Spencer Baird Parent1, William Bradfield Handler 2, and Blaine A. Chronik2
1Medical Biophysics, Western University, London, ON, Canada, 2Physics and Astronomy, Western University, London, ON, Canada
Synopsis
Using finite elements methods, an
investigation of the systematic errors in magnetic field NMR probes is
investigated. A NMR field probe is modeled and the field broadening and field
offset is investigated as a function of the susceptibility of epoxy. It is
shown that susceptibility matching the epoxy drastically reduces field
broadening with a minimal effect on field offset. Additionally the effect of
air bubbles present in cured epoxy is modeled and the results show that for
certain critical regions of the probe the presence of an air bubble can be disadvantageous
to the quality of the field probe. Purpose
The
presence of the copper solenoid RF coil in a NMR magnetometry field probe
creates a susceptibility boundary which induces a broadening of the field in
the region of interest.
1 Susceptibility matching the material that
encapsulates the solenoid, in this case epoxy, and sample vial has been shown to
reduce this field broadening.
2 Additionally the presence of air
bubbles that can occur in the epoxy is potentially a source of both field
broadening and offset. The effects of such air bubbles have yet to be discussed
in the literature. This work seeks to investigate the systematic errors due to the
susceptibility of epoxy and presence of air bubbles in NMR field probes, as a
guide to construction of such probes for accurate quantification of time
varying fields in MR systems.
Methods
Magnetic field NMR probes were modeled
using COMSOL, solved using the ‘Magnetic Fields, No Current’ physics solver. Four
separate models were created. The first was a realistic NMR field probe which
consisted of a glass cylinder of radius 1.4mm and wall thickness 0.2mm, containing
water, surrounded by a 4 turn copper solenoid with wire thickness 0.225mm. Encapsulating this assembly was an epoxy
ellipsoid with major and minor radii 15 and 10mm respectively. (Fig. 1) The second and third models
were identical to the first with the exception that the second model did not
contain the solenoid domain and the third model while containing the solenoid
domain did not contain the water domain. The fourth model differed from the
first only by the addition of a 1mm sphere of susceptibility 0 placed at
various positions w.r.t. the center of the vial. In this model the
susceptibility of the epoxy was kept at a constant value of -9.5658 ppm.
All models were simulated in a static homogenous
magnetic field.
To determine the optimal susceptibility for
the epoxy, the first 3 models were run using a range of susceptibilities of
epoxy from -20 to 1.2 ppm. To
investigate the systematic errors caused by the presence of an air bubble, model
4 was used. It was solved with
the sphere at various angles while remaining fixed in the Z=0 plane at a radius
of 5mm in addition to the sphere at different z positions while remaining fixed
in the XY plane.
Results
Values of the magnitude of the B field from
the water domain were taken and the weighted variance of those field points was
calculated using the sensitivity profile of the solenoid as a weighting factor.
The field broadening is defined as the square root of the weighted variance. The
field offset was calculated by subtracting the weighted mean of the B field
data from the applied field.
The results from varying the susceptibility
of the epoxy are shown in figure 2. The
resulting field offset for the NMR field probe as a function of epoxy
susceptibility is plotted in figure 3.
Results from introducing an air bubble at various angles of a circle of radius
5mm in the Z=0 plane are shown in figure
4. Figure 5 shows the results
from the model of the NMR field probe with an air bubble at a constant (X,Y) value but at varying positions in Z.
Discussion
For the case of susceptibility matching the
epoxy, figure 2 shows that the
optimal susceptibility (vertical black line) lies somewhere between the
susceptibility of water (vertical blue line) and copper (vertical red line) and is
not that of the susceptibility of copper,
as might have been supposed. The
best matched case is an almost tenfold improvement in field broadening compared
to having the probe in air alone, as expected. Although field offset is shown
to decrease as susceptibility increases, offset is inherent and can be
calibrated for. Broadening cannot be calibrated for, thus the main concern
should be to susceptibility match.
Figure
4 shows that the effect of the bubble on broadening
is symmetric as expected. Figure 5 shows
that the broadening is particularly severe when the bubble is close to the
solenoid.
Conclussion
It was shown that using susceptibility matched epoxy improves field broadening by a factor of almost 10 and that the
optimal susceptibility occurs when the epoxy is matched closely to that of the water. In the future, epoxy susceptibility should be matched for highest accuracy. Additionally as the presence of air bubbles is
shown to increase field broadening and is particularly severe at certain
critical regions within proximity to the solenoid, it is necessary to eliminate
them in construction though the use of a vacuum pot method.
Acknowledgements
I would like to acknowledge the members of
the Chronik lab at Western University for their helpful discussions concerning
the formatting of the figures in this abstract. References
1. De Zanche, N., Barmet, C., Nordmeyer-Massner,
J. a., & Pruessmann, K. P. (2008). NMR Probes for measuring magnetic fields
and field dynamics in MR systems. Magnetic Resonance in Medicine, 60(1),
176–186.
2. Stoll, M. ., & Majors, T. . (1982).
Reduction of magnetic susceptibility broadening in NMR by susceptibility
matching. Journal of Magnetic Resonance (1969), 46(2), 283–288.