Henrik Odéen1, Bradley Bolster2, Eun Kee Jeong1, and Dennis L Parker1
1Utah Center for Advanced Imaging Research, Department of Radiology, University of Utah, Salt Lake City, UT, United States, 2Siemens Healthcare, Salt Lake City, UT, United States
Synopsis
Measurements
of changes in signal intensity and T1 relaxation time with temperature has been
suggested for temperature monitoring in cortical bone during MR guided focused
ultrasound treatments. In this study we compare changes in signal intensity,
T1, and T2* with temperature using a 3D ultrashort echo time pulse sequence and
a 2D gradient recalled echo pulse sequence with short TE. The effects of T1 and
T2* change with temperature counteract each other making the change in signal intensity
small, and therefore T1 and T2* appears to have the greatest sensitivity to
changes in temperature.Introduction/Purpose
For MR guided focused ultrasound treatments in
or close to bone, such as for transcranial applications focusing through the
intact skull bone and treatments of bone metastases, significant heating can
occur in the bone. MRI of bone is in general challenging due to the short T2 of
bone. For MR temperature imaging the short T2 also severely decreases the
accuracy of the standard proton resonance frequency shift method. Instead
researchers have investigated the temperature dependence of the MR signal
intensity (SI) and T1 relaxation time for temperature monitoring1–4. Miller1 and Fielden3 et al showed that the SI from cortical bone decreases
with increasing temperature using ultrashort echo time (UTE) pulse sequences,
and Ramsay2 et al found that, contrary to what Miller and Fielden
observed, the SI increases with increasing temperature using a short TE
gradient recalled echo (GRE) pulse sequence. Han4 et al further showed that T1 increases with
temperature, also using UTE.
In this work we investigate the temperature
dependence of the SI (dSI/dT) and the T1 and T2* relaxation times (dT1/dT and
dT2*/dT, respectively) using 3D UTE and short TE 2D GRE to investigate which
parameter has the highest sensitivity to temperature change.
Methods
All
imaging was performed on a 3T MRI scanner (MAGNETOM PrismaFit, Siemens
Healthcare, Erlangen, DE) using a 3D UTE and a 2D GRE pulse sequence. The UTE
sequence utilizes radial, ramped sampling of k-space in 3D starting at the
k-space center after a 80 μs hard RF pulse, allowing TEs down to 50 μs. T2* was measured by an exponential fit to data acquired at TE=50,
90, 130, 170, and 250 μs (other scan parameters are listed in Table 1). The 50 μs TE was also used for dSI/dT calculations. T1 was measured using the variable
flip angle (VFA) method5 with FA 8 and 36° (Table 1).
The
spoiled GRE sequence utilizes an asymmetric echo to allow TE down to 1.20 ms. For
T2* measurements 6 contrasts with TEs between 1.20 and 7.50 ms were acquired,
and then repeated 4 more times with TEs starting at 1.30, 1.40, 1.50, and 1.60
ms, Table 1. The 1.20 ms echo was used for dSI/dT calculations. T1 measurements
were performed using the VFA method with FA 8 and 36°.
An approximately 4-cm long bovine femur bone (marrow
and connective tissue removed) was placed in a phantom holder that allowed
heated water to circulate around the bone, Figure 1. 1 fiber optic probe measured the
water temperature, and 3 probes were inserted in 1-mm diameter, 2-cm deep,
drill holes in the bone to measure the temperature of the bone. The water was
warmed to 4 temperatures (~22, 35, 50, and 65 °C) and the data was collected when all 4 probes measured
within 1 °C. The whole setup was places in a 20-channel RF
head coil.
Results
Figure
2 shows 2D maps of T1 and T2* relaxation times for the 4 different temperatures
for the UTE and GRE scans. Mean and standard error values from a 9x9 ROI close to
each probe is shown in Figure 3, together with calculated changes in %/°C. From the UTE data a decrease in SI of 0.3-0.5 %/°C,
and increases in T1 and T2* of 0.5-0.9
%/°C
and 0.6-0.9 %/°C,
respectively, was observed. From the GRE data a slight increase in SI of 0.06 %/°C, and decreases in T1 and T2* of 0.4-0.5 %/°C and 0.3-0.4
%/°C,
respectively, was observed.
Using
the temperature dependent spoiled GRE signal equation6 and the observed values for dT1/dT and dT2*/dT,
dSI/dT can be closely predicted for the UTE case, whereas the GRE case predicts
a small decrease in dSI/dT.
Discussion
and Conclusions
The
decrease in SI for UTE, and increase in SI for GRE, is in accordance with
previously published results. However, we observed a lower change in the GRE SI (0.06%/°C) than the 0.9 %/°C reported by Ramsay. The measured change in T1
using UTE agrees well with the 0.6 %/°C
reported by Han.
The
lower dSI/dT compared to Ramsay may be partly explained by our slightly longer
TE (1.20 versus 1.05 ms) which will detect more long T2-component protons. For
both UTE and GRE the effect of T1 and T2* on SI are counter-acting each other
(both increasing for UTE, and both decreasing for GRE), which reduces the
sensitivity of dSI/dT. This may suggest that dT1/dT
and dT2*/dT are more suitable candidates for bone MR thermometry, and Figure 3
also shows higher sensitivity for relaxation times that for SI.
Acknowledgements
This work was supported by The Focused Ultrasound Surgery Foundation,
Siemens Healthcare, The Ben B. and Iris M. Margolis Foundation, and NIH
grants R01s EB013433 and CA134599References
1. Miller W. Toward
T1-Based Thermometry in Cortical Bone Using Ultrashort Echo-Time MRI. In:
International Symposium on Focused Ultrasound. Bethesda, Maryland; 2012. p.
65–BN.
2. Ramsay E, Mougenot C, Kazem
M, Laetsch TW, Chopra R. Temperature-Dependent MR Signals in Cortical Bone?: Potential
for Monitoring Temperature Changes during High-Intensity Focused Ultrasound
Treatment in Bone. Magn. Reson. Med. 2014:ePub ahead of print. doi:
10.1002/mrm.25492.
3. Fielden S, Mugler III J,
Miller W, Butts Pauly K, Meyer C. Detecting signal changes in heated bone with
a 3D spiral ultra-short echo time sequence. In: Proc. Intl. Soc. Mag. Reson.
Med. Vol. 23. ; 2015. p. 1632.
4. Han M, Rieke V, Scott SJ,
Ozhinsky E, Salgaonkar VA, Jones PD, Larson PEZ, Diederich CJ, Krug R.
Quantifying temperature-dependent T1 changes in cortical bone using
ultrashort echo-time MRI. Magn. Reson. Med. 2015;ePub ahead of print. doi:
10.1002/mrm.25994.
5. Deoni SCL, Rutt BK, Peters
TM. Rapid combined T1 and T2 mapping using gradient recalled acquisition in the
steady state. Magn. Reson. Med. 2003;49:515–26.
6. Cline HE, Hynynen K, Hardy
CJ, Watkins RD, Schenck JF, Jolesz FA. MR temperature mapping of focused
ultrasound surgery. Magn. Reson. Med. 1994;31:628–636.