Xueming Cao1, Elmar Fischer1, Oliver Gruschke2, Jan Korvink2, Jürgen Hennig1, and Maxim Zaitsev1
1University Medical Center Freiburg, Freiburg, Germany, 2Karlsruher Institut für Technologie, Karlsruher, Germany
Synopsis
In receiver coil arrays, the most commonly used
decoupling methods are overlap together with low-input-impedance preamplifiers.
But very small receiver coils can not be decoupled effectively with these two
methods. A new decoupling method, which is helpful for receiver coil arrays
with very small coils, is developed here. The coil arrays decoupled with this
method have less noise correlation and better performance in
highly accelerated imaging in the sample periphery.Introduction
Receiver coil arrays are widely used in MRI for obtaining
high local SNR and improved imaging speed [1]. The most commonly used decoupling
methods are overlap (for neighboring coils) and low-input-impedance
preamplifiers (for non-neighboring coils) [2]. As the channel number in coil
arrays increases, element coils become smaller. At some point, the overlapping
method becomes impractical because of the limits on the available space for
matching networks and preamplifiers. Furthermore, overlapped neighboring coils increase
the mutual inductance between next-neighbor coils and degrade the preamplifier decoupling method for very small coils [3]. Therefore, new decoupling methods that can
decouple very small coils in receiver arrays should be developed. Here, we present
a new decoupling method together with a corresponding coil array.
Methods
The layout
of the corresponding coil array is shown first in Figure 1. The coil array has overlapped receiver coils
in the direction aligned along $$$\overrightarrow{B_{0}}$$$ and gapped receiver
coils in the direction transverse to $$$\overrightarrow{B_{0}}$$$. The layout
is designed considering the uniformity of array sensitivity. All the receiver
coils have the same shape and orientation. Only 16 channels are shown in Figure
1, but it is easy to extend the number of channels in two dimensions.
All decoupling
setups have the same structure. Each consists of four small decoupling loops connected
to the corners of four neighbor receiver coils separately. These four
decoupling loops are strongly overlapped with each other, thus, coupling among
receiver coils is counteracted by the strong negative coupling between these decoupling
loops. The layout of a representative decoupling setup is shown in Figure 2a (top-view)
and Figure 2b (substrate). The four decoupling loops stay in four different
layers, with distances between layers in agreement with current microPCB
technology (Dyconex AG, Switzerland). The topview layout (Figure 2a) of the
decoupling setup is generated by us with the purpose of counteracting the
coupling among receiver coils. By using this decoupling setup and the proposed coil
array layout, the mutual inductance between next neighbor coils can be kept small
enough and the preamplifier method can work effectively.
The
performance of the newly designed coil array is simulated and results are
compared with the traditional overlapped array (abbreviated as Overlapped Array).
Limited by computer memory, the 2D array is represented by two 1D 10-channel
arrays: one aligned along $$$\overrightarrow{B_{0}}$$$ (abbreviated as A_Array)
and the other transverse to $$$\overrightarrow{B_{0}}$$$ (abbreviated as
T_Array). Since the A_Array has overlapped coils similar to the traditional Overlapped
Array, very similar performance is found (not shown). Therefore, only the
comparison between the T_Array and the Overlapped Array are shown.
Results
First, noise correlation is compared with a Gapped
Array as a reference. Seen from Figure 3, the largest noise correlations, coming
from neighboring coils, are 44.9%, 48.2% and 56.4% in the three arrays,
respectively. Obtained intrinsic SNR maps, without acceleration (R=1) and with
strong acceleration (R=9), are compared in Figure 4 and Figure 5, respectively.
SNR was represented by $$$\frac{B_{1-xy}^-(\triangle V)}{\sqrt{R_{sample}}}$$$ for
simplicity. With strong acceleration, the T_Array has better performance than
the Overlapped Array, especially at the sample periphery.
Discussion
Both the
new decoupling method and the overlapping method use overlap to decouple
neighboring coil, but overlap in two methods contributes differently to the
noise correlation of receiver coils. In the overlapping method, the overlapped
sensitivities of receiver coils contribute a lot to noise correlation, while in
the new decoupling method, the overlapped sensitivities of small decoupling loops
only have a minor contribution to the noise correlation because of the gaps
between coils and sample. In highly accelerated parallel imaging, lower noise correlation
means lower g-factor, and higher SNR, as seen in Figure 5. It could be
predicted that with more receiver channels, higher acceleration factor can be successfully
applied.
Considering
the low noise correlation, the new decoupling method is similar to the transformer
decoupling method [4]. However, every two coupled coil loops need one pair of
transformers in the transformer decoupling method, so the new decoupling method
needs fewer decoupling setups. Similar to the transformer decoupling method, a
disadvantage of the new decoupling method is the large resistance induced by
the decoupling loops.
Conclusion
A new decoupling method has been developed which
is specifically useful for very small receiver coils in coil arrays. Compared
to the traditional overlap method, the new decoupling method contributes less
to noise correlation between receiver coils, and has better performance in highly accelerated
imaging especially at the sample periphery.
Acknowledgements
No acknowledgement found.References
[1] Keil, Boris,
and Lawrence L. Wald. "Massively parallel MRI detector arrays." Journal
of Magnetic Resonance 229
(2013): 75-89.
[2] Roemer,
Peter B., et al. "The NMR phased array." Magnetic
resonance in medicine 16.2
(1990): 192-225.
[3] Xueming Cao.
"The Design of Massive Parallel Receive Coils for MRI." PhD thesis, Albert-Ludwigs-Universität Freiburg, 2015.
[4] Tabbert, M., et
al. "An 8-channel rat-body array coil for transceive at 9.4 T." Proceedings
of the 17th Annual Meeting of ISMRM, Honolulu, Hawaii, USA. 2009.