Vitaliy Zhurbenko1, Vincent Boer 2, and Esben Thade Petersen 2
1Technical University of Denmark, Kgs. Lyngby, Denmark, 2Danish Research Centre for Magnetic Resonance, Centre for Functional and Diagnostic Imaging and Research, Copenhagen University Hospital Hvidovre, Hvidovre, Denmark
Synopsis
Transmission line
resonators are often used as coils in high field MRI. Due to distributed nature
of such resonators, coils based on them produce inhomogeneous field. This work
investigates application of series capacitors to improve field homogeneity
along the resonator. The equations for optimal values of evenly distributed
capacitors are presented. The performances of the segmented resonator and a
regular transmission line resonator are compared.Purpose
To increase longitudinal
homogeneity of the magnetic field in transmission line resonators. The described
approach to control magnetic field distribution would be useful in the design
of coils with large in terms of wavelength field of view.
Background
In order to boost the
sensitivity of transmission line coils they are usually operated as resonators.
Hence, the loading of transmission lines is either open-circuit, short-circuit,
or purely reactive. Such a loading results in infinite standing wave ratio, SWR,
(assuming lossless case), where the distance between the consecutive minima or
maxima is one-half a wavelength. To avoid dark regions on MR image and achieve
reasonable homogeneity, current minima should be avoided. For that reason, coils
based on transmission line resonators are typically shorter than half a
wavelength and proper capacitive loading is used to achieve symmetric current
distribution along the line
1. The current magnitude reduces at the
ends of the line due to destructive interference of the incident and reflected electromagnetic
waves. It is, however, preferable to have a uniform current distribution. That
would result in a more homogeneous magnetic field.
Methods
A more uniform
current distribution is achieved by inserting series capacitors Cs into the transmission line
(Fig. 1). These capacitors can
compensate for self-inductance of the transmission line by introducing opposite
phase shift 2,3. Such arrangement results in a smaller SWR than in
the conventional resonator case. The optimal in terms of SWR value of the series
capacitor Cs is expressed
as follows:
$$ C_{s}=\frac{sin(\frac{\beta l}{N})}{4 \pi f_{0}Z_{0}(1-cos(\frac{\beta l}{N}))} ,$$
where l is the total length of the segmented transmission
line coil; N is the number of segments
after inserting series capacitors (for example, using two series capacitors
will result in N = 3); β is the phase constant; Z0 is the characteristic
impedance of the implemented transmission line; f0 = 298 MHz is the operating frequency. The
value of the loading capacitor is 2Cs,
which provides symmetric current distribution.
Results
An example of a
transmission line coil using suspended microstrip technology is shown in Fig.2. Series capacitors are
formed by overlapping adjacent transmission line sections. The magnetic field
profiles for the regular transmission line resonator and coil in Fig. 2 in
free-space are illustrated by full-wave simulations in Fig. 3.
The field profiles are presented for
different
r, which is the
perpendicular distance from the coil to the point of observation. As can be
seen, segmentation improves homogeneity of the field. For example, at
r = 3 cm homogeneity,
which is defined as a ratio min {|
B(z)|} / max {|
B(z)|}, increases from 33.6 % for
regular line resonator to 56.2 % for segmented resonator with two series capacitors.
Since the field strength reduces at the ends of the line, even a higher degree
of homogeneity can be achieved by reducing the field of view.
To evaluate efficiency of the designed coil in
the presence of lossy tissue, phantom studies were conducted using a 7T system.
Photographs of the implemented prototype resonators are shown in Fig. 4. The measurement
results using a saline phantom are shown in Fig.5.
Discussion and Conclusion
As can
be seen in Fig.5, there is
an improvement in the homogeneity along the transmission line. Though the
intensity of the field closer to the coil is higher for traditional resonator,
the segmented resonator exhibits deeper penetration depth.
It is demonstrated, that the magnetic
field homogeneity for the transmission line resonators can be improved by segmenting
them with series capacitors. The higher the number of capacitors, the more
uniform field can be generated. Virtually, any degree of homogeneity is
possible. The only limitation is the finite length of the coil (which can be
overcome by reducing the field of view) and, potentially, losses in the capacitors.
Another benefit of using series capacitors is the capability to provide in a
simple manner the required uniform field over a very long length, and construct
transmission line coils, which can be even longer than one-half of a
wavelength.
Evenly distributed capacitors were considered in
this work (equal value capacitors inserted between equal length line sections).
It is expected, that higher homogeneity over a wider field of view could be
achieved using unevenly distributed capacitors with variable values.
Acknowledgements
The authors would like
to thank Danish National Research Foundation (grant DNRF124) for partial
support of the activities.References
1. Vaughan J. T., et al., Magn. Reson. in
Med., 1994, Vol. 32, Issue 2, pp. 206-218.
2. Zhurbenko V., Journal of Sensors,
2015.
3. Zhurbenko V. et al., Proc. IMOC 2015, pp.1-5.
4. Yarnykh, Magn. Reson. in Med., 2007, Vol.
57, Issue1, pp. 192-200.