High Speed, High Sensitivity MR-ARFI Using a Balanced Steady-State Free Precession Pulse Sequence
Yuan Zheng1, Michael Marx1, G. Wilson Miller2, and Kim Butts Pauly1

1Radiology, Stanford University, Stanford, CA, United States, 2Radiology and Medical Imaging, University of Virginia, Charlottesville, VA, United States

Synopsis

We have developed a novel MR-ARFI technique that makes use of transition band balanced steady-state free precession (bSSFP). Due to the strong dependence of image phase on the motion-encoded phase, this technique improves the sensitivity of MR-ARFI measurements over commonly used spoiled sequences. The proposed technique also features high speed, as an ARFI contrast image can be acquired in a few seconds. With its high speed and high sensitivity, the bSSFP-ARFI technique could be useful in confirming/calibrating the HIFU focal spot before thermal ablation treatment.

Introduction

MR acoustic radiation force imaging (MR-ARFI) provides a method to visualize the focus of high-intensity focused ultrasound (HIFU) based on tissue stiffness1. When ultrasound is absorbed by tissue, a force is exerted along the beam direction and can cause tissue displacement of a few micrometers at the focal spot when using HIFU transducers. By inserting a pair of motion-encoding gradients (MEG) into MR pulse sequences, with one lobe synchronized with a HIFU pulse, spins at the ultrasound focus accrue an additional phase that can be observed in MR phase images. MR-ARFI is mostly performed with spin-echo or spoiled GRE sequences2, though fluctuating-equilibrium balanced steady-state free precession sequences have also been proposed3. In this work, MR-ARFI using transition band bSSFP is presented with a scan time of only a few seconds. This new technique improves the sensitivity of MR-ARFI measurements over spoiled sequences, by generating a larger phase change for a given displacement.

Theory

Transition band bSSFP signal is highly sensitive to the amount of phase that spins accrue in the rotating frame during each TR. This phase βoff is usually caused by the difference between the spin resonance frequency and the scanner operating frequency. The highly sensitive transition band bSSFP signal has been used for fMRI4 and thermometry5. For bSSFP-ARFI, a pair of MEG are added to traditional bSSFP sequences (Fig. 1). By synchronizing a HIFU pulse with one lobe of the MEG, an additional phase βMEG is generated by the radiation force displacement. We define the sensitivity of MR-ARFI pulse sequences as the derivative of the image phase ϕ with respect to the motion-encoded phase βMEG. The sensitivity is always 1 with conventional pulse sequences. However, for the bSSFP-ARFI sequence in Fig.1, it can be shown that the image phase is: $$$\phi=\arctan(\frac{E_2\sin(\beta_{off}+\beta_{MEG})}{1-E_2\cos(\beta_{off}+\beta_{MEG})})+\frac{TE}{TR}\beta_{off}+\beta_{MEG}$$$ (Eq. 1), where E2=exp(-TR/T2). The ARFI sensitivity $$$(\frac{d\phi}{d\beta_{MEG}}\Bigr|_{\beta_{MEG}=0})$$$ is plotted as a function of βoff in Fig. 2. The highest sensitivity occurs on resonance and equals T2/TR+1/2. Since T2 is usually much longer than TR for bSSFP sequences, MR-ARFI sensitivity is amplified near resonance.

Methods

A homemade gel phantom was used for this study. All MR images were acquired on a 3T GE MR750 scanner (GE Healthcare, Milwaukee, WI) with a single-channel local coil. Pulse sequences were developed and implemented on the scanner using RTHawk (HeartVista, Inc., Menlo Park). 2D images were collected with slice thickness=3 mm, FOV=12.8 cm, matrix=128x128. HIFU was generated by a multi-element phased-array transducer (Exablate 2000, Insightec, Haifa, Israel). The pulse duration was 1.45 ms and the acoustic power was 64.8 W. The HIFU beam was parallel to the MEG and perpendicular to the imaging slice. The bSSFP-ARFI images were acquired with FA=2°, BW=25 kHz, TE=10.5 ms, TR=14.7 ms. Each lobe of the MEG was 3 ms long, and the amplitude and duration of the plateau were 5 G/cm and 1.25 ms. Scanner frequency and shimming gradients were adjusted to ensure the expected focal spot position on resonance. Two images were acquired. HIFU was off for the first image as a baseline and on for the second one. The two images were acquired in 3.8 s. To study the effect of possible small temperature rise due to HIFU pulses on the image phase, we repeated the measurement with the same bSSFP sequence with MEG removed. To compare the enhanced sensitivity of the bSSFP-ARFI technique with traditional MR-ARFI, we kept MEG and HIFU the same as the bSSFP-ARFI experiment and performed MR-ARFI using a standard 2DFT spGRE-ARFI sequence with FA=40°, BW=16.125 kHz, TE=11.3 ms and TR=500 ms. Images were acquired with opposite MEG polarities (with separate baseline images). The total acquisition time for spGRE-ARFI was 4 min.

Results

Fig. 3a-c shows the phase contrast calculated from bSSFP-ARFI images, bSSFP images and the spGRE-ARFI images. The average phases in a 2 by 2 pixel ROI at the focal spot are 0.38, 0.10 and 0.052 respectively. Removing phase change caused by temperature change results in bSSFP-ARFI phase contrast of (0.38-0.10)=0.28. The phase contrast from the spGRE images equals βMEG. Therefore the contrast of the bSSFP-ARFI sequence is amplified by (0.28/0.052)=5.4 times.

Conclusion

We have developed a novel MR-ARFI technique that makes use of the sharp phase transition of bSSFP signals to amplify the motion-encoded phase. With its high speed and high sensitivity, the bSSFP ARFI technique could be useful in confirming/calibrating the HIFU focal spot before thermal ablation treatment.

Acknowledgements

The authors acknowledge our funding sources: P01 CA159992, InSightec and General Electric.

References

1, McDannold N, Maier SE. Magnetic resonance acoustic radiation force imaging. Med Phys. 2008 Aug;35(8):3748-58.

2, Kaye EA, Pauly KB. Adapting MRI acoustic radiation force imaging for in vivo human brain focused ultrasound applications. Magn Reson Med. 2013 Mar 1;69(3):724-33.

3, G. Wilson Miller, Kelsie F. Timbie, and Richard J. Price. Fluctuating Equilibrium MR-ARFI. Proceedings of ISMRM, 2014 (0261).

4, Miller KL, Hargreaves BA, Lee J, Ress D, deCharms RC, Pauly JM. Miller KL1, Hargreaves BA, Lee J, Ress D, deCharms RC, Pauly JM. Functional Brain Imaging Using a Blood Oxygenation Sensitive Steady State. Magn Reson Med. 2003 Oct;50(4):675-83.

5, Yuan Zheng, G. Wilson Miller. High speed, high sensitivity MR thermometry using a balanced steady-state free precession pulse sequence. Proceedings of ISMRM, 2015 (1654).

Figures

A pair of MEG is inserted in bSSFP-ARFI sequences. HIFU pulses are synchronized with the second lobe of MEG.

Sensitivity of the bSSFP-ARFI pulse sequence as a function of the off-resonance phase with T2=100 ms and TR=15 ms, calculated by taking the derivative of Eq. 1.

Phase contrast calculated from a) the bSSFP-ARFI image pair, b) the bSSFP image pair (without MEG) , c) the spGRE-ARFI acquisitions. Contrast-to-noise ratio (CNR) are 10, 2.7 and 7 respectively. The contrast of b) can be used to subtract out the fraction of contrast in a) due to temperature-induced PRF shift.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2104