Impact of Field Strength and Image Resolution on MRE Stiffness Estimation
Eric Barnhill1, Jing Guo2, Florian Dittmann2, Sebastian Hirsch2, Michael Perrins3, Lucy Hiscox3, Tim Herrmann4, Johannes Bernarding4, Neil Roberts3, Jürgen Braun1, and Ingolf Sack2

1Institute of Medical Informatics, Charité Universitätsmedizin Berlin, Berlin, Germany, 2Department of Radiology, Charité Universitätsmedizin Berlin, Berlin, Germany, 3Clinical Research Imaging Centre, The University of Edinburgh, Edinburgh, United Kingdom, 4Institute for Biometrics and Medical Informatics, Otto von Guericke University Magdeburg, Magdeburg, Germany

Synopsis

We investigated the impact of field strength and image resolution on brain MRE stiffness results. A cohort of 18 healthy volunteer subjects was scanned at 1.5T (2mm isotropic voxels), 3T (2mm) and 7T (1mm), with a fourth set downsampling the 7T to 2mm. Means were 1634 Pa (+/-613) for 1.5T, 1743 Pa (+/-811) for 3T, 1786 (+/-634) Pa for 7T 2mm, and 927 (+/-364) Pa for 7T 1mm. In the paired sign-rank tests, there were no significant effects for field strength. Examination of histograms of example slices suggests that a different distribution of features is being captured at the higher resolution.

Introduction

Recent work has produced high-resolution Magnetic Resonance Elastography (MRE) viscoelastic parameter maps, including brain MRE at different field strengths, vibration frequencies and image resolutions [1-5]. This variability in methods caused considerable variation of stiffness estimates ranging from 1.08 kPa (at 40,50,60 Hz) [4] to 3.7 (at 45 Hz) [3] for in vivo white matter. Since brain tissue features spatial heterogeneities at multiple scales, an improved image resolution, as is attainable at ultrahigh fields, would add detail within broader shear wavelengths. We hypothesize that fine features increase wave curvature which is expected to lower wavelength-based stiffness estimates.

Purpose

Here we investigated the impact of field strength and image resolution on brain MRE stiffness results obtained by multifrequency dual elasto visco (MDEV) inversion [4].

Methods

A cohort of 18 healthy volunteer subjects was scanned at 1.5T, 3T and 7T using the protocol described in [4]. We used three drive frequencies of 30, 40 and 50 Hz. The 1.5T and 3T scans had isotropic voxel resolution of 2mm (20 contiguous transversal slices). The 7T scan had isotropic voxel resolution of 1mm (42 contiguous transversal slices). To investigate the impact of feature fineness apart from field strength, a fourth data set was evaluated in which the 7T set was downsampled to 2mm isotropic voxels using the Matlab function 'imresize' (MathWorks, Natick, MA, USA). Images were phase unwrapped using the Laplacian based technique in [7] and denoised using divergence-free wavelets [8] which were thresholded with SureShrink using median absolute deviation noise estimates, and complex dualtree wavelets [9] thresholded with Overlapping Group Sparsity [10] using a Principal Component Analysis-based blind noise estimation technique [11]. For measurements, the cortex was thresholded by measuring the mean and standard deviation (SD) of a pure noise region in each image, then masking all values below two SD's above the mean noise level. Mean and SD of each masked volume were measured. Example slices and histograms of these images were also examined. Each of the four cohorts was tested for significance of field strength effect using the Wilcoxon signed rank test.

Results

Plots of mean and SD for each subject at each field strength are shown in Figure 1. Means across the cohort for 1.5T, 3T and 7T_2mm resampling were 1634 Pa (+/-613), 1743 Pa (+/-811) and 1786 (+/-634) Pa respectively, while mean for the original 7T 1mm resolution was 927 (+/-364) Pa. In the paired sign-rank tests, there were no significant effects for field strength between any of the three 2mm acquisitions (p values of 0.13 for 1.5T vs 3T, 0.84 for 1.5T vs. 7T_2mm, and 0.38 for 3T vs. 7T_2mm) while the impact of 1mm vs. 2mm voxel edge lengths was significant with p < 0.001 in all cases.

An example slice from each field strength for a single subject chosen at random is in Figure 2. Below it are the histograms for the masked brain region in each case. It can be seen that while the three 2mm voxel edge length histograms are roughly symmetrical, the 7T 1mm histogram is qualitatively different with heavy right skew.

Discussion

In agreement to previous work, our study shows no significant impact of field strength on stiffness estimation at identical voxel size [6]. The impact of a 1mm isotropic resolution on stiffness values, however, is highly significant. Examination of histograms of example slices suggests that a different distribution of features is being captured at the higher resolution, supporting the hypothesis that the additional fine viscoelastic features detected at the higher field strength create more tortuosity in the wave, lowering a wavelength-based stiffness estimate.

Conclusion

In conclusion, the MDEV inversion protocol appears robust to field strength, with lower elastogram values in higher-resolution elastograms likely a function of finer detail in the image creating more curvature in the shear waves.

Acknowledgements

The authors gratefully acknowledge image acquisitions from the University of Magdeburg.

References

[1] Zhang J, Green MA, Sinkus R, Bilston LE. Viscoelastic properties of human cerebellum using magnetic resonance elastography. J Biomech 2011;44(10):1909-1913.

[2] Johnson CL, McGarry MD, Gharibans AA, Weaver JB, Paulsen KD, Wang H, Olivero WC, Sutton BP, Georgiadis JG. Local mechanical properties of white matter structures in the human brain. NeuroImage 2013;79:145-152.

[3] Clayton EH, Genin GM, Bayly PV. Transmission, attenuation and reflection of shear waves in the human brain. J R Soc Interface 2012;9(76):2899-2910.

[4] Braun J, Guo J, Lützkendorf R, Stadler J, Papazoglou S, Hirsch S, Sack I, Bernarding J. High-resolution mechanical imaging of the human brain by three-dimensional multifrequency magnetic resonance elastography at 7T. NeuroImage 2014;90:308-314.

[5] Murphy MC, Huston J, 3rd, Jack CR, Jr., Glaser K, Senjem ML, Chen J, Manduca A, Felmlee J, Ehman RL. Measuring the Characteristic Topography of Brain Stiffness with Magnetic Resonance Elastography. PLoS One 2013; 8(12): e81668.

[6] Hamhaber U, Klatt D, Papazoglou S, Hollmann M, Stadler J, Sack I, Bernarding J, Braun J. In vivo magnetic resonance elastography of human brain at 7 T and 1.5 T. J Magn Reson Imaging 2010;32(3):577-583.

[7] Barnhill, E., Kennedy, P., Johnson, C. L., Mada, M., & Roberts, N. (2015). Real-time 4D phase unwrapping applied to magnetic resonance elastography.Magnetic Resonance in Medicine, 73(6), 2321-2331.

[8] Ong, F., Uecker, M., Tariq, U., Hsiao, A., Alley, M. T., Vasanawala, S. S., & Lustig, M. (2015). Robust 4D flow denoising using divergence-free wavelet transform. Magnetic Resonance in Medicine, 73(2), 828-842.

[9] Liu, X., Tanaka, M., & Okutomi, M. (2013). Single-image noise level estimation for blind denoising. Image Processing, IEEE Transactions on, 22(12), 5226-5237.

[10] Selesnick, I. W., Baraniuk, R. G., & Kingsbury, N. G. (2005). The dual-tree complex wavelet transform. Signal Processing Magazine, IEEE, 22(6), 123-151.

[11] Chen, P. Y., & Selesnick, I. W. (2014). Group-sparse signal denoising: non-convex regularization, convex optimization. Signal Processing, IEEE Transactions on, 62(13), 3464-3478.

Figures

Figure 1. Mean complex modulus magnitude |G*| of each subject at each field strength, with standard deviation error bars.

Figure 2. An example slice of a subject chosen at random, and the histogram of the masked brain for each slice. a-c: 2mm edge length, d: 1mm edge length showing a qualitatively different distribution than the others.



Proc. Intl. Soc. Mag. Reson. Med. 24 (2016)
1952