Yu Li1, Fangfang Tang1, Bassem Henin1, Fabio Freschi1,2, Feng Liu1, and Stuart Crozier1
1School of ITEE, The University of Queensland, Brisbane, Australia, 2Department of Energy, Politecnico di Torino, Torino, Italy
Synopsis
Eddy currents are inevitably
induced in the electrically conductive surroundings including the magnet
cryostat vessel, RF coil, RF shield and other peripheral metallic structures. This results in image distortion and
artefacts. In order to control eddy currents, some studies have also discussed
the issue of eddy currents on magnet cryostat vessel and RF shields. In this
work, the eddy currents on a 12 channel micro-strip RF array head coil for 7T MRI
were analysed and compared with that of a slotted, double-sided copper RF
shield. A preliminary approach is proposed to reduce the eddy current effect,
without compromising RF performance. Target audience
People interested in
electromagnetic field analysis with specific interest in RF array and gradient
coil designs for high field MRI.
Purpose
To improve high field MRI imaging quality by mitigating the deleterious effects
of eddy currents which were generated on the RF array coil when imaging with
rapid gradient pulse and sequences in high field MRI.
Introduction
In MRI, eddy currents are inevitably induced in the electrically
conductive surroundings including the magnet cryostat vessel, RF coil, RF
shield and other peripheral metallic structures. This is due to the time-varying magnetic
fields generated by the switching of gradient coils. The eddy currents
sequentially distort the spatiotemporal qualities of the primary gradient
fields within the imaging region. This results in image distortion and
artefacts. In order to control eddy currents on the magnet cryostat vessel,
many studies have introduced constrains into gradient coil design [1]. Some
studies have also discussed the issue of eddy currents on RF shields and the deleterious
effect from RF shields has been demonstrated through imaging experiments [2]. However,
analysis of eddy current effects on RF array coils has had little discussion to
date. In high field MRI RF coil design, micro-strip structured RF coils are
well utilised as they provide obvious advantages [3]. The micro-strip RF coil structures
however generally have a wide strip conductor and a large patch of copper
ground, which can potentially create the issue of unwanted eddy currents. In
this work, the eddy currents on a 12 channel micro-strip RF array head coil for
7T (300MHz) MRI were analysed and compared with that of a slotted, double-sided
copper RF shield. A preliminary approach is proposed to reduce the eddy current
effect, without compromising RF performance.
Method and Results
In order to study the eddy currents
on the RF array coil, a 12 element micro-strip array head coil (fig.1d) was modelled
with a 300mm inner diameter. Each RF coil element (Fig.1a) is 200mm long and constructed
with a 20mm wide strip conductor and a 60mm wide ground panel. The equivalent
circuit diagram is shown in Fig.1b and the magnetic field (H field) is calculated and displayed in Fig.1c.
A slotted RF coil shield with double-sided copper was modelled to
compare the eddy currents on the RF array coil. The RF shield was designed with
a 300mm inner diameter, 24 slots and 0.2mm thickness between two sides. Fig.2
illustrates the orientation of the RF shield (a), the RF head array coil (b)
and the gradient coil. Only half of the X-gradient coil is illustrated for ease
of viewing.
The X-gradient coil was energized
with a 476Amp current at a frequency of 1 kHz. The eddy current distribution on
the RF shield and RF coil was then calculated, shown in Fig.3a and b (only half
quarter of subjects were plotted here). We found that the eddy currents distribute evenly
on the RF shield but can be concentrated at some points on the RF array coil. As
illustrated in Fig.3d, the maximum deviations of eddy current induced gradient fields
are 2.23 µT and 1.15µT for the RF shield and the RF array coil respectively. The
numerical simulation results demonstrate that the eddy current effect on the
micro-strip array coil is an important issue to be considered.
In order to reduce gradient
field-induced eddy currents on the RF array coil, slots cut on the ground of
the micro-strip element were utilised. As shown in Fig. 4b, the coil ground
panel was cut along the longitude direction, with three 1mm width slots which
were180mm in length, the micro-strip conductor (Fig. 4a) was kept the same. The
modified RF coil was then modelled and simulated. The H-field was calculated
and shown in Fig. 4c-d. Compared to that of the initial coil structure without
slots, there is no visible change in the RF field pattern, strength and
S-parameters. The eddy current distribution and gradient field distortion were
then calculated and shown in Fig. 4c-d. The maximum field deviation is
significantly reduced by 90% and down to 0.13 µT.
Conclusion and Discussion
In this work, the eddy current issue for high field RF array coil was analysed.
The numerical simulations indicate that this issue is worth considering in
relation to high field RF coil design, particularly for micro-strip type array
coils. A ground slotting method was proposed which shows promising potential
for RF coil eddy current control. Experimental demonstrations will be carried
out in future work.
Acknowledgements
No acknowledgement found.References
[1] Sanchez Lopez H, ect. “Multilayer Integral Method for Simulation of
Eddy Currents in Thin Volumes of Arbitrary Geometry Produced by MRI Gradient
Coils”, Magnetic Resonance in Medicine, 2014 May; 71(5):1912-22
[2] Yujuan Zhao, etc. “Dual
optimization method of radiofrequency and quasistatic field simulations for
reduction of eddy currents generated on 7T radiofrequency coil shielding”,
Magnetic Resonance in Medicine, Volume 74, Issue 5, pages 1461–1469
[3] Bing Wu, etc. "Shielded
Microstrip Array for 7T Human MR Imaging," in Medical Imaging, IEEE
Transactions on , vol.29, no.1, pp.179-184, Jan. 2010