Zhoujian Li1, Sajad Hosseinnezadian1, Geneviève Guillot1, Georges Willoquet1, Laurène Jourdain1, Marie Poirier-Quinot1, Luc Darrasse1, and Jean-christophe Ginefri1
1Laboratoire d'Imagerie par Résonance Magnétique Médicale et Multi-Modalités, Université Paris-Sud, Orsay, France
Synopsis
We have implemented a piezo-motor based
automation system for contactless impedance matching of a monolithic Transmission
Line Resonator (TLR) operating at 4.7 T. The automation system successfully achieved inductive matching to more
than -30dB of the TLR inside the magnet and no artifacts was observed on the
image of a rectangular box-shaped water phantom. A second image, acquired
in the same condition but in the presence of another piezo-motor fixed on a
side of the sample revealed that the close proximity of the piezo-motor to the
sample brings B1-field inhomogeneity.Purpose
The monolithic
design of self-resonant transmission line resonators (TLRs)
1 is
attractive for developing small-sized coils that can be made flexible or
fabricated with superconducting material and thus offers several strategies for
improving the SNR in NMR experiments
2. In a previous work
3,
we investigated the performance of contactless tuning and matching techniques based
on electric and inductive coupling and we evaluated on the bench the
feasibility of implementing these techniques using an automation system based on
piezoelectric actuator. In the present work, we have implemented a piezo-motor
based automation system for imaging experiments at 4.7 T and we present the
first image acquired with a TLR being automatically matched.
Methods
This work was
conducted with a 6 turn – 4 gap TLR 4 tuned at 199.8 MHz (1H at 4.7 T) by dielectric
coupling with a FR-4(Glass-enforced epoxy, εr = 4.6) slice. The TLR was inductively coupled to a pickup-loop
attached to a piezoelectric motor (PiezoLEGS®, PiezoMotor, Sweden). The
matching procedure was performed using the automation system so as to adjust
the horizontal distance between the pickup-loop and the coil. The automatic
system consists of a graphical user interface (GUI) developed with MATLAB (MathWorks,
Massachusetts, USA) and a driver board for controlling the piezoelectric motor
and communicating with the computer. The TLR was successfully matched to more
than -30 dB inside the bore of the magnet using the set-up shown in the figure
1.
MRI was done on a home-assembled 4.7 T scanner controlled by
an Apollo sequencer (Tecmag, Texas USA) equipped with a BGA12 gradient bore
(400 mT/m, Bruker Biospin, Germany). The parameters of the 3DGE acquisitions
were TR/TE 29.78/2.57 ms, FOV (6 cm * 6 cm * 6 cm) and pixel bandwidth 398 Hz. The coil was
loaded with a rectangular box-shaped phantom (5.9 cm * 4.2 cm * 6 cm) filled with water.
The TLR was fixed to the bottom of a 6 mm thick Plexiglass support with the
sample placed on top of it (Figure .1). In order to evaluate the influence of
the piezoelectric actuator on the image quality, we acquired a first image
using the piezo-motor of the automation system only, placed 7 cm horizontally away
from the center of the TLR (Figure 1). We acquired a second image in the same
condition than the first one but adding another piezo-motor fixed on a side of
the sample. The later configuration corresponds to a worst case regarding the
position of the piezo-motor. Both experiments were done after achieving the desired
matching level using the automation system.
Results
Figure 2
shows the image obtained in the axial plane in the presence of the piezo-motor of
the automation system only. No significant artefacts were observed. Figure 3
shows the image acquired in the presence of the second piezo-motor fixed to the
sample wall. It demonstrates that the close proximity of the piezoelectric
motor to the sample brings inhomogeneity to the B1-field of the coil.
Conclusion
The
piezo-motor based automation system successfully achieved inductive matching of
a TLR inside a 4.7 T NMR scanner and no artifacts were observed on the image.
However piezoelectric motor should be placed at a minimum distance from the
coil and the sample so as not to perturb the B1-field homogeneity. Since
matching the TLR can be done remotely, it facilitates the matching procedure in
small-sized bores. Furthermore, such a contactless automation system is of
interest when the accessibility to the RF coil is limited such as using HTS TLRs
in a cryogenic environment. In future work, extending the system for both automatic
tuning and matching simultaneously and compensating for their mutual influence
is of concern.
Acknowledgements
No acknowledgement found.References
[1] Serfaty S et al, Mag Reson Med, 38(4), 687-689,1997.
[2]
Darrasse L, Ginefri J.-C. Biochimie,
85:915–937, 2003.
[3] Li Z et al, Proc ISMRM, 2014.
[4] Kriegl R et al, Proc ESMRMB, 2013.