Chi Ma1, Zaiyang Long1, Diana M Lanners1, Donald J Tradup1, Joel P Felmlee1, David A Woodrum1, Nicholas J Hangiandreou1, and Krzysztof R Gorny1
1Department of Radiology, Mayo Clinic, Rochester, MN, United States
Synopsis
The suitability of a compact
Samsung ultrasound (US) system for real-time imaging guidance of treatment
device positioning inside 1.5T interventional magnetic resonance imaging (iMRI)
suite was assessed. The US system was tested in a proposed site-specific
configuration. Magnetic displacement forces exerted by the static magnetic
field on each of the US system components were estimated at the proposed
locations. Image quality of both MRI and US systems with the US system set to
different operating modes were evaluated.
Results demonstrate that this particular US system is suitable for use
in the site-specific configuration at our 1.5T iMRI suite. Introduction
MRI-guided interventional
procedures are gaining increasing clinical acceptance
1,2.
Positioning of treatment devices under real-time MRI is challenging and time
consuming due to tight space inside MRI scanner bore. Compact Ultrasound (US)
imaging system provides an option for real-time guidance because of its small
foot print and low amount of magnetic components. The purpose of this study was
to define and test site-specific conditions for safe use of a compact Samsung UGEO
model HM70A US system in an interventional magnetic resonance imaging (iMRI)
suite hosting a 1.5T Siemens Espree MRI scanner. The investigation also included evaluation of the
effects of the presence of US system in the iMRI suite on both MR and US image
quality.
Methods
The proposed configuration for
clinical use included placement of the US scanner with transducer connector at
the foot of the MRI table and the transducer to be operated at the opening of
the MRI scanner bore (Figure 1). During
MRI scanning, the US scanner and transducer were repositioned onto an MR safe
stool at the foot of the table. A series of tests were performed to compare magnetic
displacement (MDF) and static friction (SFF) forces exerted on each of the US
system components at the proposed locations to assess the likelihood of
displacement towards the bore.
Components that
did not exhibit any magnetic attraction under a hand-held magnet test were positioned
on a smooth Plexiglas surface, and moved to proposed location to directly
assess if field-induced displacement occurs.
For US system
components displaying magnetic attraction, a separate measurement protocol,
listed below, was followed, to obtain a conservative estimate of the ratio, R,
of the SFF to MDF at the proposed location,zloc.
1. MDF exerted
by the hand-held magnet on US system components and a steel wool test pad were measured using a scale outside iMRI suite.
2. Magnetic
field strength along the scanner z-axis
inside the iMRI suite was mapped using a Gauss meter and the field
gradient at each location was derived, gradientB(z).
3. Steel
wool test pad was used inside iMRI suite, to determine the location zdisp where the pad starts to get displaced by
B0 of MRI scanner. (Figure
2)
4. Using
data from the above experiments, R was calculated as follows,
$$R=\frac{F_{test\ pad}}{F_{US}}\times\frac{m_{US}}{m_{test\ pad}}\times\frac{gradient_{B}(z_{disp})}{gradient_{B}(z_{loc})}$$
where FUS and Ftest pad are MDF exerted by the hand-held magnet on the US system component
and the test pad, respectively, and mus ,mtest pad are
their masses. A minimum R value of 10 is set as the criteria for the
tethered US system components to be safely advanced towards the proposed
location.
Image quality (IQ) tests of MRI
system in the proposed configuration were evaluated with the US system in
different operating modes, using an ACR MRI accreditation phantom to assess MRI
IQ degradation caused by the presence of the US scanner. Similar US IQ
assessments were conducted to detect any field-induced degradation of US image
quality.
Results
Using a hand-held magnet, US
scanner, transducer connector and A/C adapter demonstrated attraction, while
transducer scan head did not. Displacement tests for the transducer scan head demonstrated
no displacement at scanner bore opening. Magnetic field strength measurements are
shown in Figure 3. Estimated R ratios for US scanner, transducer
connector, and A/C adapter were found to be 1063, 66, and 3850, demonstrating
the SFF to be at least an order of magnitude greater than MDF at the proposed
locations, which met our criteria. The US scanner, transducer connector and A/C adpater were
subsequently tethered and advanced towards the proposed locations inside iMRI
suite, at which no tendency towards magnetic displacement was found.
MR image quality met the American
College of Radiology (ACR) standards3 for clinical scanning. Increases
in noise levels and RF interference of varied degrees were observed, depending
on operating mode of the US scanner. The “sleep mode” was identified as having been
associated with absence of RF interference and minimal noise level increases (Figure 4). This mode allows the physician
to begin and end US scanning quickly, without significant delay. At the
proposed locations, no clinically-significant impact in US image quality was
observed.
Conclusion
Our results suggest that the Samsung
UGEO model HM70A US system is suitable for use in our well-defined site-specific
configuration inside the iMRI suite. Following the tests presented above, the
US system has been successfully used in our clinical practice for image
guidance of treatment device positioning (Figure
5) during MRI-guided intervention procedures.
Acknowledgements
No acknowledgement found.References
1. Bomers, J.G., J.P. Sedelaar, J.O. Barentsz, et al., MRI-guided interventions for the treatment
of prostate cancer. AJR. American journal of roentgenology, 2012. 199(4): p. 714-20.
2. Woodrum, D.A., A. Kawashima, R.J. Karnes, et al., Magnetic resonance imaging-guided cryoablation
of recurrent prostate cancer after radical prostatectomy: initial single
institution experience. Urology, 2013. 82(4):
p. 870-5.
3. Radiology., T.A.C.o., Phantom
test guidance for the ACR MRI accreditation program. Reston (VA), USA; ,
2005.