Tomoya Nakamura1, Shuhei Shibukawa2, Yuma Sainokami2, Tomohiko Horie1, Isao Muro2, Terumitsu Hasebe1, Yutaka Imai2, and Tetsuo Ogino3
1Tokai University Hachioji Hospital, Hachioji, Japan, 2Tokai University Hospital, Isehara, Japan, 3Philips Healthcare Asia Pacific, Shinagawa, Japan
Synopsis
The purpose of this study is to
assess the fractional anisotropy (FA) of heart using ECG gating and second
moment nulling pulse which is intrinsically insensitive to motion.
The FA at motion correction (MC) gradient was significantly higher than
at acceleration motion correction (aMC) gradient, therefore, cardiac motion artifact resluts in an overestimation of FA.
In
conclusion, the use of second order motion correction gradient enables the
quantification of FA at heart and has the potential to contribute to clinical
cardiac imaging.Introduction
Although conventional-diffusion-weighted imaging (C-DWI)
of the human heart is sensitive to motion, we reported the simultaneous
use of electrocardiogram (ECG) gating and motion correction (MC), including
second-order velocity compensation makes it possible to assess the water
molecular dynamics [1,2]. The purpose of this study is to assess the fractional
anisotropy (FA) of heart using ECG gating and second moment nulling pulse which
is intrinsically insensitive to motion.
Methods
MC is achieved by a bipolar gradient pulse that compensates
the phase dispersion caused by the velocity term of motion. Further, motion correction
until the acceleration term of motion is achieved by second-moment nulling
pulse, shape is shown as Figure 1. Cardiac diffusion tensor imaging was
done with ECG gating and free breathing on a 1.5T scanner (Achieva nova dual,
Philips Healthcare) using a 32-channel coil. Each b-value of 300 and 600s/mm
2
was set for the acceleration motion correction DWI (aMC-DWI) and the MC-DWI,
and TE value was 78ms (shortest TE of aMC-DWI, b-value of 600s/mm
2). Six
healthy volunteers (age 23-28) were scanned and written informed consent was
obtained from all volunteers. FA was calculated and compared among aMC and MC
sequence, and the statistical difference between each secquence was determined
by Mann Whitney-U test. Moreover, FA color map was compared between aMC and MC
sequence. Further parameters were; TR: 5 heartbeats, FOV: 280mm, RFOV:
80%, voxel size: 8×2.19×2.43 mm
3, SENSE fc: 2, half scan fc: 0.6,
NSA: 6, fat suppression: SPIR.
Results
Figure 2 shows the FA at aMC and
MC in each b-value, the FA at aMC was significantly lower than at MC in each b-value. Figure 3 shows the FA color map at aMC and MC, fiber orientation in each
sequence was also varied.
Discussion and Conclusion
The FA at MC was significantly higher
than at aMC, therefore, cardiac motion artifact resluts in an overestimation of
FA. Moreover, first order motion correction gradient which is sensitive to bulk
motion also caused the visualization of fiber architecture to vary as well. In
conclusion, the use of second order motion correction gradient enables the
quantification of FA at heart and has the potential to contribute to clinical
cardiac imaging.
Acknowledgements
No acknowledgement found.References
[1] Nakamura et.al, 24th Annual Meeting of ISMRM, Toronto,
2015 (Abstract2015), [2] Nakamura et.al, Nichidoku-Iho 2015 (in press).