Can Cramer-Ráo Lower Bound be used to find optimal b-values for IVIM?
Oscar Gustafsson1,2, Maria Ljungberg1,2, and Göran Starck1,2

1Department of Radiation Physics, Institute of Clinical Sciences, The Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden, 2Department of Medical Physics and Biomedical Engineering, Sahlgrenska University Hospital, Gothenburg, Sweden

Synopsis

Cramer-Ráo Lower Bound is commonly used in experiment design optimization. Here we use it to find optimal b-value schemes for IVIM imaging. The optimization was generalized with regard to averaging and the characteristics of the results, given the input and the constraints, were studied. The resulting schemes never included more than the minimum number of four unique b-values, even though multiple sets of tissue parameters were included in the optimization. The optimized b-value schemes were compared to a typical one using simulations.

Introduction and purpose

Intravoxel incoherent motion (IVIM) imaging1 has gained renewed attention during the last years due to its potential of simultaneously monitoring tissue diffusion and perfusion completely non-invasively. However, to this date there is no consensus with regard to which b-values should be used. In this study we investigate the potential of Cramer-Ráo Lower Bound (CRLB), which is a commonly used tool for optimal experiment design2. The effect of averaging is considered and generalized compared to previous studies3.

Theory

IVIM is described by a biexponential model1: $$ S=S_0((1-f)e^{-bD}+fe^{-b(D+D^{*})})$$ where D is the diffusion coefficient, D* is the pseudodiffusion coefficient, f is the perfusion fraction and S0 is the signal without diffusion weighting. Given a model the CRLB sets a lower limit of the parameter uncertainties: $$\sigma ^2_{w_i} \leq (F^{-1})_{ii}$$ where $$$\sigma^2_{w_i}$$$ is the actual parameter uncertainty, $$$w \in\{D,D^*,f,S_0\}$$$ and F is Fisher's information matrix. Given Gaussian noise and considering the effect of averaging, F has elements: $$ F_{ij} = \sigma^{-2}\sum_{k=1}^K n_k \frac{\partial S_k}{\partial w_i}\frac{\partial S_k}{\partial w_j}$$ where $$$\sigma^2$$$ is the noise variance, nk is the number of averages for the k:th b-value, Sk is the expected signal for the k:th b-value given the IVIM model. The noise level and the total number of measurements $$$(N=\sum^K_{k=1}n_k)$$$ are constants, therefore a matrix M may be used in the optimization instead of F: $$ M_{ij}=\frac{\sigma^2}{N}F_{ij}=\frac{\sigma^2}{N}\sigma^{-2}\sum_{k=1}^K n_k \frac{\partial S_k}{\partial w_i}\frac{\partial S_k}{\partial w_j}=\sum_{k=1}^K \alpha_k \frac{\partial S_k}{\partial w_i}\frac{\partial S_k}{\partial w_j}$$ where $$$\alpha_k = \frac{n_k}{N}$$$, which implies that $$$0\leq\alpha_k\leq1$$$ and $$$\sum^K_{k=1}\alpha_k=1$$$. Considering multiple sets of tissue parameters and adjusting for the different magnitude of the model parameters the total relative uncertainty, which is to be minimized, is given by: $$C = \sum^S_{s=1}\gamma_s\sum_{i=1}^3\frac{(M_s^{-1})_{ii}}{w_{i,s}^2}$$ where $$$\gamma_s$$$ is a weighting factor. The uncertainty of S0 is omitted in the inner sum as it is not of primary interest.

Materials and methods

Constrained minimization was performed in MATLAB for three sets of tissue parameters (table 1) separately and combined ($$$ \gamma = [\frac{1}{200^2}, \frac{1}{30^2}, \frac{1}{20^2}] $$$ weights from Lemke4) to obtain a total of four optimal b-value schemes. To assess the effects of boundary conditions the optimizations was performed for multiple lower and upper limits of b-values (blower = [0,10,20], bupper = [600,1000,2000]).

To assess the effect of b-value scheme on the model parameter uncertainties data was simulated for all tissue parameter sets by adding Rician noise to signal values given by the IVIM model. The model parameters were obtained by weighted non-linear least squares fitting. The uncertainties were characterized by IQR normalized to the true tissue parameter values. This was compared to simulated data using a b-value scheme comparable to those commonly seen in the literature (blit=[0,10,20,30,40,50,75,100,200,300,400,600]). To match the total number of acquisitions the proportions of the optimal schemes were multiplied by 12 and b-values were rounded to the nearest multiple of ten.

Results and discussion

For all sets of tissue parameters the optimal b-value scheme included no more than four unique b-values with non-zero proportion (up to 10 unique b-values were allowed) (table 2). This is in agreement with previous observations3. The same result was produced when the optimization was performed over a weighted sum of multiple tissue parameter sets ($$$b_{sum} = [0,30,220,600],\alpha_{sum}=[0.18,0.32,0.32,0.18] $$$). Limits had small effect on the α:s for most settings. Increased lower limit/decreased upper limit mainly affected the lower/higher b-values respectively (table 2).

The CRLB b-value schemes gave smaller model parameter uncertainties in 11 of the 18 cases (table 3). The best gain in using the CRLB scheme was seen for the medium perfusion parameter set, which is characterized by relatively high f and low D*.

Conclusions

CRLB optimization favors averaging of the minimum number of unique b-values rather than adding more unique ones. In practice the optimized b-value schemes were found to decrease the model parameter uncertainty in simulated data with mid-range perfusion.

Acknowledgements

This research was supported by the Swedish Research Council, the Swedish Cancer Society, the King Gustaf V Jubilee Clinic Cancer Research Foundation in Gothenburg, Sweden and grants from the Swedish state under the LUA/ALF agreement.

References

1. Le Bihan et al. 1988 Separation of Diffusion and Perfusion in Intravoxel Incoherent Motion MR Imaging. Radiology

2. Alexander. 2008 A General Framework for Experiment Design in Diffusion MRI and Its Application in Measuring Direct Tissue-Microstructure Features. Magnetic Resonance in Medicine

3. Leporq et al. 2014 Optimization of Intra-voxel Incoherent Motion Imaging at 3.0 Tesla for Fast Liver Examination. Journal of Magnetic Resonance Imaging

4. Lemke et al. 2011 Toward an optimal distribution of b values for intravoxel incoherent motion imaging. Magnetic Resonance Imaging

Figures

Table 1. Tissue perfusion parameters used in this study and previously by Lemke4

Table 2. The effect of b-value limits on optimal b-values and proportionality coefficients (b and α). The first three rows show the effect of changing the upper limit whereas the last three rows show the effect of changing the lower limit. b-values were rounded for better readability (nearest 5 for b < 50 s/mm2 and nearest 10 for b > 50 s/mm2) Note: in some cases the displayed proportions do not add up to 1 due to rounding error.

Table 3. Relative uncertainties for the model parameters, from fitting of simulated data, calculated as IQR/true value. Diff was calculated as (IQR(Literature)/IQR(CRLB) – 1). As a result, positive numbers indicate better performance for the CRLB scheme and vice versa.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
2043