Ryota Yamada1, Makoto Tsuda1, Katsumi Kose1, and Yasuhiko Terada1
1Institute of Applied Physics, University of Tsukuba, Tsukuba, Japan
Synopsis
A
multi-circular shim coil (MCSC), which consists of a set of localized circular
current coils, provides the flexibility to design and produce linear and
higher-order magnetic fields that compensate for a given B0
imhomogeneity both statically and dynamically. However, the concept of the MCSC
has currently been realized only for cylindrical base geometries. Here we
translated the concept of the MCSC to a biplanar geometry, and a planar-type
MCSC was designed and fabricated for an open, 1.0 T permanent magnet system. We
concluded that the planar MCSC is a useful devise to achieve field homogeneity
with reasonable accuracy.INTRODUCTION
A multi-circular
shim coil (MCSC) [1,2], which consists of a set of localized circular current
coils, provides the flexibility to design and produce linear and higher-order
magnetic fields that compensate for a given B
0 imhomogeneity both
statically and dynamically. However, the concept of the MCSC has currently been
realized only for cylindrical base geometries. Here we translated the concept
of the multi-circular shimming to a biplanar geometry. To test the concept, a
planar-type MCSC was designed and fabricated for an open, 1.0 T permanent
magnet system. We verified the validity of the planar MCSC.
MATERIALS AND METHODS
Figure
1 shows the MCSC and MRI systems. The planar MCSC consisted of circular coils (24
coils, wire diameter 0.26 mm, 3 turns for each coil) fixed on two fiber-reinforced
plastic plates (20 cm × 20 cm, thickness 0.5 mm) (Fig. 1(a)). There is a
trade-off between the shimming performance and the maximum coil current,
depending on the center diameter of circular coils (Fig. 2). Here the center diameter
was determined to be 30 mm. Then the MCSC was fixed on the home-built coplanar
gradient coils (Fig. 1(b)). The solenoidal coil (diameter 55 mm, 60 mm long) and
the RF shield box (22 cm × 18 cm × 7 cm) made of the brass plates were inserted
into the gradient coil and MCSC set. The magnet used was a Halbach 1.0 T
permanent magnet (Fig. 1(c); NEOMAX Engineering, Tokyo, Japan; gap 90 mm,
homogeneity 7 ppm over 4 cm diameter spherical volume (DSV), weight 980 kg).
The
field imhomogeneity over 4 cm DSV at the center of the magnet was corrected
using MCSC as follows. The spatial variation of the magnetic field ΔB0 was
measured using a conventional phase shift method with a CuSO4–doped
water phantom. Then the constrained leased-squared fitting based on the
Levenberg-Marquardt method was applied to decompose the measured field
variation into the fields generated by the circular coil elements and to
determine the coil currents that were necessary to correct the field variation.
The maximum coil current was limited to 1 A. The coil currents were generated
by a home-built 24ch current power supply and controlled by a home-written C
software run on a laptop windows PC (Fig. 1(d)). For comparison, conventional
second-order shim coils (xy, yz,
xz, z2, and x2-y2) were fabricated.
RESULTS AND DISCUSSION
Figure 3
shows the center slices of ΔB0
measured with and without the shimming. The field imhomogeneity was largely
compensated for by the planar MCSC with the maximum coil current of 390 mA. Without
shimming, the root mean square (RMS) and peak-to-peak (PP) ΔB0
were 0.91 and 8.2 ppm, and these were largely reduced with the planar MCSC
(RMS: 0.56, PP: 4.8 ppm). The measured RMS and PP values with the MCSC were
close to the theoretical values (RMS: 0.41 ppm, PP: 4.0 ppm). The performance
of the MCSC shimming was compatible with the conventional second-order shimming
(RMS: 0.38 ppm, PP: 3.7 ppm). Figure 4 shows the capability of inhomogeneity
correction with high-order terms. The theoretical field maps generated by the
MCSC agreed with the ideal second and third-order spherical harmonics. The
errors were acceptable for the second order terms. They were not small for the
third-order terms, but were still acceptable for the practical use. More
accurate field correction may be achieved by increasing the maximum coil
current or by optimizing the configuration of the circular coils.
In
conclusion, the planar MCSC is a useful devise to achieve field homogeneity
with reasonable accuracy.
Acknowledgements
No acknowledgement found.References
[1]
C. Juchem et al. Magnetic field modeling with a set of individual localized
coils, J. Magn. Reson. 204 (2010) 281-289.
[2] C. Juchem et al. Dynamic
multi-coil shimming of the human brain at 7 T, J. Magn. Reson. 212 (2011)
280-288.