Hai Luo1, Bin Wang1, Gaojie Zhu1, Wenzhou Wang1, Xiang Zhou1, Ziyue Wu1, and Leping Zha1,2
1AllTech Medical Systems, Chengdu, China, People's Republic of, 2AllTech Medical Systems, Cleveland, OH, United States
Synopsis
3D dual echo gradient echo sequence is commonly used
to obtain the field map for B0 shimming. The maps contain the true B0 fields mixed
with eddy currents induced magnetic field changes averaged over the echo time
difference, which compromise the shimming accuracy. A calibration sequence with
alternating gradient polarities is proposed to measure the eddy currents term. Quadratic surface fitting is then applied to produce smooth eddy currents
calibration maps over the full imaging volume containing only the first and
second order components. The actual rapid in-vivo shimming sequence runs later,
using the calibration maps to remove the eddy currents influences during
the post-processing, with partial Fourier acquisition
on phase encoding and slice encoding directions to reduce the scan time. The fast
method provides means of eddy currents insensitive shimming, as well
as reduced sensitivity to motion.Purpose
To develop a method of rapid B0 field estimation
for robust and accurate B0 field shimming, which is insensitive to the eddy currents.
Introduction
Chemical
shift based fat suppression techniques as well as spectroscopic applications do
not work well without proper B0 shimming. 3D gradient echo (GRE) sequences acquiring a
pair of in-phase echoes were proposed and widely used to measure the B0 map
1, in which the common effects of non-static field induced by eddy currents
2 lead to inaccurate shimming. We propose a robust method of rapid B0 field
map estimation, with a single calibration scan in which a full gradient
polarity reversal is used to estimate the influence of eddy currents, before
the sub-sequential in-vivo mapping scans and the eddy current induced error correction in the post-processing.
Method
Fig.1(a) depicts a routine GRE dual-echo sequence for raw (uncorrected) B0 field
mapping, with the so called “fly-back” unipolar gradient readouts to reduce
phase inconsistencies due to effects such as gradient delay and eddy currents. Phase
difference
of the two echoes is the B0 field plus the eddy currents
induced field
Beddy encoded over the echo time difference
ΔTE. As well known, the polarity of
Beddy will be flipped if all of the applied
field gradients are reversed, as shown
in Fig.1(b).
The phase difference between images acquired in the
two parts of the calibration scan can be written as$$\Delta\phi_{positive}={\gamma}B_{0}\Delta{TE}+{\gamma}B_{eddy}\Delta{TE}{\hspace{4cm}}(1)$$ $$\Delta\phi_{negative}={\gamma}B_{0}\Delta{TE}-{\gamma}B_{eddy}\Delta{TE}{\hspace{4cm}}(2)$$ From which
Beddy can then be calculated as $$B_{eddy}=(\Delta\phi_{positive}-\Delta\phi_{negative})/2\gamma\Delta{TE}\hspace{3.3cm}(3)$$In a given MR scanner, the eddy currents patterns are determined by the hardware material and structure, and system tuning
including the eddy currents compensation (ECC), while the fast varying field
changes form the patients are basically negligible. Thus, the
Beddy field distribution measured on a regular sphere phantom in the calibration
scan, and later extrapolated onto the full imaging volume can be considered the
same as in the in-vivo shimming.
The eddy currents calibration combining sequences in Fig.1(a)
and Fig.1(b) runs once, before running the actual B0 mapping sequence in Fig.1(a)
in vivo. The
Beddy map calculated through (1)–(3) then goes through a 3D phase unwrapping algorithm
3, decomposed into 0
th and higher spatial order terms, while in general the 3
rd order and above are negligible. Surface fitting with quadratic basis
function is thus generally enough to obtain smooth eddy currents induced field maps over the full imaging space
B’eddy. Finally, the extrapolated and refined
B’eddy estimation is applied to correct the raw B0 field maps for shimming$$B_{0}=(\Delta\phi_{positive}-{\gamma}B_{eddy}\Delta{TE})/\gamma\Delta{TE}\hspace{3.9cm}(4)$$
Results
The B0 mapping sequences and the processing procedure were implemented on an Alltech
EchoStar 1.5T scanner (AllTech Medical Systems, Chengdu, China). Fig.2(a)
shows the center slice of a measured 3D
Beddy distribution on a 190mm sphere
water phantom. The sequence parameters were TR/TE1/TE2/θ=13.4ms/4.5ms/9.0ms/20°, with a 320mm cubic
FOV and a matrix size of 32x32x32, zero-filled
to 64x64x64. The ECC was turned off to make the comparisons
more obvious. Fig.2(b) shows the
B’eddy correction map from Fig.2(a) through phase unwrapping and quadratic fitting, with the 0
th term discarded as it does not affect shimming.
To validate the method’s effectiveness, a proton
density (PD) weighted FSE sequence with chemical shift selective (CHESS) fat
saturation pre-pulse was acquired for pelvic imaging. The imaging parameters were
TR/TE=1800/27ms, ETL=5, FOV380x300mm, acquisition matrix size316x188, and 6mm slice thickness. Only the
sequence in Fig.1(a) run in-vivo for shimming, with a partial Fourier ratio of 0.85 in the both phase encoding (PE) and the slice encoding (SE)
directions. The total duration of shimming is about 10 seconds. Fig.3 shows the
resulting images after
applying the shimming results, with (a) and without (b) corrections of the
B’eddy map. Fat signal in Fig.3(a) was almost fully saturated, while in Fig.3(b) there are residue
non-saturated fats and unintended water suppression from the insufficiently homogeneous B0 distribution.
Discussion
The proposed calibration method with full
gradient polarity reversal estimates the eddy currents induced field
distribution with reasonable accuracy, in practical situations from our
experiences. It was found that with matching calibration and shimming sequence
parameters, the B’eddy patterns stay essentially the same.
Moreover, translations of the shimming FOV do not change the gradient
waveforms, so that the B’eddy map is not
changed. Therefore, multiple shimming protocols with changing shimming volume
definitions can share the same B’eddy correction map. In summary, the proposed B0 mapping procedure provides an
eddy currents insensitive method without increasing the in-vivo shimming time
from the normal mapping methods. This also reduces errors from motion induced
phase changes because of the relatively short mapping time.
Acknowledgements
No acknowledgement found.References
[1] Kanayama et al., MRM 36 : 637-642 (1996), [2] Bartusek et al., Meas. Sci. Tech. 21 : 105601 (9pp., 2010), [3] Cusack et al.,
NeuroImage 16 : 754-764 (2002).