled, Symmetric EPI for
Rapid Dynamic Metabolic Imaging of Hyperpolarized 13C Substrates on a Clinical
MRI Scanner. in Proceedings of the International Society for Magnetic
Resonance in Medicine 23 (2015) 4605.
6. Scheffler, K. & Lehnhardt, S.
Principles and applications of balanced SSFP techniques. Eur. Radiol. 13,
2409–2418 (2003).
7. Von Morze, C. et al.
Frequency-specific SSFP for hyperpolarized 13C metabolic imaging at 14.1 T. Magn.
Reson. Imaging 31, 163–170 (2013).
8. Varma, G. et al. Selective
spectroscopic imaging of hyperpolarized pyruvate and its metabolites using a
single-echo variable phase advance method in balanced SSFP. Magn. Reson.
Med. 00, n/a–n/a (2015).
9. Von Morze, C.
et al. Multi-band frequency encoding method for metabolic imaging with
hyperpolarized [1-(13)C]pyruvate. J. Magn. Reson. 211, 109–13
(2011).