Maggie Mei Kei Fung1, Ek Tsoon Tan2, David Soon Yiew Sia3, and Darryl Sneag3
1MR Apps & Workflow, GE Healthcare, New York, NY, United States, 2MR, GE Global Research Center, Niskayuna, NY, United States, 3MRI Research Lab, Hospital of Special Surgery, New York, NY, United States
Synopsis
Diffusion tensor imaging (DTI) can potentially
be helpful in visualizing peripheral nerves and assessing nerve damages.
However, upper extremity DTIs (wrist, elbows & arm) are susceptible to
distortion and fat suppression failure, especially in arms-down position where
the area of interest is far from iso-center and can have more B0 inhomogeneity.
In this study, we aim to investigate whether a combination of B0 correction
methods can help reduce fat suppression failure, improve spatial misalignment
and thus improve nerve tracking. We observed consistent fat suppression
improvement at the wrist, but no significant improvement in spatial accuracy.Purpose:
DTI has been utilized in imaging nerve
injuries and peripheral nerve tumors[1].However, upper extremity(UE) DTI is
susceptible to distortion and fat suppression failure, especially in the
patient-friendly arms-down position where the area of interest is far from
iso-center and has increased B0 inhomogeneity. Spatial mismatch makes it difficult to do proper seeds placement and tracks
the nerve in its entirety in fiber tracking software. In this study, we aimed
to investigate the utility of a realtime B0 correction(RTB0) method[2], and an image registration-based distortion correction
(ADDC) method [3,4] in UE-DTI.
Methods:
To reduce B0 inhomogeneity & fat
suppression failure, we applied the RTB0 technique, which examined the FID for
each slice to determine the optimal center frequency and apply it to each slice
in realtime (Fig 1). To further reduce
misalignment due to tissue susceptibility and eddy-current-induced distortion,
the ADDC post-processing technique is used , where a rigid registration of the DTI
T2 image to the PD image was performed. To account for susceptibility &
eddy current effects, an additional refinement of alignment was performed using 20-parameter cubic-polynomial and 4-parameter linear basis-function
respectively.
Peripheral nerve DTIs with and without
RTB0 were performed on 3 consented healthy volunteers on a GE 3T 60cm bore
scanner (MR750) using a 16channel Flexible extremity array in arms down
position. Optimized DWI parameters were:
FOV:12cm(LR)x9.6cm(AP), Matrix: 80(freq) x 64(phase),
TR/TE:5000ms/60.5ms (non-RTB0), 66.5ms(RTB0), single spin echo, BW:250kHz, Slice
thickness:3mm, # slices:25/station (i.e. 9cm per station), Adiabatic IR fat
suppression , TI=79ms, b-value=0s/mm2(4 NEX), 600s/mm2 (4
NEX), diffusion direction: 15, scan time: 5:25min. Since the most common injury sites of the
upper extremity are wrist, elbow and upper arm, we performed DTI at these 3
sites for each volunteer (adjusting the coil placement for each region of
interest). Both datasets were then processed through ADDC.
To assess the spatial accuracy, an
experience radiologist identified the median, ulnar and radial nerve positions
on the proximal, mid, distal locations of each imaging site. The nerve positions
were identified on the following datasets: 1. Anatomical PD images, 2. DTI
without RTB0, 3. DTI with RTB0, 4. DTI without RTB0 & ADDC, 5. DTI with
RTB0 & ADDC. Using the PD image as gold standard, the absolute displacement
errors were computed for datasets 2 to 5 in AP and LR direction, and compared
using Wilcoxon rank sum test. Fat suppression quality was also noted in the
comparison.
Results &
Discussion:
Figure 2 shows the center frequency offset calculated from RTB0 at awrist,
elbow and arm respectively on a volunteer. For elbow & arm, since they are
relatively cylindrical in shape, the CF offset is mainly driven by magnet B0
inhomogenity, which is minimal in the 9cm SI coverage range. While in wrist,
where there is shape changes at the wrist joint, there is a higher CF offset
due to changes in tissue susceptibility. This large CF offset contributes to
fat suppression failure at the distal wrist with conventional DTI in all 3
volunteers. RTB0 eliminates this fat suppression failure since it enables the spectrally selective water excitation
pulse to operate at the optimal water peak even at off-isocenter location (Figure 3).
Figure 4 shows the absolute displacement error in AP & LR for datasets
2 to 5. We did not observed improvement in spatial accuracy, and upon further inspection, it could be
due to the differences in X/Y/Z shim values & CF in DTI vs RTB0 series, which
contributed to stretching, skewing and shifting of the image. In future
experiement, we should fix the CF & shimming parameters to eliminate this
confounding factor.
Figure 5 shows a 3D rendering of the peripheral nerve around a nerve
sheath tumor at the wrist in a patient. Note that nerve bundles were best tracked in
the dataset with both RTB0 & ADDC applied in this case, showing that ADDC
& RTB0 could potentially help with visualization of nerve bundles.
Conclusion:
In this work, we demonstrated fat
ghosting reduction using the RTB0 acquisition method. And in our initial
investigation, the combination of RTB0 & ADDC method does not provide
significant difference or improvement in alignment in this small SI coverage scenario.
Future investigation can look at larger SI coverage & comparison with fixed
shimming parameters to better assess the spatial alignment differences.
Acknowledgements
No acknowledgement found.References
[1] Naraghi el al, Semin
Musculoskelet Radiol 2015,19:191–200. [2] Fung et al, ISMRM 2015 Proceeding,
1606, [3] Jenkinson et al, NeuroImage 2002, 17(2), 825-841, [4]Klein et al, IEEE
Transactions on Medical Imaging 2010, 29(1):196
- 205.