Chang-Hoon Choi 1, YongHyun Ha1, Arthur W. Magill1, and N. Jon Shah1,2
1Institute of Neuroscience and Medicine-4, Research Centre Juelich, Juelich, Germany, 2Faculty of Medicine, Department of Neurology, JARA, RWTH Aachen University, Aachen, Germany
Synopsis
A novel double
tuned (1H/23Na) butterfly/loop surface coil using LCC
traps was designed whereby the sodium mode was operated in a quadrature. The
performance of this coil was evaluated on a 4T whole-body scanner and compared
with a single-tuned butterfly and a loop coil. Images obtained by the quadrature-compensated
double-tuned RF coil were more uniform in each slice and the SNRs were slightly
higher over the selected ROIs compared to those from the reference coils.Purpose/Introduction
Combining X-nuclei
with proton MRI allows accessing valuable cellular and metabolic information.
1, 2 Due to lower intrinsic sensitivity of
non-proton NMR, compared to
1H, it is important to ensure that the signal-to-noise
ratio (SNR) of X-nuclei is as high as possible. Quadrature coils are widely used
due to the benefit in SNR.
3 In this work, a novel, double-tuned (
1H/
23Na) surface
coil using LCC traps
4,5 was designed in which
23Na was driven
in quadrature. The performance of this coil was evaluated and compared with
geometrically identical single-tuned butterfly and loop coils.
Methods
As shown in Figure
1, a double-tuned loop coil on the top of a single-tuned
23Na
butterfly coil was designed and constructed. These two coils were geometrically
isolated and LCC traps were added only to the loop in order to achieve dual
frequencies and not to lose the sensitivity of the butterfly sodium coil. This
allowed quadrature operation at the sodium frequency. A T/R switch including a
23Na
hybrid coupler was built and three reference coils of single-tuned
1H
loop,
23Na loop and
23Na butterfly were also prepared for
comparison. These probes were tuned and matched to a uniform water phantom
containing 44.8 mM NaCl and 4.7 mM NiSO
4, and the loaded/unloaded Q
factors were measured on the bench using a double-sniffer loop positioned at ~15
mm from the coil (Figure 2). All the experiments were carried out on a 4T whole-body
system corresponding to the Larmor frequency of 168.2 MHz for proton and 44.5
MHz for sodium. Shimming was adjusted using the
1H coil and the obtained
shim values were applied for the sodium measurements.
Results
1H and
23Na images were acquired using
a 2D and a 3D FLASH sequence, respectively. The imaging parameters for
proton/sodium were TR = 150/130 ms, TE = 4.34/2.78 ms, averages = 2/8, 5 mm
slice thickness and the acquisition time = 0:38/13:19 minutes. SNR values were
measured at the selected regions of interest (ROIs) – large, centre and side (top
right in Figure 3) and plotted across the imaging volume (top left in Figure 3).
The first and last two slices were discarded for the SNR calculations as these
were either too close to the coil or too far. It is known that the loop coil
provides high signal intensity around the coil surface but not at the centre
while the butterfly coil does the opposite. The SNR profiles (bottom in Figure 3)
demonstrate this and also indicate that the proposed concept can compensate for
this effect with maintaining comparable SNR to the single-tuned reference coils
at the side, centre and overall.
Discussion/Conclusions
It was shown that the
images obtained by the quadrature compensated double-tuned RF coil were more
uniform in each slice and the SNRs were slightly higher over the selected ROIs
compared to these from the reference coils. The loss in the proton side due to
the inserted traps was less than 10% (e.g. slice #3:
SNR
single = 2901, SNR
double = 2704), which may also be minimised by optimising the
LCC traps. The LCC traps may also be included only in the butterfly coil rather
than in the loop. We intend to apply this concept to an animal surface coil
and/or a human volume array and to carry out further investigation.
Acknowledgements
No acknowledgement found.References
1. Shah N.J., et
al. Imaging of sodium in the brain: a brief review. NMR Biomed. 2015.
2. Bottomley P.A.,
et al. Human in vivo Phosphate Metabolite Imaging with 31P NMR, Magn. Reson. Med.
1988;7:319-336.
3. Hoult D.I., et
al. Quadrature detection in the laboratory frame. Magn Reson Med.
1984;1:339-353.
4. Meyerspeer M.,
et al. An improved trap design for decoupling multinuclear rf coils. Magn.
Reson. Med. 2014;72(2):584-590.
5. Roig E.S., et al.
A double-quadrature radiofrequency coil design for proton-decoupled carbon-13
magnetic resonance spectroscopy in humans at 7T. Magn Reson Med.
2015;73:894-900.