Ian Gavin Murphy1, Martin Graves2, Scott Reid3, Andrew Patterson2, Ilse Gavin Joubert1, Andrew N Priest2, and David J Lomas2
1Radiology, Cambridge University Hospitals NHS Foundation Trust, Cambridge, United Kingdom, 2Radiology, Cambridge University NHS Foundation Trust, Cambridge, United Kingdom, 3GE, Little Chafont, United Kingdom
Synopsis
In patients with liver disease, MR elastography (MRE) is a non-invasive method for evaluating fibrosis. MRE is phase-based and sensitive to motion artefact, and is typically performed in end expiration. We found that navigator timed MRE shows no statistical difference to breath-held techniques for stiffness and reproducibility in 6 healthy volunteers, and may prove superior in patients unable to adequately hold their breath Background
MR elastography is a fast-evolving field with increasing evidence for its
use in the diagnosis of liver fibrosis. Fibrotic livers have, amongst other
factors, a higher collagen content which results in a increase in stiffness that can be quantified using
magnetic resonance elastography (MRE).
Technical difficulties
exist however, which may lead to variability in the measured liver stiffness. Standard
techniques are phase-based and are therefore sensitive to motion artefact, such
as can be seen with breathing.1 MRE in the liver is typically
performed at end-expiration, and requires four breath holds.2, 3 Sequential breath
holds may result in slightly differing diaphragmatic positions and therefore
different position of the liver and other viscera, which may result in
misregistration. In
addition, some patients may also not be able to accommodate the breath-holds. Respiratory
triggering, using navigators, is an alternative method to breath-hold
acquisitions but is not supported in the product MRE sequences. The aim of
this study is to compare
liver stiffness measurements using breath-hold, navigator-gated, and free breathing MRE
Methods
Ethical approval was provided for the study. The studies were carried out on six healthy volunteers, 3
male and 3 female, with a mean age of 28 years, with full informed consent.
Imaging was peformed using Examinations
were performed
on a
1.5T whole-body
MRI
scanner
(Discovery MR450,
General Electric,
Waukesha ha,
WI)
using an
eight-channel
receive-only
phased-array
coil.
Examinations
were performed
on a
1.5T whole-body
MRI
scanner
(Discovery MR450,
General Electric,
Waukesha ha,
WI)
using an
eight-channel
receive-only
phased-array
coil.
Examinations
were performed
on a
1.5T whole-body
MRI
scanner
(Discovery MR450,
General Electric,
Waukesha ha,
WI)
using an
eight-channel
receive-only
phased-array
coil.
Examinations were performed on a 1.5T whole-body MRI scanner (Discovery
MR450, GE Healthcare, Waukesha, WI) using an eight-channel
array coil.
For the MRE acquisition a passive 18.5-cm-diameter pneumatic driver was
placed anteriorly over the right lower ribs superficial to the right lobe of the
liver. The passive driver was connected to an active drive unit producing shear waves at 60 Hz. The product gradient-echo based MRE sequence was modified to incorporate a 2D cylindrical
excitation navigator. Sequence parameters were TE/TR =22/50ms, matrix 256x64,
FOV = 40x36cm, slice thickness=8 mm, gap=5mm, bandwidth = ±31.25kHz, and flip
angle =30°. A parallel imaging (ASSET) acceleration factor of 1.5 was used. Four slices were acquired with four phase offsets. In the breath-hold
acquisition each offset was acquired in a separate breath-hold. MRE shear modulus-based stiffness and confidence interval maps were subsequently computed.
Each subject was imaged twice using the free-breathing, breath-held and navigator-triggered methods. The order in which the
sequences were acquired was randomized.
The stiffness and the percentage of analyzable liver area were
calculated at a per subject level by determining the mean across all 4 slices
(Figure 1).
The repeatability of each technique was determined by computing the
absolute difference in MRE liver stiffness between repeat scans 1 and 2.
The Kruskal-Wallis test was performed to assess if there was an overall
difference between each the MRI acquisition strategies.
Pair-wise comparisons were performed using Wilcoxon’s rank-sum test.
Results
The percentage mean area of liver
where the stiffness was quantifiable (confidence interval >95%) and mean
stiffness values are summarized in Table 1.
No statistically significant differences in liver stiffness were noted between
acquisition strategies (p=0.199) (Figure 2). We note that liver stiffness
during inspiration is elevated relative to expiration as previously noted in
the ultrasound literature 4 , however the difference was not
significant (p=0.240). The navigator method was not significantly different to
the standard expiration (p=0.589) or the inspiration (p=0.180) methods.
A comparison of analyzable liver areas (where CI >95%) demonstrates
that the overall group difference is significantly different (p=0.002) (Figure
3). The free breathing acquisition is notably lower than the other methods.
The group-wise comparison of MRE repeatability (Figure 4) was not
statistically significant (p=0.142), however, the free breathing method
demonstrates the greatest variability. The navigator method has comparable
repeatability to the expiration and inspiration methods (p=0.520 and p=0.521
respectively).
The mean acquisition time for the navigator
triggered sequence was 4 minutes 55 seconds.
Discussion
The results show that MRE using
navigator-triggering gives results similar to the gold-standard end-expiratory breath
hold technique in healthy volunteers. The technique may therefore be useful in
patients unable to perform extended breath-holds.
Acknowledgements
ACT
MRIS Radiographers, Addenbrooke's Hospital, Cambridge, UK
NIHR Cambridge BRC
References
1. Zhuo J, Gullapalli RP. Radiographics.
2006 Jan-Feb;26(1):275-97
2. Venkatesh SK et al, JCAT, 2013 Nov-Dec;37(6):887-96
3. Godfrey EM et al, Eur
Radiol. 2012 Dec;22(12):2790-7
4. Ling W et al, Eur J Radiol. 2013
Feb;82(2):335-41