Design and simulation of a nested 4 channel 1H and 3 channel 13C coil for glycogen NMR experiments in the calf muscle at 7 T
Sigrun Goluch1,2,3, Roberta Kriegl2,3, Elmar Laistler2,3, Martin Gajdošík 4,5, and Martin Krššák 1,4,5

1Division of Endocrinology and Metabolism, Department of Internal Medicine III, Medical University of Vienna, Vienna, Austria, 2MR Center of Excellence, Medical University of Vienna, Vienna, Austria, 3Center for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria, 4High-Field MR Center, Department of Biomedical Imaging and Image-Guided Therapy, Medical University of Vienna, Vienna, Austria, 5Christian Doppler Laboratory for Clinical Molecular MR Imaging, Medical University of Vienna, Vienna, Austria

Synopsis

Due to the inherently low sensitivity of carbon-13 NMR, 13C spectroscopic experiments at 7T require specifically optimized double tuned local RF transceive arrays for high SNR, exhibiting sufficient electrical isolation between the arrays to enable 1H decoupling and high SAR efficiency as to not invoke SAR limits during proton decoupling. In this work we present the simulation and optimization of a 7 channel nested 1H/13C RF transceive coil array for 13C metabolic studies in the human calf muscle at 7 T.

Purpose

Carbon-13 MRS is a validated method for the assessment of glucose metabolism in skeletal muscle. It can be used together with an exercise challenge to uncover the dynamics of glycogen stores, or to study glucose/lipid oxidation in different pathologies. However, as many other NMR feasible nuclei, 13C NMR suffers from intrinsically low sensitivity, which is further decreased due to the 1H-13C hetero-nuclear J-coupling. To enhance its overall sensitivity, transmission on proton frequency during 13C reception can be employed (proton decoupling). Another possibility to gain sensitivity is to move to higher static magnetic field strength, resulting in further increase of RF power deposition posing some limitation to 1H decoupled 13C NMR experiments. Therefore, 13C NMR at 7T requires specifically optimized double tuned local RF transceive arrays for high SNR, exhibiting sufficient electrical isolation between the arrays to enable 1H decoupling and high SAR efficiency as to not invoke SAR limits during proton decoupling. In this work, we present the design for a 7 channel nested 1H/13C RF transceive coil array for 13C metabolic studies in the human calf muscle at 7T.

Methods

Coil Design: The coil array consists of a 4 channel 1H array atop of a 3 channel 13C configuration. The channels are mounted on an half-cylindrically shaped ABS former with an outer diameter of 15 cm. The distance between the 13C and 1H arrays was chosen to be 1 cm. Previously built 13C phased arrays [1,2] often employ so-called LCC traps in order to decouple the concentrically arranged channels, in doing so, incorporating additional loss mechanisms and decreased Q-ratios [3]. By choosing 3 and 4 channels, respectively, we are able to geometrically decouple both arrays, while maintaining a sufficiently large FOV and high SNR. The element sizes are dictated by 3 variables, namely the overall length of the 1H array, and the optimum overlap between next nearest 1H and 13C channels, respectively. The overall length of the 1H array was chosen to be 24 cm, which covers almost the whole half cylinder. Depending on the overlap factor for 1H (olH), this yields a 1H element size of w1H=24/(1+3*olH). In order to remain geometrically decoupled with the 1H array, the element size of the carbon array is dictated by the choice of the overlap factor (olC) and the fact that olH*w1H=olC*w13C. A schematic of this setup can be seen in Fig. 1.

Simulation: The coil design was realized within XFdtd 7.4 (Remcom, State College, PA, USA) using 2 mm thick coil wire, modeled as perfect conductor. It was loaded with a cylindrical phantom filled with tissue-like gel, as well as Ella’s calf (Virtual Family). All capacitor gaps were connected with 50Ω voltage sources to enable RF co-simulation in ADS (Agilent Technologies) [4]. Inductive counter-wound decoupling was realized between all next-nearest neighbors in co-simulation. Realistic loss estimations for inductances, capacitances, and solder joints were modeled as series resistances [5]. The optimal overlap factor was determined beforehand via simulation of 2 channels with varying overlap Static B1+ shimming was achieved using the SimOpTx toolbox (Medical University of Vienna, Austria) for both arrays by maximizing SAR efficiency (B1+/√SAR10g, max) in the region of interest (gastrocnemius muscle, depicted in black in Fig. 3) as a function of the relative channel phases, incremented in 5° steps for 13C (5184 phase sets) and 10° steps for 1H (46656 phase sets).

Results

The optimized overlap factor for the 1H array was 0.8, yielding an element size of 7.1x12.5 cm2, and 0.9 for the carbon array, resulting in an element size of 6.2x10 cm2. Decoupling between coil elements was below -14.5 dB and -14.4 dB for the 1H and 13C array, respectively. Cross-coupling was always below -29.4 dB at the proton Larmor frequency (297.2 MHz) and -27.9 dB at the carbon Larmor frequency (74.7 MHz), indicating good geometric decoupling (see Fig. 2). Optimized static B1+ shim could be achieved with a phase set of [120°/40°/-40°/-120°] for 1H and [50°/0°/-55°] for 13C. B1+, SNR and SAR maps can be seen in Fig. 3. The maximum 10g averaged SAR for 1 W input power is 1.5 W/kg for 1H and 1.3 W/kg for 13C.

Conclusion

The design and simulation results for a nested 7 channel 1H/13C transceive array, aiming at high SAR efficiency are presented. The shifted setup results in good decoupling between 1H and 13C channels, and therefore seems a promising candidate for a coil enabling proton decoupling to further increase 13C SNR in metabolic studies of the human calf at 7T. The physical implementation of the coil is in progress.

Acknowledgements

This work was supported by the OeNB Anniversary Fund (grant number P 15363).

References

[1] Donati et al, ISMRM 2015; [2] Roig et al, MRM 2015; [3] Meyerspeer et al, MRM 2013; [4] Kozlov at al, JMR 2009. [5] Kumar et al, MRM 2009.

Figures

Fig.1. (Left) 2D schematic of the proposed RF coil design. Red: 4 channel 1H array, blue: nested 3 channel 13C array. Element sizes were chosen according to the formulas. (Right) 3D view of the nested RF coil array, with the overlap-optimized element sizes of 7.1 cm and 6.2 cm for 1H and 13C elements respectively.

Fig.2. Full S-parameter matrix at 74.7 MHz (upper triangle) and 297.2 MHz (lower triangle) for the phantom load (top) and voxel model (Ella) load (bottom), respectively.

Fig.3. Performance of the 3 channel 13C array (74.7 MHz), from left to right: B1+ over square root of power, signal to noise ratio and the maximum intensity projection of the SAR distribution. Top row are the results for the phantom load, bottom row are the results with Ella’s calf as load. In both cases, the same static B1+ shim set was used. The black half-moon shaped area is the ROI corresponding to gastrocnemius muscle.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2229