Yohei Itoh1, Yasuo Takehara2, Naoki Ooishi2, Masanori Kawade2, Tetsuya Wakayama3, Mikio Suga4, Takasuke Ushio1, Yuki Hirai1, Nobuko Yoshizawa1, Shuhei Yamashita1, Hatsuko Nasu1, and Harumi Sakahara1
1Diagnostic Radiology & Nuclear Medicine, Hamamatsu University school of medicine, Hamamatsu, Shizuoka, Japan, 2Department of Radiology, Hamamatsu University Hospital, Hamamatsu, Shizuoka, Japan, 3GEHCJ, Hino-shi, Tokyo, Japan, 4Center for Frontier Medical Engineering, Chiba University, Chiba, Japan
Synopsis
To achieve the high spatial resolution MR elastography(MRE), it has been reported that combining SS-SE-EPI with a spatially selective excitation provides an efficient way of reducing FOV. Using this technique, we performed phantom study and pancreatic MRE. We report the result of our study and the point we found when performing the reduced-FOV MRE.Introduction
To achieve higher spatial resolution in MR elastography (MRE), it has been reported that the combination of SS-SE-EPI with a spatially selective excitation efficiently reduce FOV in the phase encoding direction without wrap around artifacts thereby increase the spatial resolution [1]. This local excitation technique is called FOCUS (FOV Optimized and Constrained Undistorted Single-shot) that may extend the application for relatively small organs such as the pancreas [2]. In this study, we sought to optimize FOCUS-MRE to the pancreas using homogenous phantom and a volunteer.
Purpose
1) To compare the shear stiffness value of the homegeneous phantom between conventional-MRE and FOCUS-MRE .
2) To optimize the parameters for FOCUS-MRE.
3) To optimize the form of FOV placement for the pancreatic FOCUS-MRE.
Materials & Methods
In this study,
we used a 2D echo-planar RF pulse instead of spatial spectral RF pulse for
excitation in the SS-SE-EPI sequence. Only the region of interest in the phase
FOV direction can be excited without incurring aliasing artifacts, as a result
of which fewer phase encodings are required to achieve the desired spatial
resolution compared to the conventional-MRE. All examinations were conducted on
a clinical MR scanner (Discovery 750w 3.0T, GE Healthcare,) with body AA and PA
phased array coil with 32 elements. The parameters for each method were as
follows: TR/TE = 1000/53.4-77.3 msec (conentional-MRE) and 1000/59.5-77.5 msec (FOCUS-MRE),
motion-encoding gradients (MEG) direction = z. In both cases, all MRE images
were acquired axial with three slices. The wave images and elastograms were
automatically generated by MR-Touch system.
We used the polyacrylamide
gel (70 wt% glycerin) phantom for standardwith storage modulus (G') of 3.1 kPa (rheometer)
[3]. The phantom is 16*18cm cylindrical
shape. The acoustic driver was placed on the top of the phantom.
1) Shear
stiffness was measured on the phantom for the conventional-MRE and FOCUS-MRE varying
MEG and driver frequencies. 2) In FOCUS-MRE, various size and the position of
FOV placement was tested.
3) Using
FOCUS-MRE, pancreatic MRE was tested on a normal volunteer.
Result
1) In
conventional-MRE, the larger the driver frequency was, the higher the measured
shear stiffness. In FOCUS-MRE, the measured stiffness were about 8% lower than
that of normal-MRE. In both MRE, the difference of MEG did not affect shear
stiffness (Table 1).
2) Figure
1&2 show how elastogram and wave images vary according to the forms of the FOV
placement with FOCUS-MRE. As for FOCUS-MRE, when the whole FOV was placed inside
of the phantom (AB), the elastogram suffered from inhomogeneity and the artifacts.
Waves were not parallel on the wave images. When the FOV was placed off-set
including the remote side of the phantom (CDEF), the elastograms were more homogeneous
and the waves were parallel on the wave images. When the FOV was placed off-set
including the lateral half side of the phantom, liner artifact occurred in the
edge of magnitude image and this artifact seemed to affect the wave image and
elastogram.
3) Figure
3&4 show the result of volunteer study for the pancreatic MRE. When the whole
FOV was placed inside the body (J), the elastogram and wave image was
unsatisfactory similar to the phantom study (AB). When the FOV was placed
including the lateral side of the body (K), the artifact occured in some cases
(arrow ; magnitude image of K) and the analysis was seems to be affected (under-bar
; elastogram and wave image of K). This phenomenon was reproduced similar to
the phantom study (GH). When the FOV was placed off-set including the proximal
and remote side of the body (LM like phantom study CE), the obtained image of
elastogram and wave image were better than others (JK).
Conclusion
In the phantom
study, the measured stiffness value of FOCUS-MRE is about 8% lower than those
of conventional-MRE; however, the values were closer to that measured with
rheometer. In FOCUS-MRE, FOV should be placed off-set including proximal or remote
side of the object, furthermore, inclusion of the lateral side of the object is
not recommended, because it may result in the artifacts and the resultant
stiffness values may be affected. Pancreatic MRE using FOCUS-MRE, FOV may be
better placed off-set including ventral or dorsal side of the body.
Acknowledgements
No acknowledgement found.References
1. Emine US, et al, MRM 60: 468-473 (2008)
2. Feng Z, et al, Clin Imaging. 2015 Sep-Oct;39(5):851-5
3. Suga M, et al, ECR,C-0757(2015)