Ina Vernikouskaya1,2, Alexander Pochert3, Mika Lindén3, and Volker Rasche1,2
1Internal Medicine II, University Hospital of Ulm, Ulm, Germany, 2Small Animal MRI, University of Ulm, Ulm, Germany, 3Inorganic Chemistry II, University of Ulm, Ulm, Germany
Synopsis
Quantification of 1H
MR contrast agents (CA) is limited by the only indirect visualization of the changes
of the relaxation properties of the surrounding tissue. Using alternative
nuclei such as fluorine (19F) as CA enables direct and quantifiable
readout of local CA aggregations, since the 19F signal linearly
correlates with its local concentration. However non-uniformity of the
transmit/receive radiofrequency fields impact the resulting absolute signal,
leading to wrong quantification results. Application of an easy-to-use time-efficient
B1+/B1--mapping technique for
correction of the 19F signal in vivo is presented in this
work.PURPOSE
The
19F concentration
in a specific tissue can be quantified by direct comparison of the resulting
image intensity with that of a reference sample. However, this approach relies on
uniform spin excitation and signal reception. Due to tissue properties and MR
coil characteristics, the local deterioration of the transmit (
B1+)
and receive (
B1-) fields between the targeted
region of interest and the location of reference introduce errors in the
quantified
19F signal. In previous work
1, we applied a fast
and accurate method for
B1+ and
B1-
fields mapping, based on the spoiled gradient echo (
SPGR) approach
2.
This work shows the preclinical
in vivo
evaluation of the proposed
19F quantification technique with
B1+/B1--correction
after systemic injection of hollow mesoporous silica sphere (
HMSS) nanoparticles loaded with
perfluoro-15-crown-5-ether (
PFCE) (
HMSS-PFCE)
3.
METHODS
Considering B1+
and B1- effects, the resulting
measured signal intensity for SPGR acquisition with selected flip angle
(FA) αn is:
$$I_{\alpha_{n}}(r)=cM_{0}(r)B_1^-(r)\frac{\sin(B_1^+(r)\cdot\alpha_{n})(1-E_{1})}{1-E_{1}\cos(B_1^+(r)\cdot\alpha_{n})}E_{2}, [1]$$ where M0(r)
is the equilibrium longitudinal magnetization, c a constant, $$$E_{1,2}=e^{-\frac{TR}{T_{1,2^{*}}}}$$$, B1-(r) the relative receive coil characteristic, and B1+(r) the
relative FA variation. For quantitative imaging, an image intensity I0(r) proportional to M0(r) is
required. Neglecting c and E2 in Eq. (1), I0(r) can be approximated by: $$I_0(r) \approx \frac{I_{\alpha_{n}}(r)}{w_{ss}(r, \alpha_{n})\cdot B_1^-(r)}, [2]$$ where wss(r,αn) is a steady-state factor depending on B1+(r). Moreover,
in case of using a transmit/receive coil B1-(r)
= B1+(r)
allowing to compensate for transmit and receive effects after calculation of B1+(r) from ratio of the signal
intensities acquired with two different FAs.
The technique was tested in one NMRI mouse
anesthetized using i.p. ketamine/xylazine injection to avoid any signal
contamination from inhaled anesthesia gases.
About 200µL of the HMSS-PFCE dispersion (theoretical
load of 11.23mg PFCE) was injected systemically into the tail vein. A reference
probe of 20mM (11.7mg/mL) concentration of PFCE diluted in n-Hexane was placed
into the field-of-view (FOV). MR
scanning was performed directly after injection (20 to 40min
post-injection), 24h, and 96h post-injection. Two SPGR acquisitions with the
following parameters: FOV = 3.2x4cm, matrix = 32x40, slice thickness 3mm, TE/TR
= 1.7ms/350ms, NEX = 56, and FA = 30/45° resulting in 13min acquisition time
per FA - could be performed within the anesthesia time.
19F
signal was quantified based on the data acquired 24h
post-injection. B1+ mapping was performed on a slice-by-slice
approach using previously measured PFCE T1 value of 650ms. Pixel-wise concentration was
calculated based on the non-corrected and B1+/B1--corrected
signal intensity images acquired at 30° and 45° respectively. Calculation of
the total PFCE volume was done by integrating the 19F signal over
all slices.
RESULTS
No fluorine signal
from the mouse abdomen could be detected directly after intravenous injection
of HMSS-PFCE nanoparticles. 24h post-injection pronounced fluorine signal could
be detected exclusively in the mouse liver (Figure 1), which was completely gone 96h
post-injection. Respective
correction data considering transmit and receive effects could be calculated
from 30° and 45° data. Resulting concentration maps derived from the signal
intensity images acquired at 30° and 45° flip angles before and after
correction are shown in Figure 2. An overview of the mean PFCE concentrations in
the liver calculated for fluorine containing slices for corrected and
non-corrected data is provided in Figure 3. Correction yielded similar
19F
concentrations for different FAs, whereas without correction the resulting
concentrations differed by up to 25%. The PFCE quantity in the liver as
evaluated from the
B1+/B1--corrected
concentration maps resulted to 9.73mg (for 30° FA image) and 9.59mg (for 45° FA image), corresponding to more than 85% of
the injected amount of PFCE. Quantification from non-corrected
data yielded high differences between the two FAs and lower absolute values of 7.10mg
(for 30° FA image) and 8.23mg (for 45° FA image).
DISCUSSION
Without correction, the inhomogeneous
B1+ profile
yielded inaccurate quantification results as obvious by the huge differences of
the
19F concentrations derived from two FAs. With the proposed
correction, the resulting concentrations were almost independent on the FA
indicating good correction of the transmit and receive effects. The PFCE amount in the liver
calculated after
B1+/B1--correction
(~10mg) does still not perfectly match the theoretical value of 11.23mg.
Remaining differences may be attributed to actually lower amount of particles
in injected solution, non-complete loading of the particles, variations of the
particle properties and partly excreted or still circulating particles 24h
after injection.
CONCLUSION
In
conclusion, proposed method for correction of
B1+ and
B1- inhomogeneities
enables accurate quantification of local
19F aggregations with
acquisition times suitable for
in
vivo applications without requiring any modifications of the imaging
sequence.
Acknowledgements
No acknowledgement found.References
[1] Vernikouskaya I, Pochert A, Rasche V.
19F MRI Quantification using B1 Correction. ISMRM 2015; 3697.
[2] Sung K, Saranathan M, Daniel BL,
Hargreaves BA. Simultaneous T(1) and B(1) (+) mapping using reference region
variable flip angle imaging. Magn Reson
Med. 2013;70(4):954–61.
[3] Vernikouskaya I,
Pochert A, Linden M, Rasche V. Perfluoro-15-Crown-5-Ether-Loaded Hollow
Mesoporous Silica Spheres for 19F In Vivo MRI. ISMRM 2015; 1902.