Chi Ma1, Krzysztof R Gorny1, Christopher P Favazza1, Robert E Watson1, and Heidi A Edmonson1
1Radiology, Mayo Clinic, Rochester, MN, United States
Synopsis
Exclusion of scanning with transmit RF body coil may
prohibit access to life-saving diagnoses for patients with MR-conditional
implantable devices. Manufacturer provided plots of RF B1-field indicate that
RF energy over the implant may be significantly reduced if the implant is kept
outside of the 50-55cm long transmit RF body coil. Scanner-specific B1-field measurements and
RF-induced heating measurements confirm reduction in heating as conductive
material moves away from scanner isocenter.
B1-field measurements lateral to the central scanner axis demonstrate
local peaks in the B1-field that would not be identified from the IEC-required
manufacturer plots.Introduction
Manufacturers of many implantable active devices provide
conditions for safe MRI scanning that exclude use of the transmit
radiofrequency (RF) body coil. This limits
the types of MRI exams available to patients with such devices, possibly prohibiting
access to life-saving diagnoses. However,
the stray RF field rapidly decreases outside of the body coil
1, and the
typical 50-55 cm electrical length of the body coil is significantly shorter
than the 105-130 cm MR scanner bore length. A conservative MR exam using body coil
transmit might have low theoretical risk of implant heating if centering the
imaged anatomy places the implant sufficiently far from isocenter, beyond the
location of an approximate 10 dB reduction in B1-power
2. IEC60601-2-33 requires manufacturers to
provide locations along the z-axis where the RF transmit field is reduced by 3
and 10 dB
1 . Here we set out to measure the RF B1-field on
a specific MRI scanner for comparison against the manufacturer-provided data,
as shown in Figure 1
3. The ultimate goal is to better inform clinical risk/benefit
assessment for patients with MR-conditional implanted devices that exclude use
of the body transmit coil, but otherwise allow for MR exams with volume head
and extremity transmit coils.
Methods
MRI measurements
were performed on a 1.5T Optima 450W scanner (GE Healthcare, Milwaukee, WI). Transmit
gain (TG) for a 90° pulse was set to 150, a value commonly seen for normal BMI
patients on this particular scanner. To map
changes in the transverse RF B1-field relative to isocenter, induced voltages
in a single-loop RF pick-up coil tuned to 64 MHz (1.5T) were measured using an
oscilloscope; the voltages induced by an unloaded body coil were measured with
the pick-up coil oriented in a sagittal and coronal plane at 10 cm increments
along the z-axis, as described in IEC60601-2-331. B1-field was also mapped along a line shifted
15 cm lateral to the z-axis to correspond to a typical position of an
Implantable Pulse Generator (IPG) and associated leads. B1-field measurements were repeated with a
body TLT sphere and body-coil loader phantom within the RF coil.
Reduction
in the RF field should correspond to reduction in RF-induced heating. To demonstrate this, the temperature rise was
measured at the lead tips of insulated copper leads placed in a head-and-torso
ASTM phantom (ASTM F2182-11) filled with gelled saline4 . Two 18-gauge, 15 cm long wires, parallel to the scanner z-axis, were submerged in the phantom, as
depicted in Figure 2 A. Fluoroptic temperature probes (STF-2, model 750;
LumaSense Inc., Santa Clara, CA, USA) were positioned at both ends of each of
the wires. An RF spin echo with 4-echoes, TR/TE1/TE2/TE3/TE4
100/17/34/51/68 ms, with an estimated SAR of 1.2 W/kg (input weight=150 lbs) was
scanned for 259 seconds to generate lead heating. The phantom was initially landmarked and
scanned with the middle of the wires at the center of the RF transmit coil, and
subsequently at 10 cm increments as the phantom was advanced out of the RF
transmit coil. The maximum temperature
increase for each wire was recorded as a function of landmark location.
Results
B1-field power and max temperature elevations as a function of
distance along the z-axis from isocenter are shown in Figure 2B. Along
the Z-axis, B1-field power falls below 10 dB at a distance of 30 cm from
isocenter. At the 40 cm lateral location from isocenter, the B1-field power has
a small, reproducible, local peak, at 13%/9 dB reduction relative to isocenter. This
power increase closer to the bore wall is consistent with B1-field modeling of
a high-pass birdcage transmit coil.
Temperature
measurements show a marked reduction in RF-induced heating: temperature
elevations were measured to be 3% of those at isocenter when the wires were
positioned at a landmark location 40cm from isocenter. Again, these temperature
measurements were for demonstration purposes only. Actual temperature increases would be
dependent on device, patient, lead path, scan location and scan technique.
Conclusions
RF
power over an implanted device, and the resulting heating of said device, are expected
to drop dramatically when the implanted device is located far away from magnet isocenter.
Our scanner specific measurements demonstrated localized B1-field power
increases near the end-rings of the body coil, information that would not be
immediately apparent from the manufacturer’s provided data. The scanner specific measurements provide
information that can help with conservatively scanning and positioning a
patient for whom risk/benefit analysis favors proceeding with MRI despite the
implanted device’s exclusion of body coil transmit from its MR conditions for
use.
Acknowledgements
No acknowledgement found.References
1. International
Electrotechnical Commission, Medical electrical equipment –Part 2-33:
Particular requirements for the basic safety and essential performance of
magnetic resonance equipment for medical diagnosis, 2013
2. Kanal E, MR Safety Course: MR Medical
Director/MR Safety Officer, Las Vegas, NV 06/23/2015 and Orlando, FL
10/20/2015.
3. OptimaTM MR450w 1.5T 25.0 English Operator
Manual/Training Tool, GE Healthcare, p. 214, Figure 2-6.
4. Favazza C, King D,
Edmonson H, et al. Use of a radio frequency shield during 1.5 and 3.0 Tesla
magnetic resonance imaging: experimental evaluation, Medical Devices: Evidence
and Research 2014; 7:363-370.