Masamitsu Hatakenaka1, Koichi Onodera1, Naomi Koyama1, and Mitsuhiro Nakanishi2
1Diagnostic Radiology, Sapporo Medical University, Sapporo, Japan, 2Division of Radiology, Sapporo Medical University Hospital, Sapporo, Japan
Synopsis
To
evaluate reproducibility of voxel-based ADC, voxel-based ADC of the phantom was
measured using clinical 3T MRI system. The direction of motion probing gradient
affected the voxel-based ADC significantly in both echo planar
imaging and turbo spin echo diffusion-weighted imaging. Also voxel-based ADC
differed both between identical positioning examinations and between slightly
different positioning examinations. Voxel-based ADC could not be reproduced sufficiently even in a phantom study. It would
be recommended to pay enough attention when performing voxel-based ADC study
like histogram analysis for tumor ADC.Purpose
To
evaluate reproducibility of voxel-based apparent diffusion coefficient (ADC).
Materials and Methods
ADC of
the phantom containing 0.2 mM gadolinium and 80% polyvinyl alcohol (Nikkofines (90-401 type),
Tokyo, Japan) was measured in a whole-body 3T clinical MRI system (Ingenia,
Philips Medical Systems, Best, The Netherland). The measurements were performed
three times for diffusion weighted imaging with both single shot echo planar
imaging (DWI-EPI) (Fig. 1) and multi-shot turbo spin echo (DWI-TSE) (Fig. 2) . 1
st and 2
nd
examinations were performed at an identical positioning and 3
rd examination
was done at a slightly different positioning. The sequence parameters were as
follows: field of view=220 mm, matrix size=112x112, number of averages=4, slice
thickness=4 mm, SENS factor=2, repetition time=3000 ms, echo time=81.1 ms for
DWI-EPI and 87.6 ms for DWI-TSE, b-factor=0 and 1000 s/mm
2, three
orthogonal motion proving gradients (MPG). Each voxel ADC was calculated by fitting signal
intensity change within the region of interest containing 197 voxels into mono-exponential
curve. ADCs measured by DWI-EPI and DWI-TSE with right-to-left, anterior-to-posterior,
foot-to-head motion probing gradients, and three synthesized imaging were
described as ADC-EPI-RL, ADC-EPI-AP, ADC-EPI-FH, ADC-EPI-ISO, ADC-TSE-RL, ADC-TSE-AP,
ADC-TSE-FH, and ADC-TSE-ISO, respectively. And ADCs from 1
st, 2
nd,
and 3
rd examinations were differentiated by adding 1, 2, and 3 at
the end, for example ADC-EPI-RL1 means ADC calculated by 1
st DWI-EPI
with right-to-left MPG. The
voxel-based ADCs measured by several sequences were compared using paired
t-test and t-test for evaluating MPG direction
effect (ADC-EPI-RL vs. ADC-EPI-AP vs. ADC-EPI-FH, and ADC-TSE-RL vs. ADC-TSE-AP
vs. ADC-TSE-FH) and reproducibility using both identical (1
st ADC-EPI
vs. 2
nd ADC-EPI, and 1
st ADC-TSE vs. 2
nd ADC-TSE)
and not identical positioning (1
st ADC-EPI vs. 3
rd
ADC-EPI, and 1
st ADC-TSE vs. 3
rd ADC-TSE). p<0.05 was
considered statistically significant.
Results
Regarding
effect of different MPG directions (Fig. 3), significant difference
was observed between ADC-EPI-RL1 and ADC-EPI-AP1, between ADC-EPI-AP1 and
ADC-EPI-FH1, between ADC-TSE-RL1 and ADC-TSE-AP1, and between ADC-TSE-RL1 and
ADC-TSE-FH1. Regarding reproducibility for identical positioning (Fig. 4), significant difference was
observed between ADC-EPI-RL1 and ADC-EPI-RL2, between ADC-EPI-ISO1 and
ADC-EPI-ISO2, and between ADC-TSE-FH1 and ADC-TSE-FH2. Regarding reproducibility for different
positioning (Fig. 5), significant difference was observed in all DWI-EPI and DWI-TSE comparisons
between 1
st and 3
rd examinations except between
ADC-TSE-AP1 and ADC-TSE-AP3.
Discussion
The present
study demonstrated that voxel-based ADC could not be reproduced using even phantom.
Many studies have been reported regarding histogram analysis based on
voxel-based ADC distribution1-4. We consider enough attention should be payed when analyzing voxel-based
ADC distribution.
Voxel-based ADC values differed among different MPG
directions in both DWI-EPI and DWI-TSE. As phantom was considered having no
anisotropy, these differences could be originated in the difference of MPG
strength and/or image distortion. Voxel-based ADC values differed between 1st
and 2nd identical position examinations. As DWI-EPI showed
significant difference in 2 comparisons, between ADC-EPI-RL1 and ADC-EPI-RL2
and between ADC-EPI-ISO1 and ADC-EPI-ISO2 but DWI-TSE showed significance in
only between ADC-TSE-FH1 and ADC-TSE-FH2, this difference may be due to strong distortion artifact in DWI-EPI and/or
low signal-to noise ratio of DWI-TSE. Voxel-based ADC values differed between 1st and 3rd
slightly different position examinations even in synthesized images. We consider this
result should be kept in mind when analyzing sequential voxel-based ADC
histogram changes of tumor, e.g. pre- and post-chemoradiotherapy.
Conclusion
The reproducibility of voxel-based ADC is not high enough even in a phantom
study. It would be recommended to pay enough attention when performing
voxel-based ADC study like histogram analysis for tumor ADC.
Acknowledgements
No acknowledgement found.References
1. Donati OF, Mazaheri Y, Afaq A, et al. Prostate cancer
aggressiveness: assessment with whole-lesion histogram analysis of the apparent
diffusion coefficient. Radiology 2014;271(1):143-152.
2. Liang HY, Huang YQ, Yang ZX, Ying D, Zeng MS, Rao SX.
Potential of MR histogram analyses for prediction of response to chemotherapy
in patients with colorectal hepatic metastases. European radiology 2015.
3. Lin Y, Li H, Chen Z, et al. Correlation of histogram analysis
of apparent diffusion coefficient with uterine cervical pathologic finding. AJR
American journal of roentgenology 2015;204(5):1125-1131.
4. Rodriguez Gutierrez D, Manita M, Jaspan T, Dineen RA, Grundy
RG, Auer DP. Serial MR diffusion to predict treatment response in high-grade
pediatric brain tumors: a comparison of regional and voxel-based diffusion
change metrics. Neuro-oncology 2013;15(8):981-989.