Christiaan G. Overduin1, Jurgen J. Fütterer1,2, and Tom W.J. Scheenen1
1Radiology, Radboud University Medical Centre, Nijmegen, Netherlands, 2MIRA Institute for Biomedical Engineering and Technical Medicine, University of Twente, Enschede, Netherlands
Synopsis
Our study
assessed the feasibility and accuracy of 3D ultrashort TE (UTE) MR thermometry to
dynamically track temperatures across frozen tissue during cryoablation on a
clinical MR system at 3T. We demonstrated 3D UTE imaging to achieve measurable
MR signal from frozen tissue down to temperatures as low as -40°C within a
clinically realistic time-frame (~1min) and with sufficient spatial resolution
(1.63mm isotropic). Using a calibration curve, we could derive 3D MR-estimated
temperature maps of the frozen tissue, which showed good agreement with matched temperature sensor
readings on statistical analysis.Introduction
MRI-guided
cryoablation is a promising minimally invasive therapy with applications in musculoskeletal,
liver, kidney and prostate cancer
1. To assure effective treatment, temperature
feedback during these procedures is desirable but a non-invasive approach is
still warranted. Previous studies have demonstrated measurable MR signal from
frozen tissue using ultrashort TE (UTE) imaging and explored its potential for
MR thermometry
2,3. As an important next step, we assessed in this
work the feasibility and accuracy of 3D UTE MR thermometry to dynamically track
temperatures across frozen tissue during cryoablation on a clinical MR system
at 3T.
Methods
Four
identical cryoablation experiments were performed. An MR-compatible cryoneedle
(IceRod, Galil Medical, Yokneam, Israel) was inserted into ex vivo porcine
muscle specimens at room temperature on a 3T clinical MR system (MAGNETOM Trio,
Siemens, Erlangen, Germany). Fiber optic sensors (T1, Neoptix, Quebec, Canada) were
used for temperature reference (Figure 1). Two cycles of 10:3 min. freeze-thaw
were applied. Continuous MR monitoring of ice progression was performed using a
3D radial ramp-sampled UTE sequence (TR/TE/FA = 59.5ms/70μs/15°, voxel size =
1.63x1.63x1.63mm, acq. time = 1:14min). Data of three experiments were used as
reference sets. Signal intensity (SI) values were normalized to the baseline
value before cooling and related to temperature. Data points for subzero
temperatures were fitted by an exponential function. In a separate validation
set, the obtained fit was used to generate MR-estimated temperature maps of the
frozen tissue. Statistical analysis was performed to determine accuracy of the
estimated temperature maps.
Results
In the
reference sets, next to the known T1-related signal increase during cooling
from room temperature to 0°C
4, normalized SI decreased
mono-exponentially with temperature for subzero conditions with the signal
decay fitted by normalized SI = 1.26e
0.05T (R
2=0.93)
(Figure 2). Using the fit as a calibration curve, we could obtain MR-estimated
temperature maps of the frozen tissue in 3D at each imaging time point in the
validation set (Figure 3). MR-estimated temperatures strongly correlated with
sensor readings at matched time points over the course of the cryoablation
experiment (r=0.977, p<0.001) (Figure 4a). Bland-Altman analysis demonstrated
good agreement between the two measures (Figure 4b). Mean difference between
MR-estimated and sensor measured temperatures was –1.7±2.8°C with upper and
lower limits of agreement of –7.1 and 3.8°C respectively.
Discussion
In this
work we demonstrated the feasibility of 3D UTE MR thermometry to dynamically
track temperatures in frozen tissue during cryoablation. With currently no
other method existing to non-invasively measure temperatures within frozen
tissue, the accuracy we found in this work could already be of value in a
clinical context. From a pragmatic approach, a relative signal level as
compared to baseline before cooling may be identified for which there is a high
degree of certainty that temperatures are below a certain threshold, e.g.
-40°C. Nevertheless, to apply this method clinically it should be validated how
ex vivo calibration curves translate to the in vivo setting. Notably, in the
signal peak observed with temperature decreasing from room temperature to the
freezing point we observed an offset of approximately +5°C. Theoretically,
maximum signal would be expected around 0°C, directly before the
freezing-related signal loss. This discrepancy probably results from a mismatch
between the positions of the temperature probes and the voxels used for the SI
measurements. Another contributing factor that inherently affected our
measurements is temperature averaging, both within a voxel due to the spatial
temperature gradient as well as over time due to temperature changes occurring during
image acquisition. Despite this though, a temporal resolution of 1:14min seemed
to provide adequate accuracy in tracking the temperature-induced MR signal
changes, even for the faster variations occurring at the transition from the
freeze to thawing phase.
Conclusion
3D MR
thermometry of frozen tissue using UTE signal intensity was feasible within the
time frame of a typical cryoablation procedure on a clinical MR system at 3T.
Down to temperatures as low as -40°C, accuracy of the MR-estimated temperature
maps was within clinically acceptable limits.
Acknowledgements
No acknowledgement found.References
(1) Morrison et al. JMRI 2008; (2) Wansapura et al. Acad
Radiol 2005; (3) Kaye et al. JMRI 2010; (4) Overduin et al. Med Phys 2014.