Lumeng Cui1,2, Conrad Yuen3, Ted Lynch4, Paul Babyn5, Francis M. Bui6, and Niranjan Venugopal1,5
1Department of Medical Physics, Saskatchewan Cancer Agency, Saskatoon, SK, Canada, 2Division of Biomedical Engineering, University of Saskatchewan, Saskatoon, SK, Canada, 3Department of Medical Physics, BC Cancer Agency, Vancouver, BC, Canada, 4Non-ionizing Radiation, CIRS Inc., Norfolk, VA, United States, 5Department of Medical Imaging, University of Saskatchewan, Saskatoon, SK, Canada, 6Department of Electrical & Computer Engineering, University of Saskatchewan, Saskatoon, SK, Canada
Synopsis
Magnetic Resonance Elastography (MRE) is a new imaging technique
that combines the acoustic waves and MRI to retrieve elastic properties of
tissue. Because MRE is non-invasive, there is great clinical interest for its
use in the detection of cancer. In this work, we focus on the design of an MRE
phantom to be used in the clinical commissioning of an MRE System. With the aid
of newly designed pulse sequences and inversion algorithms we have developed a
quality assurance process to validate the efficacy of MRE for applications to many
clinical sites (i.e. prostate, cervix, uterus).Introduction
Magnetic Resonance Elastography (MRE) is a new imaging technique
that combines the acoustic waves and Magnetic Resonance Imaging (MRI) to retrieve
elastic properties of tissue
1.
Because MRE is non-invasive, there is great potential and interest for its use in the detection of cancer
1–4. Many
institutes have dedicated a number of studies determining the optimal
experimental parameters and methods to perform MRE in vivo
2,5–8. At
present, the only commercial FDA approved system available for clinical use is
the Resoundant System (Rochester,
MN). To commission the MR Elastography system, prior to clinical implementation,
a set of quality control (QC) tests are needed to verify the efficacy of the MR
elastography data (i.e. reproduce predefined stiffness values from phantom
data). At present, there are no commercially available phantoms
that fit the dimensions of the Resoundant system. As well, limitations using scanner
software do not allow for a thorough evaluation over a range of stiffness’s
required for quality assurance. In this work, we focus
on designing a practical MRE phantom that works alongside the Resoundant MRE
System. With the aid of newly designed pulse sequences and inversion algorithms
we are developing a quality assurance process to validate the efficacy of MRE
for applications to other clinical sites (i.e. prostate, cervix, uterus).
Methods
and Materials
To verify the efficacy of the MRE technique in a clinical setting,
a set of QC tests was performed with two phantoms: an elasticity phantom
containing rods of varying stiffness (049A, CIRS Inc., Norfolk, VA) and a
prostate phantom containing intraprostatic lesions (053L CIRS Inc., Norfolk,
VA). All experiments were carried out on a 3T Skyra platform (Siemens
Healthcare, Erlangen, Germany) with the Resoundant system. We employed the
Siemens production pulse sequence which encodes the acoustic motion using a 2D
gradient echo-based imaging technique. MRE image acquisitions were also repeated
using a newly developed spin-echo planar imaging technique (Siemens WIP 923,
VE11A). Elastograms were reconstructed using the Siemens scanner inversion
program, and MRE/Wave inversion software
9.
Results
and Discussions
The dimensions of the paddle used
by the Resoundant system is 18.5 cm in diameter, however the surface dimension for
049A phantom is 17×10 cm. This
mismatch in the size of paddle and the phantom causes a number of problems.
Firstly, the housing of phantom is not flat with the scanning surface as shown
in Figure 1 (A). The air gap between the paddle and scanning surface creates an
interface between the wave source and the phantom. Additionally, the small housing causes apparent
reflections of the shear waves at the sides of housing wall, and therefore generates
significant interference and loss of data integrity
10 (as seen by the checkered region
in Figure 2). In order to address these issues, a novel phantom was designed based
on the 049A phantom, as seen in Figure 1 (D). The original 049A gel block
(white area) was placed in a larger housing filled with a dispersive Zerdine
gel (grey area) which minimizes wave reflections from the side walls. As well, the
new housing was designed with a larger diameter to fit Resoundant paddle. In
figures 3 B and C, we present reconstructed stiffness maps calculated by two
different inversion algorithms, the Siemens product software, and the MRE/Wave
inversion software. For Lesion #1, the
mean shear stiffness was 2.59±0.70 kPa versus
11.82±1.10 kPa respectively. The urethra was 1.63±0.81 kPa versus 14.03±1.72 kPa, and for lesion #2, it was 2.20±0.83 kPa versus 11.04±1.67 kPa. The reason for this difference is
because the current scanner’s inversion program was designed for the Liver, and
as such limits the stiffness range to 0-8 kPa. The large discrepancies observed
demonstrates the limited applications of the MRE system to other sites (i.e.
cervix cancer, prostate cancer, etc.) using the system configuration as is. But
using the newer Siemens WIP, Resoundant system, and MRE/Wave together expands
the range of applications for which MRE can be used. Utility of the modified
phantom allows us to perform QC tests to verify the reproducibility of the MRE
technique for varying stiffness.
Conclusion
In this work we have identified the limited utility of current
commercial phantoms available for routine QC tests using the Resoundant MRE
system. In collaboration with the manufacturer, we developed a phantom for
robust verification of the MRE system over a wide range of stiffness’s, thus
expanding the clinical utility to other clinical sites (i.e. prostate, cervix, uterus).
Acknowledgements
No acknowledgement found.References
1. Mariappan,
Y. K., Glaser, K. J. & Ehman, R. L. Magnetic resonance elastography: A
review. Clin. Anat. 23, 497–511 (2010).
2. McKnight,
A. L. et al. MR elastography of breast cancer: preliminary results. Am.
J. Roentgenol. 178, 1411–1417 (2002).
3. Frangioni,
J. V. New technologies for human cancer imaging. J. Clin. Oncol. 26,
4012–4021 (2008).
4. Venkatesh,
S. K. et al. Magnetic resonance elastography of liver tumors-preliminary
results. AJR Am. J. Roentgenol. 190, 1534 (2008).
5. Kruse,
S. A. et al. Tissue characterization using magnetic resonance
elastography: preliminary results. Phys. Med. Biol. 45, 1579
(2000).
6. Plewes,
D. B., Bishop, J., Samani, A. & Sciarretta, J. Visualization and
quantification of breast cancer biomechanical properties with magnetic
resonance elastography. Phys. Med. Biol. 45, 1591 (2000).
7. Green,
M. A., Bilston, L. E. & Sinkus, R. In vivo brain viscoelastic properties
measured by magnetic resonance elastography. NMR Biomed. 21,
755–764 (2008).
8. Sahebjavaher,
R. S., Baghani, A., Honarvar, M., Sinkus, R. & Salcudean, S. E.
Transperineal prostate MR elastography: Initial in vivo results. Magn.
Reson. Med. 69, 411–420 (2013).
9. MRE/Wave
- Magnetic Resonance Imaging Lab: Richard L. Ehman - Mayo Clinic Research. at
<http://www.mayo.edu/research/labs/magnetic-resonance-imaging/mre-wave?_ga=1.251021446.1640793623.1433887874>
10. Yin, M., Manduca, A. & Grimm, R. C. in Magnetic
Resonance Elastography (eds. Venkatesh, S. K. & Ehman, R. L.) 19–37
(Springer New York, 2014).