Bo and SAR calculation for a full-ring human head PET system integrated with an 8-element Birdcage coil at 3T
Md Shahadat Hossain Akram1, Takayuki Obata1, Mikio Suga2, Fumihiko Nishikido1, Eiji Yoshida1, and Taiga Yamaya1

1National Institute of Radiological Sciences, Chiba, Japan, 2Chiba University, Chiba, Japan

Synopsis

Simultaneous PET/MRI system has attracted much because of its both functional and anatomic imaging capability. In our laboratory, we have developed a human head-size PET/RF-coil integrated modality to be used with existing clinical 3T MRI system (Siemens Magnetom Verio). Eight PET detector modules are integrated with a cylindrical 8-element Birdcage RF coil for simultaneous PET and MRI imaging. In the design each detector has been integrated in between two coil elements. RF interference to PET detector circuits affects PET performance. Also noise generated from PET circuits affects MR image quality. For proper simultaneous operation, PET circuits in each detector-module were installed inside a copper-shielded box. But shielding materials very close to RF coil elements and close to imaging region adversely affects MR imaging quality. In this study we performed Bo and SAR calculation of our hybrid system. Though there have noticeable changes in the Bo values, the SAR remains very low.

Introduction

Dedicated PET inserts for the existing MRI systems have got a great deal of interest in the recent years. PET inserts for the brain imaging has got the most attention over the last few years [1-3]. A head size PET insert have several benefits both in terms of cost and image quality: (a) it will be close to the imaging region - means better sensitivity and image resolution; (b) with fewer PET detectors and electronics - means lower cost than the larger diameter body PET ring. In this study we took the challenge of inserting the PET detectors in between the 8 elements of a Tx/Rx Birdcage coil in a plan to make the PET detectors closer to the imaging region. In a previous report in [2], we published the performance of our concept-prototype for 2 oppositely positioned PET modules. Here we have developed a complete cylindrical PET ring with 8 PET detectors integrated with the RF coil.

Methods and Materials

Corresponding electronic circuits of eight detectors were installed inside copper shielded rectangular cubic boxes with the scintillation crystal outside the shield boxes and each box is positioned in between 2 coil elements (rungs). This way the PET detectors were closer to the imaging region with the inner diameter of 26 cm whereas the coil diameter is 27 cm (Fig. 1a-b, Fig. 2). To avoid parallex error of PET modules and to get reasonable image quality even near the PET detectors, we have implemented depth-of-interaction (DOI) detectors with multi-pixel photon counter (MPPCs) and application specific integrated circuits (ASICs).

For the scintillation crystal, we implemented 4-layer depth-of-interaction (DOI) LGSO (Lu1.8Gd0.2SiO5:Ce, Hitachi Chemical Co., Ltd., Japan) detectors. The scintillators were arranged as 12X4X4 matrix (Fig. 1(a)) block in which the dimension of each crystal cube was 2.9 mmX2.9 mmX2.9 mm and the gap between the crystal blocks were filled up with 0.065 mm thick multilayer polymer mirror of 98% reflectivity (Sumitomo 3M, Ltd.). The axial FOV of the PET detectors was 11.73 mm. Two such sets of scintillator detectors were arranged side-by-side as shown in Fig. 1(c-d), and 6 MPPCs (3 MPPCs for each block of scintillators; S11064-050P, Hamamatsu Photonics, K.K.) were connected with DOI scintillator sets to readout the PET signals. Each MPPC of dimension 3 mm X 3mm had 44 readout channels. A weighted sum circuit (WSC) concept was implemented to reduce the total 96 readout channels of 6 MPPCs into 4 channels and amplifier circuits were implemented with the readout circuit board. The readout signals were collected in the data acquisition system outside the MRI magnet room by 10-meter long coaxial cables.

Experiments were conducted in a 3 T clinical MRI system (Siemens Magnetom Verio) by using a homogeneous nickel chloride solution phantom of diameter 20 cm and axial length 11 cm. Bo map was generated by the phase difference of two images with two different echo times - 30 ms and 35 ms [5]. Following the electric properties tomography (EPT) method, we calculated SAR from the B1 field distribution [6]. Also for the B1 field measurements, we implemented double angle method [7] with the following sequence parameters: TR = 3 s, TE = 12 ms, Slice thickness (ST) = 5 mm, 128 X 128 matrix, FOV = 300 mm. For the two images the excitation flip angles were 60o and 120o and, refocusing flip angles were 180o for both.

Results and Discussion

Fig. 3 illustrates the Bo map for the central slices. Units are given in microT. For comparison purpose we have given Bo maps for both the cases of RF coil with and without PET modules. It is clearly seen that due to PET module Bo values increase. At the center slice the inhomogeneity increases from 0.93 ppm without PET to 2.2 ppm due to PET detectors and, at 5 cm off-center it increases from 2.5 ppm to 4.1 ppm. The SAR maps (Fig. 5) for both transverse and coronal plane were calculated from the B1 distribution maps (Fig. 4). The value of local SAR and average SAR was seen much smaller than maximum limit. Change in SAR value for PET detectors were very negligible (from .0021 W/Kg to .0027 W/Kg at the center slice with spin-echo imaging for 3 minutes). We have found a significant change in the Bo field. This can be because of the ferromagnetic elements on the PET circuits and eddy currents in the shielding materials. We are working to further improve the system in near future.

Acknowledgements

No acknowledgement found.

References

[1] A. Kolb, et al, Eur Radiol 22 (2012) 1776–1788; [2] F. Nishikido, et al, Nucl Instru Method Phys Res A 756 (2014) 6–13; [3] P. Olcott, et al, Phys Med Biol 60 (2015) 3459–3478; [4] M. C. Leifer, Magn Reson Med, 38 (1997) 726-732; [5] AAPM Report No. 34, Medical Physics, 19(1992)217-229.[6] U. Katscher, et al, IEEE Trans Med Imag, 28 (2009) 1365-1374. [7] E.K.Insko et al, JMR 103(1993)82-85.

Figures

Fig. 1. (a) The outer and (b) inner view of the full ring PET integrated with a Tx/Rx Birdcage RF coil. (c) The PET circuits are shielded with copper in which the crystals were kept outside the shield box

Fig. 2. Schematic representation of the integrated PET/RF-coil insert for using with the MRI system

Fig. 3. Bo map for the central transverse slice for both with and without cases

Fig. 4. B1 field maps for the central transverse and coronal slices. This field distributions were implemented for SAR calculation by EPT method

Fig. 5. SAR maps for the central transverse and coronal slices. The average SAR is given on top of each map.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2193