Johanna Maria Mijntje van Breugel1, Martijn de Greef2, Charles Mougenot3, Maurice AAJ van den Bosch2, Chrit CW Moonen2, and Mario AAJ Ries2
1Radiology, University Medical Center Utrehct, Utrecht, Netherlands, 2University Medical Center Utrehct, Utrecht, Netherlands, 3Torontp, Canada
Synopsis
Hypothesis: MR-ARFI can be deployed in the kidney as an alternative for the thermal test shot at low power.
The employed respiratory gated MR-ARFI sequence
in combination with a 450 W excitation tone-burst is sensitive
enough to exceed the noise level and to clearly display the focal point of the HIFU
beam. Both at
450W and at 1000W the displacement due to the radiation force coincided with
the location of the temperature rise due to thermal ablation at equivalent
power. Hence, radiation force in combination with a pencil beam navigator to
compensate for respiratory motion is a reliable indicator of the location of
the thermal lesion and might be an alternative to the low power thermal test
shot in highly perfused organs such as the kidney.
Purpose
MRI-based interventional guidance of
High-Intensity Focused Ultrasound Therapy (HIFU) relies on the stereotactic
correspondence between the MR-image and the location of the HIFU ablation area.
However, discrepancies between the planned focal point location and the actual
focus of the HIFU beam are frequently observed, predominantly due to acoustic
wave-front aberrations induced by variations of the acoustic impedance in the
beam propagation path. As a consequence, current clinical practice employs a “low power thermal test shot” before the
start of the ablation, to provide a 3D correction of the
exact focus location. The power level of this test-shot is generally chosen in
such a way that it allows observing a distinct temperature elevation without
causing irreversible tissue damage. Although this approach is safe and accurate
in low perfused target tissues such as uterine fibroids, the high
heat-extraction rate renders this approach of limited value in high-perfused
organs such as liver or kidney. Furthermore, for therapeutic HIFU ablations in
the non-linear pressure range [1], shock-wave formation and non-linear
absorption adds further discrepancy between linear test-shots and therapeutic sonications.
Recently, MR-acoustic radiation force imaging
(MR-ARFI) has been proposed as an alternative method to locate the focal point
non-invasively without giving rise to non-desired tissue damage [2-5]. MR-ARFI
evidences the tissue
displacement due to the acoustic radiation force of the HIFU beam instead of
the temperature. However, since MR-ARFI employs a highly motion sensitized sequence
and displays limited sensitivity, it is challenging in abdominal organs and
in-vivo experience with this type of this approach is currently limited to the liver.
The first aim of this study was to investigate
the feasibility of respiratory-gated MR-ARFI on clinical HIFU equipment for the
exact beam localization in the kidney. Of particular interest was hereby
potential interference with other motion sources, such as cardiac pulsations in
the vicinity of larger vessels such as the renal artery. The second goal was to
validate if MR-ARFI can perform this task for both linear and non-linear acoustic
energy delivery.
Methods
A porcine model was chosen for its similarities
in size and perfusion to the human kidney. The local animal
board approved the study. The pig was anesthetized and mechanically
ventilated at a frequency of 13/min. A clinical Sonalleve MR-HIFU
therapy system (Philips Healthcare, Finland) integrated with a 1.5T Achieva MRI
(Philips Healthcare, The Netherlands) with minor modifications was used for
MR-HIFU. An MR-HIFU treatment cell was positioned at 4.5cm from the skin in the
cortex of the kidney. A gradient-recalled echo planar imaging 2D dynamic scan
(TR 100ms, TE 30ms, flip angle 20°, FOV 168x168mm2, matrix
112x108, voxel size 1.51x1.51mm2, slice thickness 7mm) was performed
to visualize displacement due to the acoustic radiation force similar to [2].
Thermometry data was obtained using a gradient echo with EPI (TR 100ms, TE 15ms, flip angle 20°, matrix
160x160, FOV 400x400mm2, voxel size 2.5x2.5mm2, slice
thickness 5mm). All MR sequences, including anatomical images, and the acoustic
energy delivery were respiratory gated based on pencil beam navigator images
of the contralateral kidney. For comparison
of linear and non-linear effects, four experiments
at two different power levels were performed: Two thermal ablations with 450W
CW and 1000W pulsed (30% duty cycle, tone-burst of 10500 cycles), and two
MR-ARFI experiments at equivalent acoustic intensity (tone-burst of 2400 cycles,
bipolar MR-ARFI gradient 20mT/m, 3ms).
Results
Displacements due to the radiation force could
be reproducibly measured for both acoustic intensities and coincided with the
location of the temperature elevation of the equivalent 450W and 1000W
ablations, respectively (Figure 1). No interference/artifacts with the cardiac
cycle were observed during MR-ARFI. The focus of the beam cone was evidenced
more posteriorly than planned for the 1000W, most likely due to shock-wave
formation and non-linear absorption. No significant heating was observed during
the MR-ARFI experiments (<2°C).
Discussion & Conclusion
The employed respiratory-gated MR-ARFI sequence
in combination with a 450W excitation tone-burst of 2400 cycles is sensitive
enough to exceed noise and to clearly display the focus of the HIFU
beam at a fraction of the energy compared to the 450W thermal ablation. Both at
450W and at 1000W the displacement due to the radiation force coincided with
the location of the temperature rise due to thermal ablation at equivalent
power. Hence, radiation force in combination with a pencil beam navigator to
compensate for respiratory motion is a reliable indicator of the location of
the thermal lesion and might be an alternative to the low power thermal test
shot in highly perfused organs such as the kidney.
Acknowledgements
No acknowledgement found.References
[1] Khokhlova VA, Fowlkes JB,
Roberts WW, Schade GR, Xu Z, Khokhlova TD, Hall TL, Maxwell AD, Wang YN, and
Cain CA. Histotripsy methods in mechanical disintegration of tissue: towards
clincal applications. Int J Hyperthermia 2015;31(1):145-62
[2] Auboiroux V, Viallon M,
Roland J, Hyacinthe J, Petrusca L, Morel DR, Goget T, Terraz S, Gross P, Becker
CD, Salomir R. ARFI-prepared MRgHIFU in liver: simultaneous mapping of
ARFI-displacement and temperature elevation, using a fast GRE-EPI Sequence. Mag
Reson in Med 2012;68:932-946
[3] Holbrook AB, Ghanouni P,
Santos JM, Medan Y, Butts Pauly K. In vivo MR acoustic radiation force imaging
in the porcine liver. Med Phys 2011;38(9):5081-5089
[4] Nightingale K, Soo MS,
Nightingale R, Trahey G. Acoustic radiation force impulse imaging: In vivo
demonstration of clinical feasibility. Ultrasound Med Biol 2002;28:227-235
[5] McDannold N, Maier SE.
Magnetic resonance acoustic radiation force imaging. Med Phys 2008;34:3748-3758