Removing Gradient Induced Voltages from 12-lead ECGs acquired during DW-EPI and fMRI brain Imaging
Mikayel Dabaghayan1, Shelley Hua Lei Zhang1, Zion Tsz Ho Tse2, Charles L Dumoulin3, Ronald Watkins4, Wei Wang1, Jay Ward5, and Ehud Jeruham Schmidt6

1Radiology, Brigham and Womens Hospital, Boston, MA, United States, 2Engineering, University of Georgia, Athens, GA, United States, 3Radiology, Cincinatti Childrens Hospital Medical Center, Cincinatti, OH, United States, 4Radiology, Stanford University, Stanford, CA, United States, 5E-Trolz Inc., North Andover, MA, United States, 6Radiology, Brigham and Womens Hospital, Newton, MA, United States

Synopsis

We developed a technique to restore the ECG signals distorted by MRI gradient-induced voltages (GIV) acquired during fMRI and DW-EPI brain imaging sequences. Brain EPI sequences produce the largest ECG artifacts, presenting a large challenge to GIV removal. We used a theoretical equation with 19 parameters, which characterized the GIVs at each ECG electrode based on the simultaneously recorded gradient waveforms. A rapid training sequence permitted computing the equation coefficients, followed by real-time gradient-induced voltage removal during imaging. FIR notch filters were subsequently applied to remove some residual spikes. The method succeeded in removing most GIVs, excluding artifacts at the beginning and end of imaging periods, which resulted from amplifier non-linearity.

Target Audience Neurologists, radiologists and scientists interested in obtaining high-fidelity 12-lead ECGs for physiological monitoring during diagnostic or interventional brain functional (fMRI) or Diffusion-weighted (DW-EPI) EPI studies.

Purpose ECG traces inside MRI bore are distorted by both the magneto-hydrodynamic effect and by strong (~5V) MRI gradient- induced electric fields. We previously developed and validated [1] a prototype MRI-compatible 12-lead ECG platform that utilized a theoretical equation with 19 parameters to remove gradient-induced-voltage (GIV) artifacts created during high-gradient-duty-sequences (SSFP, FSE, GRE) on 12-lead ECGs. The system acquired the 12-lead ECG traces, along with the three gradient waveforms that produce the GIV. A training procedure (duration 3-6 seconds) acquired ECGs during periods without gradients, followed by parallel-imaging accelerated sequences, to determine the 19 equation parameters. Thereafter, GIVs were removed in real-time from ECGs acquired during the imaging sequences. When the 12-lead ECG electrodes are not positioned close to magnet iso-center, such as during brain imaging, the electric fields induced in the torso are larger [2], so the GIVs induced on ECGs traces can be 3-4 times larger than observed in cardiac/abdominal imaging. Additionally, during DW-EPI’s diffusion-encoding the highest gradient strengths are employed, while during both fMRI and DWI read-out gradients are modulated rapidly at the scanner’s maximal slew rate, producing larger GIVs than other sequences. In this study, we test the ability of the theoretical method [1], together with an improved acquisition system, to remove GIVs in this most-difficult case. DW-EPI and fMRI are used clinically to diagnose patients undergoing acute stroke and brain seizures in MRI-equipped emergency rooms, as well as detect tumor and fiber location during MRI-guided brain surgery, so the availability of on-line diagnostic-quality 12-lead ECG monitoring can be critical. Additionally, ECG-gating is increasingly used to acquire more repeatable EPI-based diffusion and perfusion parameters.

Methods A theoretical equation for the gradient-induced voltages on each ECG electrode was derived [1], which depends on the three gradients and their time derivatives. It has 19 terms, and requires the determination of 19 parameters. The ECG traces were collected using a commercial GE (Waukesha, WI) CardioLab system. In this study, we used the Cardiolab catheter input channels, instead of the surface ECG channels [1], thus preventing Cardiolab’s Right-Leg voltage drive from sending out strong voltages to the subjects, and enabling measurement of the GIVs induced on each limb lead independently. We used an amplification value of 50 to prevent amplifier saturation, high-pass filters of 0.5 Hz to prevent respiratory motion and 1st-order low-pass filters of 150 Hz. The training protocol was identical to that used previously [1], acquiring ECG traces during GRAPPA=5-6 accelerated single-slice EPI sequences, followed by acquisition of ECGs during 3 QRS cycles without gradient activity. Training determined the 19 equation parameters. Since some strong GIV spikes remained in the training data after “cleaning”, we developed a method to remove them by Fourier-transforming the cleaned traces, determining the spike characteristic frequencies (all >50 Hz), and designing notch filters to remove signals at those frequencies whose amplitude was above a threshold (~15mV). Full resolution multi-slice EPI imaging (GRAPPA=2) was performed after the training protocol, with real-time subtraction of the computed gradient-induced voltages based on the 19 coefficients and the simultaneously recorded gradient waveforms, followed by the notch IIR filters. Healthy volunteer images were acquired on a 3T Siemens Skyra (Gradients: 45mT*m-1, 200 T*m-1*sec-1 slew rate). fMRI sequence parameters: TR/TE/flip=700ms/51ms/900, 102x128,epi factor=51,GRAPPA=2, 27 cm fov, 6 mm slice, 10 slices/TR, BW=2170 Hz/pxl, DW-EPI parameters: B==0 and1000 s1mm-2, TR/TE/flip=800ms/101ms/300, 192x192,epi factor=71, GRAPPA=2, 20 cm FOV, 6 mm slice, 6 slices/TR, BW=2170Hz/pxl. 12-lead ECG electrodes were placed at standard torso locations.

Results Figure 1 shows accelerated (training) and the dual-echo DW-EPI images. Figure 2A shows 9 independent ECG traces (Precordial leads:V1-V6, Limb leads: Left-Arm, Right-Arm, Left-Leg), acquired during an fMRI sequence, along with the simultaneously-acquired X, Y and Z gradient waveforms, before and after GIV removal. Figure 2B is a magnified view of 4 ECG traces. Raw traces with GIV (red), after cleaning with the equation (Green), and after cleaning with the equation followed by the IIR filter (Blue). >95% of the GIV voltage is removed. Some residual GIV remains at the start and end of each imaging period, due to the amplifier’s non-linear response to the strong GIV signal.

Discussion We demonstrated removal of gradient-induced-voltages from 12-lead ECG traces acquired during demanding brain fMRi and DW-EPI sequences. Availability of on-line 12-lead ECG in the MRI is critical for monitoring acutely-ill patients during neurological diagnosis and MRI-guided therapy.

Acknowledgements

Supported by an industrial grant from E-TROLZ and NIH P41EB015898

References

[1] Zhang SHL et al, MRM 2015 early view June 23. [2] Bencsik M. et al, Phys. Med. Biol. 2007;52, 2337-2353

Figures

Figure 1: Dual-echo DW-EPI (A) Rapidly (6 sec) acquired training data set using single slice acquisition and GRAPPA=6, and (B) subsequent full-resolution acquired imaging data with GRAPPA=2. ECGs were continuously acquired during these sequences. Note that the MRI-compatible 12-lead ECG system does not add RF noise to the images.

Figure 2: 12-lead ECG traces acquired during a brain fMRI scan. (A) 9 independent traces acquired during imaging (left of 11.2 sec time point), relative to traces acquired after imaging (right of 11.2 sec time point) . Lower three rows show acquired gradient waveforms, used to clean the ECGs. Traces shown; raw with high GIV (Green), after cleaning with equation (Red), and after cleaning with the equation and IIR notch filter (Blue). (B) Magnified view of 4 traces, detailing cleaning efficiency.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2196