Michaela A U Hoesl1, Peter R Seevinck1, Matteo Maspero1, Gert J Meijer2, Jan J W Lagendijk1, Bas W Raaymakers1, and Cornelis A T van den Berg1
1Center of Image Sciences, University Medical Center Utrecht, Utrecht, Netherlands, 2Radiotherapy, University Medical Center Utrecht, Utrecht, Netherlands
Synopsis
Pseudo-CT (pCT) generation for Head and Neck region
based on ultrashort echo time and radial under sampling is investigated in
order to reach a clinical acceptable time frame for image acquisition. Two
different UTE sequences, a 3D radial “kooshball” and a 3D radial “stack-of-stars”
k-space acquisition are compared for image acquisition and pCT result using
tissue classification and bulk density assignment. The results suggest that
radial undersampling is feasible and thus results in a time frame of clinical
relevance of 3 min for image acquisition plus 1 min for post-processing pCT
generation.Purpose
MRI is increasingly integrated into the radiotherapy treatment
workflow because of high soft tissue contrast and tumor visibility. To fully
exploit the benefits of MRI a stand-alone MRI workflow for radiotherapy is
desirable. However, CT remains the standard since it provides electron density information,
which is a requirement for treatment planning. Therefore the generation of pCT from
MR images is investigated e.g. by tissue classification. In head and neck separating
air and bone voxels is especially challenging because of neighboring location
and fast signal decay of bone.
In this work the goal is to address geometrical
accuracy of different UTE based sequences, its effect on pCT generation and the
possibility to accelerate acquisition by undersampling.
Methods
Image acquisition: Two triple echo UTE sequences were tested: a 3D in-plane
radial k-space acquisition “stack-of-stars” (SOS) sequence (TR/TE/ΔTE
(ms) 6.9/0.14/1.1; FOV 232x232x256 mm^3,
FA 10°, 1.5x1.5x2.0mm3) and a 3D radial “kooshball” (KOOSH) sequence
(TR/TE/ΔTE (ms) 6.5/0.14/1.1; FOV (232x232x232
mm3), FA 10°, 1.5x1.5x1.5 mm3) on a 3T scanner (Ingenia, Philips, Best, Netherlands) with a 8-channel head coil. For both sequences radial
undersampling up to 60% was tested in order to decrease scanning time. Each
sequence was tested on two volunteers and compared for image acquisition
convenience. pCT results are compared to the corresponding result from
undersampled data.
Image-processing: An image-processing pipeline for synthesizing a
pCT based on a triple echo UTE sequence was established (Fig.1) based on a single scan. First, a
bone-enhanced image is derived, using the first and third echo magnitude images
2(M1+M3)/(M1+M3) [1](Fig.2a).
The third echo was used to derive both a body mask for
segmentation and a water-only pCT. A two-point Dixon decomposition [2] identified
soft tissue voxels in the bone-enhanced image (Fig 2b,c). Air voxels within the patient are classified by Otsu-thresholding
and morphological filtering on M1 (Fig.2d,e).
A threshold operation was applied to isolate bone voxels. Finally body, bone
and air-inside mask are assigned bulk density according to literature values
[3]. Image processing was performed using IPython and the Insight Toolkit (SimpleITK)
[4,5].
Geometric accuracy: For verifying geometric accuracy, the sequences
were tested on a realistic agar head phantom comparing body outline
differences for different echoes and parameter settings to a high resolution CT
image.
Results
Fig.2f shows the final bone segmentation resulting in a bone
mask visible in
Fig.2i.
Fig.2g presents the resulting pCT and
Fig.2h the Digital Reconstructed Radiography
(DRR) image for position verification. pCT generation takes 1:05 min (8 core CPU computer), the imaging
time employing radial undersampling is 3:01min.
Fig.3 displays the results comparing the SOS and KOOSH
sequence including the result for radial under sampling. Tissue classification is
compared in percent to the body volume. Radial undersampling leads to
a percentage difference in the order of <1% for the KOOSH sequence and to ≈ 4% difference in SOS bone and
water classification for volunteer1 and <1% difference for volunteer2.
Geometrical accuracy, tested on the agar-phantom
showed a match in body outline of CT to M1(M2,
M3) to 95.0%(98.2%, 98.8%). Comparing
body outline for the same echo time but different trajectory delays (1,2,10,-10
μs) resulted in a difference of
1%.
Discussion
Image acquisition and image processing time in total
is in the order of 4 min if radial undersampling is used, which is within the requirements
for MR-linac applications [6]. Comparing the feasibility of the sequences, the SOS
gives more freedom and flexibility in tradeoffs for resolution and scanning
time, especially for large FOV including shoulders.
The image-processing pipeline has to be tested on an
increased number of datasets to ensure robustness. It can be potentially
improved employing a conversion of HU based on intensity levels instead of a
threshold on the final bone-enhanced image in combination with bulk assignment [7]. Also air
segmentation robustness has to be validated and potentially improved [8]. An
open problem are dental implants and metal wire artifacts, which were observed both
datasets. It has to be seen if manual bulk assignment in the artifact area is
clinically feasible and dosimetrically accurate.
Conclusion&Outlook
The results suggest that pCT generation for Head
and Neck within clinical acceptable time is feasible. As a next step we aim to verify
the tissue classification with CT data and compare dose outcome of a clinical
plan on pCT and CT. The workflow will be adapted and tested for realistic
radiotherapy treatment, which involves imaging in immobilization mask.
Acknowledgements
This research is funded by ZonMw IMDI Program, “RASOR sharp: MRI based radiotherapy planning
using a single MRI sequence”, project number: 10-10400301References
[1] Berker Y, Franke J, Salomon a., Palmowski
M, Donker H C W, Temur Y, Mottaghy F M, Kuhl C, Izquierdo-Garcia D, Fayad Z a.,
Kiessling F and Schulz V 2012 MRI-Based Attenuation Correction for Hybrid PET/MRI
Systems: A 4-Class Tissue Segmentation Technique Using a Combined
Ultrashort-Echo-Time/Dixon MRI Sequence J. Nucl. Med. 53 796–804
[2] Dixon W T 1984 Simple
proton spectroscopic imaging. Radiology 153 189–94
[3] International Commission
on Radiation Units and Measurements (ICRU): Photon, Electron, Proton and
Neutron Interaction Data for Body Tissues. ICRU Report 46. Bethesda, MD:
International Commission on Radiation Units and Measurements; 1949
[4] Pérez F and Granger B E
2007 IPython: a System for Interactive Scientific Computing Comput. Sci.
Eng. 9 21–9
[5] Lowekamp B C, Chen D T,
Ibáñez L and Blezek D 2013 The Design of SimpleITK. Front. Neuroinform. 7
45
[6] Lagendijk J J W, Raaymakers B W, Van den
Berg C A T, Moerland M A, Philippens M E and van Vulpen M 2014 MR guidance in
radiotherapy Phys. Med. Biol. 59 R349–69
[7] Juttukonda M R, Mersereau B G, Chen Y, Su
Y, Rubin B G, Benzinger T L S, Lalush D S and An H 2015 MR-based attenuation
correction for PET/MRI neurological studies with continuous-valued attenuation
coefficients for bone through a conversion from R2* to CT-Hounsfield units Neuroimage
112 160–8
[8] Zheng W, Kim J P, Kadbi
M, Movsas B, Chetty I J and Glide-Hurst C K 2015 Magnetic Resonance–Based
Automatic Air Segmentation for Generation of Synthetic Computed Tomography
Scans in the Head Region Int. J. Radiat. Oncol. 93 497–506