Emad Ahmadi1, Reza Atefi1, Emad Eskandar2, Alexandra J. Golby3, Michael H. Lev1, Rajiv Gupta1, and Giorgio Bonmassar1
1Radiology, Massachusetts General Hospital, Boston, MA, United States, 2Neurosurgery, Massachusetts General Hospital, Boston, MA, United States, 3Neurosurgery, Brigham and Women's Hospital, Boston, MA, United States
Synopsis
Electrocorticography
grids are routinely implanted over the cortex for pre-surgical planning in
epilepsy surgery. We propose that MR imaging at 3T might cause heating injury in
patients with implanted electrocorticography grids.Purpose
To study
the MR heating caused by conventional electrocorticography grids
Methods
We placed an electrocorticography
grid (FG64C-SP10X-000, Ad-Tech Medical, Racine, WI) over a head phantom with MR
thermodynamic properties similar to human tissues and measured the temperature
increase around the grid during 30 minutes of MR imaging (Figure 1) (1). For comparative purposes we also measured
temperature increases for the head phantom with no grid. A turbo-spin echo sequence was used on a 3T Siemens Trio
scanner to deliver a specific absorption ratio of 4 W/kg to the phantom. Temperature
changes were measured using eight optical sensors that
were placed within the phantom underneath the grid. To place each sensor in its location, a tunnel was
drilled inside the phantom and the sensor was passed
through the tunnel. The air gap between sensors and the tunnel walls was filled
with thermal conductive grease.
Temperature changes
during MR imaging were measured for the two cases of the phantom with and
without the grid. These
two experiments were done on two separate days. We kept the phantom in the MR
scanner room for 48 hours before each experiment to allow the phantom
temperature to reach to equilibrium with the scanner room temperature. Each
temperature increase measurement was followed by a re-test measurement using
the same procedures. This was done to ensure the reliability of the measurements.
Results
Table
1 shows sample temperature increases for the head phantom after 15 and 30
minutes of MR imaging for the head phantom without and with the
electrocorticography grid. Table 2 shows the re-test results. Figure 2 shows
the temperature increases
Conclusion
The MR
heating of the head phantom with the conventional grid was more than twice as
much of the phantom without the grid, which may lead to heating injury in
patients with electrocorticography grids undergo MR imaging.
Acknowledgements
This
research was supported by the grants 5R43NS071988-02 (NIH/NINDS), U01-NS075026 (NIH/NINDS),
1R21EB016449-01A1 (NIH/NIBIB), the National Center for Research Resources (NCRR,
P41-RR14075), the National Center for Image Guided Therapy (NCIGT,
P41EB015898), and by the MIND institute. References
1. Angelone LM, Makris N, Vasios CE, Wald
L, Bonmassar G. Effect of transmit array phase relationship on local Specific
Absorption Rate (SAR). ISMRM Fourteenth
Scientific Meeting. Seattle, USA2006.